We present an adaptive optics spectral domain optical coherence tomography (AO-SDOCT) with a long focal range by active phase modulation of the pupil. A long focal range is achieved by introducing AO-controlled third-order spherical aberration (SA). The property of SA and its effects on focal range are investigated in detail using the Huygens-Fresnel principle, beam profile measurement and OCT imaging of a phantom. The results indicate that the focal range is extended by applying SA, and the direction of extension can be controlled by the sign of applied SA. Finally, we demonstrated in vivo human retinal imaging by altering the applied SA.
We present an adaptive optics spectral domain optical coherence tomography (AO-SDOCT) with a long focal range by active phase modulation of the pupil. A long focal range is achieved by introducing AO-controlled third-order spherical aberration (SA). The property of SA and its effects on focal range are investigated in detail using the Huygens-Fresnel principle, beam profile measurement and OCT imaging of a phantom. The results indicate that the focal range is extended by applying SA, and the direction of extension can be controlled by the sign of applied SA. Finally, we demonstrated in vivo human retinal imaging by altering the applied SA.
Entities:
Keywords:
(110.1080) Active or adaptive optics; (170.4470) Ophthalmology; (170.4500) Optical coherence tomography
Optical coherence tomography (OCT) provides cross-sectional imaging of the internal
microstructures of biological tissues [1]. It is a
powerful imaging technique, especially in ophthalmology [2-6]. Among several implementations of OCT,
spectral domain OCT (SDOCT) provides high resolution, high sensitivity and high speed imaging
[7-10]. Such
implementations enable visualization of the microstructures of the human eye by non-invasive and
non-contact measurement. High sensitivity of OCT detection enables imaging of structures with
very low optical scattering, such as the retina. The high axial and lateral resolution enables
the precise morphological investigation of retinal pathologies. Axial resolution is proportional
to with being the center wavelength and being the spectral bandwidth of the light source. However, lateral
resolution is linked to the focusing property of the illumination on the sample. The lateral
resolution at the focal plane is proportional to the inverse of numerical aperture (NA) if there
is no aberration.For high-resolution retinal imaging, the lateral resolution can be improved by using higher NA
illumination, which is typically realized by enlarging the probe beam diameter on the cornea.
However, the aberrations of the anterior part of the eye increase with the incident beam
diameter. Ocular aberration deteriorates lateral resolution and leads to reduced image quality
[11,12], so it is
not possible to achieve diffraction limited resolution of a few micrometers without correction
of ocular aberrations. Therefore, correction of ocular aberration is essential for cellular
level retinal imaging.Adaptive optics (AO) technology has been used to solve this problem. Ocular aberration can be
corrected dynamically by using AO [13-16]. AO has been combined with several type of ophthalmic
imaging modalities such as a flood-illumination camera [13,15,17,18], a scanning laser ophthalmoscope (SLO)
[19-21], and an
OCT [22-27]. By
using these techniques, in vivo cellular level investigation of retinal
disorders and functions has also been demonstrated [28,29]. Thus, these AO retinal imaging devices
have provided improved lateral resolution.However, a larger probe beam diameter leads to rapid degradation of lateral resolution in the
axial direction because the depth of focus (DOF) shortens with the square of NA. There is a
trade-off between lateral resolution and DOF, and a short DOF is problematic for in
vivo AO-OCT imaging of human retina. For instance, there is temporal wavefront
instability in the form of micro-fluctuations of accommodation [16,30], tear films, a point scanning system
aberration, involuntary eye motion, and some other factors. Regardless of AO dynamic aberration
correction, the temporal fluctuations remain due to the limited closed-loop bandwidth up to 1–2
Hz [16]. These variations make it difficult to maintain
perfectly in-focus imaging conditions during measurement and deteriorate retinal image quality.
Hence, AO retinal imaging necessitates the precise defocus adjustment by monitoring the
en face projections or optimized post-processing. In addition, especially for
the OCT retinal imaging, the retina and choroid have thick multi-layered structures with a
thickness of around 0.6 mm, and hence a long focal range is required. In practice, the DOF of
typical AO system is less than 100 µm, this is too narrow to image all of retinal and choroidal
layers. Hence, it is important to achieve a long focal range while preserving high lateral
resolution.The extended DOF while preserving the lateral resolution and energy efficiency have been
studied in the fields of vision and optical imaging [31].
For instance, in the field of vision, the extended DOF is a common topic for the correction of
presbyopia. Specifically, the multifocal lenses have been used to have a long focal range.
