Validation of a new generation POCT glucose device with emphasis on aspects important for glycemic control in the hospital care.

Snježana Kos1, Arie van Meerkerk, Joke van der Linden, Theo Stiphout, Remi Wulkan.   

Abstract

BACKGROUND: Point-of-care (POC) glucose devices are widely used for insulin-dosage decision-making although such an application is not always permitted. In this study, we have evaluated a new generation of POC glucose device, the HemoCue(®) Glucose 201DMRT (201DMRT), for its suitability for (tight) glycemic control.
MATERIALS AND METHODS: This study was performed according to the CLSI/STARD criteria. The 201DMRT was compared to the laboratory hexokinase glucose method (Siemens Dimension Vista(®)). The variation among different POC devices and cuvette lot numbers was examined. Additionally, the influence of the partial pressure of oxygen and hematocrit on glucose measurement was investigated.
RESULTS: The 201DMRT showed a good agreement with the laboratory reference method. This was examined using Deming regression analysis, percentage Bland-Altman plot and a modified Clarke-error grid. The total analytical error at the clinically critical glucose concentrations of 5.6, 7.0 and 11.1 mmol/L (101, 126 and 200 mg/dL) was 6.4%, 4.3% and 3.0%, respectively. The total error among the different POC devices and among different cuvette lot numbers was <6.5%. Glucose measurements on the 201DMRT were not affected by changes in partial pressure of oxygen, whereas changes in hematocrit had influence on the results (3.4% for every 0.10 L/L change in hematocrit).
CONCLUSIONS: The 201DMRT device can be used for glycemic control based on analytical results presented. However, the clinical applicability for tight glycemic control must be confirmed in a clinical study.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23092803     DOI: 10.1515/cclm-2011-0900

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  11 in total

1.  Assessing the Accuracy of Continuous Glucose Monitoring (CGM) Calibrated With Capillary Values Using Capillary or Venous Glucose Levels as a Reference.

Authors:  Mervi Andelin; Jort Kropff; Viktorija Matuleviciene; Jeffrey I Joseph; Stig Attvall; Elvar Theodorsson; Irl B Hirsch; Henrik Imberg; Sofia Dahlqvist; David Klonoff; Börje Haraldsson; J Hans DeVries; Marcus Lind
Journal:  J Diabetes Sci Technol       Date:  2016-06-28

Review 2.  Challenges of inpatient blood glucose monitoring: standards, methods, and devices to measure blood glucose.

Authors:  Kathie L Hermayer; Aundrea S Loftley; Sumana Reddy; Satya Nandana Narla; Nina A Epps; Yusheng Zhu
Journal:  Curr Diab Rep       Date:  2015-03       Impact factor: 4.810

3.  Evaluation of Reference Metrics for Continuous Glucose Monitoring in Persons Without Diabetes and Prediabetes.

Authors:  Sheyda Sofizadeh; Anders Pehrsson; Arndís F Ólafsdóttir; Marcus Lind
Journal:  J Diabetes Sci Technol       Date:  2020-10-26

4.  A Clinical Trial of the Accuracy and Treatment Experience of the Flash Glucose Monitor FreeStyle Libre in Adults with Type 1 Diabetes.

Authors:  Arndís F Ólafsdóttir; Stig Attvall; Ulrika Sandgren; Sofia Dahlqvist; Aldina Pivodic; Stanko Skrtic; Elvar Theodorsson; Marcus Lind
Journal:  Diabetes Technol Ther       Date:  2017-03-06       Impact factor: 6.118

5.  Effect of a Whey Protein Supplement on Preservation of Fat Free Mass in Overweight and Obese Individuals on an Energy Restricted Very Low Caloric Diet.

Authors:  Anne Ellegaard Larsen; Bo Martin Bibby; Mette Hansen
Journal:  Nutrients       Date:  2018-12-04       Impact factor: 5.717

6.  Reliability and criterion validity of self-measured waist, hip, and neck circumferences.

Authors:  Pamela Barrios; Jennifer Martin-Biggers; Virginia Quick; Carol Byrd-Bredbenner
Journal:  BMC Med Res Methodol       Date:  2016-05-04       Impact factor: 4.615

7.  Overweight and abdominal obesity as determinants of undiagnosed diabetes and pre-diabetes in Bangladesh.

Authors:  Dewan S Alam; Shamim H Talukder; Muhammad Ashique Haider Chowdhury; Ali Tanweer Siddiquee; Shyfuddin Ahmed; Sonia Pervin; Sushmita Khan; Khaled Hasan; Tracey L P Koehlmoos; Louis W Niessen
Journal:  BMC Obes       Date:  2016-03-18

8.  Critical-depth Raman spectroscopy enables home-use non-invasive glucose monitoring.

Authors:  Signe M Lundsgaard-Nielsen; Anders Pors; Stefan O Banke; Jan E Henriksen; Dietrich K Hepp; Anders Weber
Journal:  PLoS One       Date:  2018-05-11       Impact factor: 3.240

9.  Diagnosis of Gestational Diabetes Mellitus with Point-of-Care Methods for Glucose versus Hospital Laboratory Method Using Isotope Dilution Gas Chromatography-Mass Spectrometry as Reference.

Authors:  Karl Kristensen; Anne-Marie Wangel; Anastasia Katsarou; Nael Shaat; David Simmons; Helena Fadl; Kerstin Berntorp
Journal:  J Diabetes Res       Date:  2020-03-20       Impact factor: 4.011

10.  Standing up against office sitting: A study protocol.

Authors:  Philippe Gradidge; Merling Phaswana; Katrien Wijndaele; Nigel Crowther; Catherine Draper
Journal:  S Afr J Physiother       Date:  2020-09-04
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.