Quantifying the trigger level of the vacuum surveillance system of the Gamma-Knife eXtend™ positioning system and evaluating the potential impact on dose delivery.

Béatrice Reiner1,2, Peter Bownes3, David L Buckley1, David I Thwaites1,4.   

Abstract

OBJECTIVE: This work evaluates the precision and characteristics of the trigger level of the vacuum surveillance system of eXtend™ on Gamma Knife Perfexion and the effect of the potential displacement on the dose distribution. METHODS AND MATERIALS: A total of 20 individually moulded mouthpieces based on human dental models were used to measure translational shift and rotation until the vacuum surveillance of eXtend interrupted the irradiation. The positional accuracy of the movement was 0.01 mm using a computerized numerically controlled positioning system. Rotation was introduced by peripheral pressure in superior or inferior direction on the mould and was measured with a digital inclinometer. In 10 patients with a large brain metastasis the effect of a potential displacement of the centre of the target was recalculated. Two out of the 10 targets were located near the optic nerve and chiasm. In addition, the potential displacement of the chiasm and the optic nerve due to rotation based on their distance to the centre of rotation (COR, the mouthpiece) was evaluated. For the recalculation of the delivered dose after a displacement, GammaPlan Software® was used. Dose distribution with displacement between ±0.1 mm and ±2.0 mm was simulated for ten targets with volumes between 7.7 cm3 and 19.3 cm3. Shifts were applied to the X (lateral), Y (ventro-dorsal) and Z axis (superior-inferior). Dwell time was kept constant within <0.5% total time and 1% individual dwell time. Evaluated parameters were "minimum dose" (D(99%)), "coverage" and "Paddick Conformity Index" (PCI).
RESULTS: The vacuum between mouthpiece and dental model was broken by a mean translation of the mouthpiece of 0.15 mm (SD ±0.05 mm, range 0.05-0.29 mm) or a mean rotation of 0.33º (SD ±0.15º, range 0.05-1.0º). Rotation caused by pressure in the inferior direction was higher with 0.42º (SD ±0.16º) compared to rotation due to pressure in the superior direction with 0.21º (SD ±0.09º). The mean rotation would produce a mean displacement of 0.64 mm for the target, 0.39 mm for the chiasm and 0.33 mm for the optic nerve, for the group of patients studied. The biggest effect of positional deviation was seen in the minimum dose to the target, with 2.5% reduction for a 0.5 mm and 5% for a 1.0 mm shift in the Z-direction respectively.
CONCLUSIONS: The trigger level is sensitive and reproducible. A sudden movement by the patient, such as an attempt to get out or correct an uncomfortable position would interrupt irradiation immediately. The dominant effect for a displacement is due to rotation. The resulting displacement depends on the distance between target and COR (magnitude) and the location (direction). Therefore, lesions near chiasm or brainstem, which are closer to the COR, are less affected from a rotation than peripheral targets.

Entities:  

Keywords:  Gamma Knife; eXtend™; fractionated stereotactic radiosurgery; intra-fraction accuracy; vacuum surveillance

Year:  2016        PMID: 29296424      PMCID: PMC5658833     

Source DB:  PubMed          Journal:  J Radiosurg SBRT


  40 in total

1.  Inclusion of geometric uncertainties in treatment plan evaluation.

Authors:  Marcel van Herk; Peter Remeijer; Joos V Lebesque
Journal:  Int J Radiat Oncol Biol Phys       Date:  2002-04-01       Impact factor: 7.038

2.  Patterns of patient movement during frameless image-guided radiosurgery.

Authors:  Martin J Murphy; Steven D Chang; Iris C Gibbs; Quynh-Thu Le; Jenny Hai; Daniel Kim; David P Martin; John R Adler
Journal:  Int J Radiat Oncol Biol Phys       Date:  2003-04-01       Impact factor: 7.038

3.  Setup accuracy of the Novalis ExacTrac 6DOF system for frameless radiosurgery.

Authors:  Thierry Gevaert; Dirk Verellen; Koen Tournel; Nadine Linthout; Samuel Bral; Benedikt Engels; Christine Collen; Tom Depuydt; Michael Duchateau; Truus Reynders; Guy Storme; Mark De Ridder
Journal:  Int J Radiat Oncol Biol Phys       Date:  2011-04-07       Impact factor: 7.038

4.  Multisession cyberknife stereotactic radiosurgery of large, benign cranial base tumors: preliminary study.

Authors:  Francesco Tuniz; Scott G Soltys; Clara Y Choi; Steven D Chang; Iris C Gibbs; Nancy J Fischbein; John R Adler
Journal:  Neurosurgery       Date:  2009-11       Impact factor: 4.654

Review 5.  Management of brain metastases: the indispensable role of surgery.

Authors:  George Al-Shamy; Raymond Sawaya
Journal:  J Neurooncol       Date:  2009-04-09       Impact factor: 4.130

6.  Visual field preservation after multisession cyberknife radiosurgery for perioptic lesions.

Authors:  John R Adler; Iris C Gibbs; Putipun Puataweepong; Steven D Chang
Journal:  Neurosurgery       Date:  2008-02       Impact factor: 4.654

7.  Outcomes and cost-effectiveness of gamma knife radiosurgery and whole brain radiotherapy for multiple metastatic brain tumors.

Authors:  Wen-Yuen Lee; Der-Yang Cho; Han-Chung Lee; Hao-Che Chuang; Chun-Chung Chen; Jun-Lin Liu; Shih-Neng Yang; Ji-An Liang; Li-Hwei Ho
Journal:  J Clin Neurosci       Date:  2009-03-09       Impact factor: 1.961

Review 8.  Stereotactic radiosurgery in the treatment of brain metastases: the current evidence.

Authors:  Bodo Lippitz; Christer Lindquist; Ian Paddick; David Peterson; Kevin O'Neill; Ronald Beaney
Journal:  Cancer Treat Rev       Date:  2013-06-27       Impact factor: 12.111

9.  Gamma knife radiosurgery for orbital tumors.

Authors:  Min-Su Kim; Kwan Park; Jong Hyun Kim; Yoon-Duck Kim; Jung-Il Lee
Journal:  Clin Neurol Neurosurg       Date:  2008-07-23       Impact factor: 1.876

10.  Neurocognition in patients with brain metastases treated with radiosurgery or radiosurgery plus whole-brain irradiation: a randomised controlled trial.

Authors:  Eric L Chang; Jeffrey S Wefel; Kenneth R Hess; Pamela K Allen; Frederick F Lang; David G Kornguth; Rebecca B Arbuckle; J Michael Swint; Almon S Shiu; Moshe H Maor; Christina A Meyers
Journal:  Lancet Oncol       Date:  2009-10-02       Impact factor: 41.316

View more
  2 in total

1.  Intra-fraction motion gating during frameless Gamma Knife® Icon™ therapy: The relationship between cone beam CT assessed intracranial anatomy displacement and infrared-tracked nose marker displacement.

Authors:  Gavin Wright; Jannie Schasfoort; Natalie Harrold; Paul Hatfield; Peter Bownes
Journal:  J Radiosurg SBRT       Date:  2019

2.  Quantifying the effects of positional uncertainties and estimating margins for Gamma-Knife® fractionated radiosurgery of large brain metastases.

Authors:  Béatrice Reiner; Peter Bownes; David L Buckley; David I Thwaites
Journal:  J Radiosurg SBRT       Date:  2017
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.