Although some technical or physiological issues are still remained, the improvement of near
vision to monofocal lenses has been observed [31-34]. On the other hands, in the field of microscopy,
three-dimensional scanning method and the point spread function engineering by using wavefront
coding [35], axicon lenses [36], and annular illumination [37] have
been proposed to extend the DOF more efficiently. Similarly, in the field of OCT, dynamic
focusing during T-scan OCT detection has been presented and has successfully shown retinal
microstructures [38]. Another approach is the Bessel
illumination method which uses an axicon lens [39,40]. Although the image quality of this method is excellent,
Bessel illumination has low light efficiency, which makes it inappropriate for retinal
imaging.It is noteworthy that the axicon can be considered to generate extremely high-order spherical
aberration (SA) [41]. Several authors have shown the
effect of extension of DOF by SA in the field of laser applications [42-44]. And hence, we hypothesize that
the SA of standard orders would extend the DOF of AO-OCT imaging.In this paper, we demonstrate high-resolution volumetric imaging of retinal microstructures by
means of extended DOF, which is achieved by introducing a controlled SA. A custom-built AO-SDOCT
is utilized for the imaging. In this system, eye aberration is dynamically corrected by AO, and
additional controlled SA is introduced also by AO in order to achieve the extended DOF. Namely,
by using AO, we create phase distribution of SA on the pupil plane in addition to cancelation of
ocular aberration. The imaging properties of this method are also investigated in detail,
applying the Huygens-Fresnel principle, beam profile measurement, and OCT imaging of a
phantom.
2. System
The schematic diagram of AO-SDOCT is shown in Fig. 1
. The system is based on our previously published custom-built AO-SDOCT [27]. This AO-SDOCT consists of two subsystems of SDOCT and AO.
The details of these subsystems are described in the following sections.
Fig. 1
The schematic diagram of the optical setup of AO-SDOCT. PC: Polarization controller, FC:
Fiber coupler, (a) The optical setup of the AO retinal scanner, L#: Lenses, LP#: Linear
polarizers, D: Dichroic mirror, Ach: Achromatizer, ST: Stop, SM#: Spherical mirrors, FM#:
Flat mirrors, WS: Wavefront sensor, DM: Deformable mirror, VS: Vertical galvanometric
scanner, HS: Horizontal galvanometric scanner. (b) Spectrometer. (c) Reference arm, ND:
neutral density filter. The green bars and arrows indicate the optical conjugate planes of
the pupil, and the red crosses and arrow indicate the optical conjugate planes of the
retina.
The schematic diagram of the optical setup of AO-SDOCT. PC: Polarization controller, FC:
Fiber coupler, (a) The optical setup of the AO retinal scanner, L#: Lenses, LP#: Linear
polarizers, D: Dichroic mirror, Ach: Achromatizer, ST: Stop, SM#: Spherical mirrors, FM#:
Flat mirrors, WS: Wavefront sensor, DM: Deformable mirror, VS: Vertical galvanometric
scanner, HS: Horizontal galvanometric scanner. (b) Spectrometer. (c) Reference arm, ND:
neutral density filter. The green bars and arrows indicate the optical conjugate planes of
the pupil, and the red crosses and arrow indicate the optical conjugate planes of the
retina.
2.1. Spectral domain optical coherence tomography
The SDOCT subsystem uses an SLD light source (Superlum diode, Ireland) with 1.02-µm center
wavelength and a spectral bandwidth (FWHM) of 106 nm. The theoretical axial resolution is 4.7
µm in air and 3.4 µm in tissue. Light from the source is divided and delivered to a sample arm
and a reference arm by a 50/50 fiber coupler. In the sample arm, light is delivered to the
retina via an AO retinal scanner. Backscattered light from the retina is recoupled into the
fiber and sent back to the fiber coupler, where the backscattered sample beam is recombined
with the light from the reference arm. The recombined light is delivered to the spectrometer,
in which the light is collimated by an achromatic lens (AC254-050-B, Thorlabs Inc., NJ),
diffracted by a reflective grating (GR50-1210, 1200 lp/mm, Thorlabs) and focused by a lens pair
(AC508-050-B and AC508-150-B, Thorlabs) on an InGaAs 1024-pixel line-scan camera driven at a
line rate of 91,911 lines/s (SUI1024LDH, Sensors Unlimited Inc., NJ).The sensitivity of the system was measured to be 81.5 dB with a sample power of 1.26 mW.
