In vitro evaluation of the imaging accuracy of C-arm conebeam CT in cerebral perfusion imaging.

A Ganguly1, A Fieselmann, J Boese, C Rohkohl, J Hornegger, R Fahrig.   

Abstract

PURPOSE: The authors have developed a method to enable cerebral perfusion CT imaging using C-arm based conebeam CT (CBCT). This allows intraprocedural monitoring of brain perfusion during treatment of stroke. Briefly, the technique consists of acquiring multiple scans (each scan comprised of six sweeps) acquired at different time delays with respect to the start of the x-ray contrast agent injection. The projections are then reconstructed into angular blocks and interpolated at desired time points. The authors have previously demonstrated its feasibility in vivo using an animal model. In this paper, the authors describe an in vitro technique to evaluate the accuracy of their method for measuring the relevant temporal signals.
METHODS: The authors' evaluation method is based on the concept that any temporal signal can be represented by a Fourier series of weighted sinusoids. A sinusoidal phantom was developed by varying the concentration of iodine as successive steps of a sine wave. Each step corresponding to a different dilution of iodine contrast solution contained in partitions along a cylinder. By translating the phantom along the axis at different velocities, sinusoidal signals at different frequencies were generated. Using their image acquisition and reconstruction algorithm, these sinusoidal signals were imaged with a C-arm system and the 3D volumes were reconstructed. The average value in a slice was plotted as a function of time. The phantom was also imaged using a clinical CT system with 0.5 s rotation. C-arm CBCT results using 6, 3, 2, and 1 scan sequences were compared to those obtained using CT. Data were compared for linear velocities of the phantom ranging from 0.6 to 1 cm∕s. This covers the temporal frequencies up to 0.16 Hz corresponding to a frequency range within which 99% of the spectral energy for all temporal signals in cerebral perfusion imaging is contained.
RESULTS: The errors in measurement of temporal frequencies are mostly below 2% for all multiscan sequences. For single scan sequences, the errors increase sharply beyond 0.10 Hz. The amplitude errors increase with frequency and with decrease in the number of scans used.
CONCLUSIONS: Our multiscan perfusion CT approach allows low errors in signal frequency measurement. Increasing the number of scans reduces the amplitude errors. A two-scan sequence appears to offer the best compromise between accuracy and the associated total x-ray and iodine dose.

Entities:  

Mesh:

Year:  2012        PMID: 23127059      PMCID: PMC3482259          DOI: 10.1118/1.4757910

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  16 in total

Review 1.  New and future endovascular treatment strategies for acute ischemic stroke.

Authors:  Gary M Nesbit; George Luh; Raymond Tien; Stanley L Barnwell
Journal:  J Vasc Interv Radiol       Date:  2004-01       Impact factor: 3.464

2.  Measurement of tissue perfusion by dynamic computed tomography.

Authors:  K A Miles
Journal:  Br J Radiol       Date:  1991-05       Impact factor: 3.039

3.  Colour perfusion imaging: a new application of computed tomography.

Authors:  K A Miles; M Hayball; A K Dixon
Journal:  Lancet       Date:  1991-03-16       Impact factor: 79.321

4.  Impact of operator-selected image noise index and reconstruction slice thickness on patient radiation dose in 64-MDCT.

Authors:  Kalpana M Kanal; Brent K Stewart; Orpheus Kolokythas; William P Shuman
Journal:  AJR Am J Roentgenol       Date:  2007-07       Impact factor: 3.959

5.  A temporal interpolation approach for dynamic reconstruction in perfusion CT.

Authors:  Pau Montes; Günter Lauritsch
Journal:  Med Phys       Date:  2007-07       Impact factor: 4.071

6.  Interventional 4-D C-arm CT perfusion imaging using interleaved scanning and partial reconstruction interpolation.

Authors:  Andreas Fieselmann; Arundhuti Ganguly; Yu Deuerling-Zheng; Michael Zellerhoff; Christopher Rohkohl; Jan Boese; Joachim Hornegger; Rebecca Fahrig
Journal:  IEEE Trans Med Imaging       Date:  2011-12-23       Impact factor: 10.048

7.  On the physical and stochastic representation of an indicator dilution curve as a gamma variate.

Authors:  M Mischi; J A den Boer; H H M Korsten
Journal:  Physiol Meas       Date:  2008-02-11       Impact factor: 2.833

8.  High resolution measurement of cerebral blood flow using intravascular tracer bolus passages. Part I: Mathematical approach and statistical analysis.

Authors:  L Ostergaard; R M Weisskoff; D A Chesler; C Gyldensted; B R Rosen
Journal:  Magn Reson Med       Date:  1996-11       Impact factor: 4.668

9.  Pre- and in-hospital delays from stroke onset to intra-arterial thrombolysis.

Authors:  Krassen Nedeltchev; Marcel Arnold; Caspar Brekenfeld; Jörg Isenegger; Luca Remonda; Gerhard Schroth; Heinrich P Mattle
Journal:  Stroke       Date:  2003-04-17       Impact factor: 7.914

10.  Guidelines for the early management of adults with ischemic stroke: a guideline from the American Heart Association/American Stroke Association Stroke Council, Clinical Cardiology Council, Cardiovascular Radiology and Intervention Council, and the Atherosclerotic Peripheral Vascular Disease and Quality of Care Outcomes in Research Interdisciplinary Working Groups: the American Academy of Neurology affirms the value of this guideline as an educational tool for neurologists.

Authors:  Harold P Adams; Gregory del Zoppo; Mark J Alberts; Deepak L Bhatt; Lawrence Brass; Anthony Furlan; Robert L Grubb; Randall T Higashida; Edward C Jauch; Chelsea Kidwell; Patrick D Lyden; Lewis B Morgenstern; Adnan I Qureshi; Robert H Rosenwasser; Phillip A Scott; Eelco F M Wijdicks
Journal:  Stroke       Date:  2007-04-12       Impact factor: 7.914

View more
  1 in total

1.  Endovascular Recanalization in Acute Ischemic Stroke Using the Solitaire FR Revascularization Device with Adjunctive C-Arm CT Imaging.

Authors:  B D Mitchell; P Chinnadurai; G Chintalapani; H A Morsi; H Shaltoni; M E Mawad
Journal:  AJNR Am J Neuroradiol       Date:  2015-06-04       Impact factor: 3.825

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.