Evaluation of hemoglobin of blood donors deferred by the copper sulphate method for hemoglobin estimation.

R B Sawant1, Z S Bharucha, S B Rajadhyaksha.   

Abstract

INTRODUCTION: The copper sulphate (CuSO4) specific gravity test for Hb screening tends to give inappropriate failures. This prompted us to compare it with alternate screening methods. AIM: To study the impact of inaccuracy of CuSO4 method on donor deferral.
METHODS: Capillary and venous blood samples of 400 potential blood donors failing the primary Hb screening using appropriately standardized CuSO4 test (specific gravity 1.053) were tested by Hemocue photometer, the Hb colour scale, Cyanmethemoglobin method as well as the automated hematology analyser, which was considered as the standard reference method.
RESULTS: One hundred and sixteen donors (29%) who failed the CuSO4 test had true Hb levels >12.5 g/dl. The Hb levels of 131 (32.8%) deferred donors were between 12 and 12.5 g/dl. The sensitivity of Hemocue, Hb colour scale and Cyanmethemoglobin was 99%, 97% and 96% and their specificity was 45%, 93% and 46%, respectively. The positive predictive values (PPV) of Hemocue and Cyanmethemoglobin methods were low (43% and 44%, respectively) but their negative predictive values (NPV) were high (99%, and 97%, respectively). The Hb colour scale had an overall best performance with a PPV of 96% and NPV of 95%.
CONCLUSION: The Hemoglobin colour scale which is inexpensive, convenient for field testing and has the overall best performance, is the most suitable for donor Hb screening. Since its readability is 12 g/dl, lowering the donor Hb threshold to 12 g/dl should be actively considered.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17382593     DOI: 10.1016/j.transci.2006.11.001

Source DB:  PubMed          Journal:  Transfus Apher Sci        ISSN: 1473-0502            Impact factor:   1.764


  11 in total

1.  Validation of the haemoglobin colour scale for screening blood donors in Malawi.

Authors:  B M'baya; I Mbingwani; L Mgawi; V Mkochi; I Bates; S A White; T J Allain
Journal:  Malawi Med J       Date:  2014-06       Impact factor: 0.875

2.  Evaluation of the validity of a rapid method for measuring high and low haemoglobin levels in whole blood donors.

Authors:  Hayedeh J Shahshahani; Nahid Meraat; Fatemeh Mansouri
Journal:  Blood Transfus       Date:  2012-10-10       Impact factor: 3.443

3.  Evaluation of capillary haemoglobin determination for anaemia screening in blood donation settings.

Authors:  Massimo Daves; Roberto Cemin; Elmar M Zagler; Alexandra Joos; Stefan Platzgummer; Rudolf Hueber; Giuseppe Lippi
Journal:  Blood Transfus       Date:  2016-01-19       Impact factor: 3.443

Review 4.  Methods and analyzers for hemoglobin measurement in clinical laboratories and field settings.

Authors:  Ralph D Whitehead; Zuguo Mei; Carine Mapango; Maria Elena D Jefferds
Journal:  Ann N Y Acad Sci       Date:  2019-06-04       Impact factor: 5.691

5.  Evaluation of a portable hemoglobin photometer in pregnant women in a high altitude area: a pilot study.

Authors:  Xiaoyan Zhou; Hong Yan; Yuan Xing; Shaonong Dang; Bianba Zhuoma; Duolao Wang
Journal:  BMC Public Health       Date:  2009-07-11       Impact factor: 3.295

6.  The Sting of Rejection: Deferring Blood Donors due to Low Hemoglobin Values Reduces Future Returns.

Authors:  Adrian Bruhin; Lorenz Goette; Simon Haenni; Lingqing Jiang; Alexander Markovic; Adrian Roethlisberger; Regula Buchli; Beat M Frey
Journal:  Transfus Med Hemother       Date:  2019-05-27       Impact factor: 3.747

7.  Comparison of the accuracy of noninvasive hemoglobin sensor (NBM-200) and portable hemoglobinometer (HemoCue) with an automated hematology analyzer (LH500) in blood donor screening.

Authors:  Moon Jung Kim; Quehn Park; Myung Hee Kim; Jeong Won Shin; Hyun Ok Kim
Journal:  Ann Lab Med       Date:  2013-06-24       Impact factor: 3.464

8.  Hemoglobin estimation by the HemoCue® portable hemoglobin photometer in a resource poor setting.

Authors:  Bernard Nkrumah; Samuel Blay Nguah; Nimako Sarpong; Denise Dekker; Ali Idriss; Juergen May; Yaw Adu-Sarkodie
Journal:  BMC Clin Pathol       Date:  2011-04-21

Review 9.  Techniques used for the screening of hemoglobin levels in blood donors: current insights and future directions.

Authors:  Rajendra Chaudhary; Anju Dubey; Atul Sonker
Journal:  J Blood Med       Date:  2017-07-03

10.  Deferral pattern in voluntary blood donors on basis of low hemoglobin and effect of application of digital hemoglobinometer on this pattern.

Authors:  Ankit Mathur; Ripal Shah; Priti Shah; V Harimoorthy; Nabajyoti Choudhury
Journal:  Asian J Transfus Sci       Date:  2012-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.