Sensitivity was measured with a cat's eye consisting of an achromatic doublet (AC254-060-B,
Thorlabs) and a mirror at an optical conjugated plane of the retina. Note that during this
measurement the deformable mirror of the AO retinal scanner was flattened and system aberration
was not corrected. Because we put the mirror (reflector) as a sample, it is impossible to
correct the system aberrations with this double pass configuration. This could result in
underestimation of the sensitivity. Alternatively, we also measured the sensitivity with
aberration correction by an indirect method as follows. First, the system aberration was
corrected with a model eye using a paper (scattering target) as a retina. Second, we replace
the paper to a mirror (reflector) and realigned it. The defocus and tilts were adjusted to have
the maximum re-coupling efficiency to the optical fiber of the interferometer. Finally, the
sensitivity was measured. As a result, the sensitivity with aberration correction was 85.9 dB,
which is 4.4 dB larger than the sensitivity without the aberration correction.The optical design of the system ensures a beam diameter of 7.4 mm on the cornea, and it
results in a transform limited 1/e2 beam width of 2.9 µm on the
retina. The DOF, which is defined by 2-times Rayleigh length, was 48 µm.
2.2. Adaptive optics
2.2.1. Adaptive optics retinal scanner
The AO subsystem is based on our previously built AO retinal scanner [45]. In this paper we used it for canceling ocular aberration and creating
SA. First, a large defocus was corrected by a Badal optometer which is placed before the eye
pupil (L6 and L7). Then, the residual aberrations are measured by a Shack-Hartmann wavefront
sensor (HASO32, Imagine Eyes, Orsay, France) using a laser diode with a 700-nm center
wavelength as a beacon. A magnetic deformable mirror (Mirao52, Imagine Eyes) corrects the
measured aberrations. An eye pupil, a deformable mirror, a wavefront sensor and two scanners
are optically conjugated to each other by the off-axis reflective telescopes and Badal
optometer.To cancel longitudinal chromatic aberrations of human eye, a custom-designed broadband
achromatizer is used. The general concept of an achromatizer is described by Fernández
et al. [46]. With this achromatizer,
the axial chromatic focal shift was controlled to be less than 3 µm in the wavelength band of
970 to 1070 nm. The achromatizer is located between the input collimator and a dichroic mirror
which splits a back-scattered sample beam to the fiber tip and to a wavefront sensor. This
configuration prevents surface reflections from the achromatizer.The AO control software was implemented using LabVIEW SDK (Imagine Eyes). The influence
matrix is obtained by placing a single mode fiber at the optical conjugate plane of the
retina. The iteration frequency of the AO closed loop was around 10 Hz.The retina is scanned with a vertical-fast raster scanning protocol using two galvanometric
scanners in which the slow scan is in the horizontal plane and the fast scan is in the
vertical plane. The optical loss of the AO retinal scanner is 2.8 dB for a single pass.
2.2.2. Adaptive optics-control with arbitrary target aberration
For dynamic ocular aberration correction, a deformable mirror applies counter aberration to
the illumination wavefront based on the measured wavefront. The closed loop control algorithm
is the vital link between these two components; the measured wavefront and the wavefront
created by the deformable mirror. In AO closed loop operation, integral controller is applied.
A command vector, a set of control voltages to the electrodes of the deformable mirror, at the
time point of t + T was obtained as [47]where t is time, T is a time period of a single cycle of the AO closed loop, and
is a command vector of the one-previous cycle.
is a slope vector obtained by the Shack-Hartmann wavefront
sensor and A is the Moore-Penrose pseudo inverse matrix of the influence matrix
of the AO-setup A, and α represent a closed-loop gain. To add an
arbitrary aberration to the target of AO closed-loop control, Eq. (1) is extended and we subtract the target slope vector
from the measured slope vector,where is a target slope vector defined from the target aberration
as . is a vector of the Zernike coefficients of the target aberration
and Z is a conversion matrix from the Zernike coefficient vector to a slope
vector. In this paper, we used root mean square (RMS) value to represent the Zernike
coefficients, which are normalized by a pupil diameter, 7.4 mm.Since our AO control algorithm defines the command vector by Eq. (2), the wavefront at the pupil converges to the target aberration by the
closed-loop operation. In this study, we set a proper amount of SA as the target.
