Abstract

PURPOSE: The aim of this study was to compare the detection of brain metastases at 3 T using a 32-channel head coil with 2 different 3-dimensional (3D) contrast-enhanced sequences, a T1-weighted fast spin-echo-based (SPACE; sampling perfection with application-optimized contrasts using different flip angle evolutions) sequence and a conventional magnetization-prepared rapid gradient-echo (MP-RAGE) sequence.
MATERIALS AND METHODS: Seventeen patients with 161 brain metastases were examined prospectively using both SPACE and MP-RAGE sequences on a 3-T magnetic resonance system. Eight healthy volunteers were similarly examined for determination of signal-to-noise ratio (SNR) values. Parameters were adjusted to equalize acquisition times between the sequences (3 minutes and 30 seconds). The order in which sequences were performed was randomized. Two blinded board-certified neuroradiologists evaluated the number of detectable metastatic lesions with each sequence relative to a criterion standard reading conducted at the Gamma Knife facility by a neuroradiologist with access to all clinical and imaging data.
RESULTS: In the volunteer assessment with SPACE and MP-RAGE, SNR (10.3 ± 0.8 vs 7.7 ± 0.7) and contrast-to-noise ratio (0.8 ± 0.2 vs 0.5 ± 0.1) were statistically significantly greater with the SPACE sequence (P < 0.05). Overall, lesion detection was markedly improved with the SPACE sequence (99.1% of lesions for reader 1 and 96.3% of lesions for reader 2) compared with the MP-RAGE sequence (73.6% of lesions for reader 1 and 68.5% of lesions for reader 2; P < 0.01).
CONCLUSIONS: A 3D T1-weighted fast spin echo sequence (SPACE) improves detection of metastatic lesions relative to 3D T1-weighted gradient-echo-based scan (MP-RAGE) imaging when implemented with a 32-channel head coil at identical scan acquisition times (3 minutes and 30 seconds).

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23192164     DOI: 10.1097/RLI.0b013e318277b1aa

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  14 in total

1.  Brain Tumor-Enhancement Visualization and Morphometric Assessment: A Comparison of MPRAGE, SPACE, and VIBE MRI Techniques.

Authors:  L Danieli; G C Riccitelli; D Distefano; E Prodi; E Ventura; A Cianfoni; A Kaelin-Lang; M Reinert; E Pravatà
Journal:  AJNR Am J Neuroradiol       Date:  2019-06-20       Impact factor: 3.825

2.  Accuracy of postcontrast 3D turbo spin-echo MR sequence for the detection of enhanced inflammatory lesions in patients with multiple sclerosis.

Authors:  J Hodel; O Outteryck; E Ryo; A-L Bocher; O Lambert; D Chéchin; H Zéphir; A Lacour; J-P Pruvo; P Vermersch; X Leclerc
Journal:  AJNR Am J Neuroradiol       Date:  2013-11-07       Impact factor: 3.825

Review 3.  Brain metastases: neuroimaging.

Authors:  Whitney B Pope
Journal:  Handb Clin Neurol       Date:  2018

4.  Comparison of lesion enhancement between BB Cube and 3D-SPGR images for brain tumors with 1.5-T magnetic resonance imaging.

Authors:  Hirokazu Hasegawa; Ryuichiro Ashikaga; Kaoru Okajima; Tetsuya Wakayama; Mitsuharu Miyoshi; Yasumasa Nishimura; Takamichi Murakami
Journal:  Jpn J Radiol       Date:  2017-05-24       Impact factor: 2.374

Review 5.  The detectability of brain metastases using contrast-enhanced spin-echo or gradient-echo images: a systematic review and meta-analysis.

Authors:  Chong Hyun Suh; Seung Chai Jung; Kyung Won Kim; Junhee Pyo
Journal:  J Neurooncol       Date:  2016-06-20       Impact factor: 4.130

6.  Consensus recommendations for a standardized brain tumor imaging protocol for clinical trials in brain metastases.

Authors:  Timothy J Kaufmann; Marion Smits; Jerrold Boxerman; Raymond Huang; Daniel P Barboriak; Michael Weller; Caroline Chung; Christina Tsien; Paul D Brown; Lalitha Shankar; Evanthia Galanis; Elizabeth Gerstner; Martin J van den Bent; Terry C Burns; Ian F Parney; Gavin Dunn; Priscilla K Brastianos; Nancy U Lin; Patrick Y Wen; Benjamin M Ellingson
Journal:  Neuro Oncol       Date:  2020-06-09       Impact factor: 12.300

7.  Visualization of acute edema in the left atrial myocardium after radiofrequency ablation: Application of a novel high-resolution 3-dimensional magnetic resonance imaging sequence.

Authors:  Tarek Zghaib; Ashkan A Malayeri; Esra G Ipek; Mohammadali Habibi; Dong Huang; Muhammad A Balouch; David A Bluemke; Hugh Calkins; Saman Nazarian; Stefan L Zimmerman
Journal:  Heart Rhythm       Date:  2018-03-09       Impact factor: 6.343

8.  Comparison of contrast-enhanced modified T1-weighted 3D TSE black-blood and 3D MP-RAGE sequences for the detection of cerebral metastases and brain tumours.

Authors:  N N Kammer; E Coppenrath; K M Treitl; H Kooijman; O Dietrich; T Saam
Journal:  Eur Radiol       Date:  2015-09-03       Impact factor: 5.315

9.  T1 Signal-Intensity Increase in the Dentate Nucleus after Multiple Exposures to Gadodiamide: Intraindividual Comparison between 2 Commonly Used Sequences.

Authors:  J Ramalho; M Ramalho; M AlObaidy; R H Nunes; M Castillo; R C Semelka
Journal:  AJNR Am J Neuroradiol       Date:  2016-03-31       Impact factor: 3.825

10.  3D T1-weighted turbo spin echo contrast-enhanced MRI at 1.5 T for frameless brain metastases radiotherapy.

Authors:  Jing Yuan; Stephen C K Law; Ka Kin Wong; Gladys G Lo; Michael K M Kam; Wing Hong Kwan; Cindy Xue; Oi Lei Wong; Siu Ki Yu; Kin Yin Cheung
Journal:  J Cancer Res Clin Oncol       Date:  2021-08-06       Impact factor: 4.553

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.