3. Pseudo focal shift by introducing spherical aberration
3.1. Theory
In this study, we introduced SA to extend the focal range of AO-SDOCT. In the presence of SA,
a paraxial focal point and a marginal focal point are no longer identical, i.e., SA generates
longitudinal aberration (LA). This LA results in not only the axial extension of the focus but
also the shifting of the pseudo focal plane as discussed below. LA is given using the following
equation [48]:where f/# is the f-number, and is the third-order SA coefficient of Seidel aberration.
Conversion from Zernike to Seidel representation is given by [49]where is the third-order SA coefficient of Zernike aberration.In addition to the focus extension, the LA also shifts the depth position of the focus.
Namely, under the presence of LA, the highest lateral resolution is no longer obtained at the
paraxial focal plane. According to geometrical optics, the depth location of a minimum circle,
where the marginal ray crosses a caustic, is regarded as one of the best image plane. The
minimum circle position is located at the position distant from the paraxial focus point for
0.75 LA, in the direction of the marginal focus point [48]. However, in a numerical simulation based on the Huygens-Fresnel principle in
Zemax, we found that the minimum circle was no longer located at 0.75 LA, and it was located at
0.25 LA.Here, we defined positive and negative SAs as follows. If the marginal rays are more highly
refracted than the paraxial rays, we denote it positive SA, and vice versa. Thus, positive SA
shifts the depth position of the minimum circle to get closer to the lens, while negative SA
shifts the minimum circle position to get away from the lens. In this study, an additional
defocus is added to the target aberration in order to cancel this pseudo focal shift incurred
by the shift of the minimum circle. In this paper, we call this additional defocus “counter
defocus.”Aside from this counter defocus, we have to apply the defocus to cancel a longitudinal
chromatic aberration (LCA) between the 1-µm probe and 700-nm beacon beam. In practice, we
canceled the LCA by adjusting the distance between the relay lens in front of the wavefront
sensor (L4 and L5). However, this cancellation could not be perfect because of practical
implementation issues. The amount of defocus needed to cancel the residual LCA is about –0.2
µm. In this study, we applied the sum of the counter defocus and the defocus to cancel the
residual LCA.
3.2. Estimation of counter defocus
The counter defocus was estimated and the relevance of the estimation was confirmed by an
experiment.We set arbitrarily and the corresponding pseudo focal shift was obtained
by the numerical simulation using Zemax (Radiant Zemax, WA). Subsequently, the theoretical
prediction of the counter defocus was calculated by the following equation [49]:where is the pseudo focal shift from the focal plane,
c is dioptric power, n is the refractive index,
R is pupil radius, and is the Zernike coefficient of defocus. Here,
corresponds to 0.25 LA.In practical estimation, a lens with a 16-mm focal length was assumed as a model of a human
eye. A beam radius of 3.7 mm is assumed. The SA under consideration ranged from –1.0 µm to 1.0
µm with a 0.1-µm step.
3.3. Experimental validation
The experiment was done by using a model eye, comprised of a lens with a focal length of 16
mm and a paper target which mimicked the refractive optics of the eye (cornea and crystalline
lens) and retina. In this experiment, the counter defocus was defined as the defocus which
maximized the OCT intensity of the paper of the model eye. Then, the counter defocus was
estimated for each SA by controlling the target wavefront in the AO closed loop.Figure 2
shows the counter defocus as a function of SA. It should be noted that the defocus
offset incurred by LCA has been corrected in this plot.
Fig. 2
The counter defocus of ray tracing simulation and experiment are plotted as a function of
Zernike coefficient of third-order SA where + is for experiment, × is for simulation-1, and
* is for simulation-2.
The counter defocus of ray tracing simulation and experiment are plotted as a function of
Zernike coefficient of third-order SA where + is for experiment, × is for simulation-1, and
* is for simulation-2.The results of simulation-1 were obtained by the numerical simulation and indicate the amount
of counter defocus required to shift the position of the minimum circle to the focal plane,
which corresponds to the distance of 0.25 LA. Contrastingly, simulation-2 is the amount of
defocus required to shift the paraxial focus position to the focal plane.This experiment and the two simulation results indicate that the actual best image plane is
located between the paraxial focus position and the position of the minimum circle. A small
discrepancy between simulation-1 and the experiment can be considered to come from a difference
between the actual beam intensity distribution on the pupil plane and that assumed in the
simulation. Here, we assumed a Gaussian distribution.
4. System performance
Theoretically, the peak signals will decrease as a function of applied SA. At the same time,
residual RMS wavefront error will increase because of an inevitable property of our AO-control
algorithm; large target aberration results in unavoidable large RMS error (see Appendix A).In order to evaluate the performance, we measured the maximum OCT signals and the RMS
wavefront error as altering the applied SA. The counter defocus and LCA cancelation was done
based on the strategy described in Section 3. The same model eye used in the Subsection 3.3 was
employed, and the backscattered light from the thin paper target was recorded by OCT. The
maximum OCT signal was defined as the average of maximum signals of 100 A-lines. The residual
RMS wavefront errors were recorded by the wavefront sensor. The SA under consideration ranged
from –1 µm to +1 µm with 0.1-µm step.The maximum OCT signals are plotted as a function of applied SA, as shown in Fig. 3(a)
. The maximum OCT signal was found at SA of around 0.0 to 0.1 µm. The possible reason of
the tiny positive shift would be the system induced spherical aberration originated from the
detection path and/or wavelength differences between the probe and beacon beams.
Fig. 3
The maximum OCT singals for each SA is shown in (a). The RMS wavefront errors for each SA
is shown in (b). The‘+ and × indicate the experimental and simulation results,
respectively.
The maximum OCT singals for each SA is shown in (a). The RMS wavefront errors for each SA
is shown in (b). The‘+ and × indicate the experimental and simulation results,
respectively.The residual RMS error was increased as a function of the amount of applied SA, as shown in
Fig. 3(b). Numerical simulation results showed the
similar tendency with the experimental results. The discrepancy between the experiment and
simulation would be explained by measurement error which did not take into account in the
simulation (see Appendix A).
5. Extension of depth-of-focus by introducing controlled aberration
In order to quantitatively investigate the effect of third-order SA on extension of focal
range, we evaluated the lateral resolution along the depth. The lateral resolution and DOF were
evaluated based on the depth-resolved point spread function (PSF) obtained by a numerical
simulation and beam profiler measurement in an experiment.
5.1. Methods
A scanning slit optical beam profiler (BP104-IR: Thorlabs) was employed in this study.
However, the slit size is 2.5 µm, and the minimum measureable diameter for this beam profiler
is 10 µm, which is significantly larger than the transform limited spot size of our standard
model eye, 6.6 µm. Hence, we utilized another eye model with a lens with 30-mm focal length
(AC127-030-B: Thorlabs) as an alternative. This model eye was also utilized for the ray tracing
and experiment.We numerically simulated the PSFs for SA = +0.7, 0.0, and –0.7 µm by using Zemax. The counter
defocus was defined by Eq. (5) in Section 3. We
calculated the theoretical depth-resolved PSF at each depth over ±1000 µm, where the
beam diameter reaches a minimum at the depth of 0 µm. PSF was calculated based on the
Huygens-Fresnel principle in Zemax. We set the pupil and image sampling sizes at 256 pix × 256
pix. The lateral spacing of the simulation was set at 0.5 µm, and the result was binned to have
the same resolution with the experiment of 1.2 µm. We calculated the incoherent sum of the PSF
at 5 different wavelengths of 967 nm, 993 nm, 1020 nm, 1046 nm and 1073 nm which were adopted
as representative wavelengths of the broad band light source of a probe beam. The PSFs were not
normalized. The lateral resolution of the system was defined by
1/e2 beam width. The DOF was defined by the FWHM of central peak
intensity profile along the depth.In the experiment, we measured the intensity profile by the beam profiler at each depth. We
set a target resolution at 1.2 µm. Note that intensity profile represents the integrated
profile over the entire slit, and it is different from the simulation. After the acquisition,
the measured intensity profiles were normalized at each depth location by the maximum
intensity. This is because there was a significant intensity fluctuation by a rotational slit
profiler at each depth. Then, we plotted the entire profiles for SA = +0.7, 0.0, and –0.7 µm.
The measurement range was ±1000 µm with a 20-µm step, and the 0-µm depth was set so
that the beam diameter reaches a minimum at this depth. The counter defocus obtained by another
experiment similar to Section 3 was also added for this measurement. Note that, although the
counter defocus utilized in the simulation and the experiment are not perfectly identical due
to the difference of intensity distribution on the pupil plane as mentioned in Section 3, this
only results in a difference in the depth-offset and has no significant effect on the results.
The lateral resolution was defined by beam width. The variation of the lateral resolution as a
function of the depth was then fitted to the following function [50] which is parameterized by
, and :where is the minimum lateral resolution, z is
depth position, is depth position with the minimum lateral resolution, and
is the Rayleigh length. is used as a parameter to assess the focal range.
5.2. Results
The depth-resolved PSFs for SA = +0.7, 0.0, and –0.7 µm obtained by Zemax are shown in Fig. 4
. The profile of PSF of the case of SA = 0.0 µm (without SA) was almost symmetric with
respect to the focal plane. However, the depth profile of the PSF becomes asymmetric when SA
was applied. When positive SA was applied, the transversal width of PSF was smaller at closer
to the lens than farer from the lens, although some side peaks appeared. When negative SA was
applied, the transversal width of PSF became smaller at farer from the lens than closer to the
lens, and at the same time side peaks appeared. The DOFs are 515 µm, 125 µm, and 489 µm for SA
= +0.7, 0.0, and –0.7 µm, respectively. The DOF was extended 4 times with SAs. The
1/e2 beam widths are 12.6 µm, 7.66 µm, and 12.4 µm for SA = +0.7,
0.0, and –0.7 µm, respectively. The lateral resolution becomes worse 1.6 times with SAs.
Fig. 4
Theoretical PSFs for SA = +0.7, 0.0, and –0.7 µm. The depth positions of LA are indicated
by blue-letters.
Theoretical PSFs for SA = +0.7, 0.0, and –0.7 µm. The depth positions of LA are indicated
by blue-letters.If we simply reduce the beam diameter to have an m-times longer DOF, we
should shrink the beam diameter in times. And hence, the lateral resolution is reduced in
m times worse. Here, to have the same 4 times longer DOF, the lateral
resolution should decrease in 2 times. It means that the lateral resolution is 1.25 times
better with SA with the same DOF.The measured intensity profiles for SA = +0.7, 0.0, and –0.7 µm are shown in Fig. 5
. For all SA values, the intensity profile was deteriorated and transversal width became
larger as a function of depth with respect to the focal plane. The intensity profiles were not
cylindrically symmetric although applied SA and defocus were symmetric. The possible reasons
for this asymmetry are residual aberrations and the heterogeneity of the Gaussian distribution
of the beam intensity on the pupil plane. Note that the abrupt translation of the beam
indicated by the red arrows comes from the instability of the rotational slits of the beam
profiler. Hence it is an artifact and can be safely ignored.
Fig. 5
Normalized intensity profiles for SA = +0.7, 0.0, and –0.7 µm, which were obtained by the
scanning slit optical beam profiler. Red arrows indicate the profile translations caused by
instability of rotation scanning slits. The depth positions of LA are indicated by
blue-letters.
Normalized intensity profiles for SA = +0.7, 0.0, and –0.7 µm, which were obtained by the
scanning slit optical beam profiler. Red arrows indicate the profile translations caused by
instability of rotation scanning slits. The depth positions of LA are indicated by
blue-letters.In order to see more clear differences, we compared the transversal intensity profiles in
cross section at several depths, as shown in Fig. 6
. When positive SA was applied, the transversal width of intensity profile was smaller at
closer to the lens. When negative SA was applied, the transversal width of intensity profile
was smaller at father from the lens. The estimated were 172 µm, 126 µm, and 173 µm for SA = +0.7, 0.0, and
–0.7 µm, respectively. The DOF was extended 1.4 times with SAs. Both of positive and negative
SAs showed the extension of focal range. The minimum lateral resolutions
were 16.6 µm, 14.2 µm, and 16.2 µm for SA = +0.7, 0.0, and
–0.7 µm, respectively. The lateral resolution would become worse 1.2 times with SAs.
Fig. 6
Normalized intensity profiles for SA = +0.7, 0.0, and –0.7 µm in linear scale. The Black
arrows and numbers indicate the full width at half-maximum (FWHM).
Normalized intensity profiles for SA = +0.7, 0.0, and –0.7 µm in linear scale. The Black
arrows and numbers indicate the full width at half-maximum (FWHM).These results indicate first that the effective focal range can be extended by applying SA.
Second, lateral resolution was well preserved with SAs.
6. Phantom imaging
In this section, the extended focal range induced by SA is quantitatively evaluated by imaging
a phantom. The OCT signal intensity, the lateral resolution and DOF were evaluated by an OCT
phantom images and the numerical simulation.
6.1. Methods
In this experiment, a custom made phantom is imaged by AO-SDOCT. The phantom predominantly
consists of agar gel. Polystyrene micro-beads with a 1-µm diameter (Polybead Polystyrene
Microsphere 1.00 µm, Polysciences Inc.) are dispersed in the gel with a concentration of 91
million-beads/ml. The phantom is put on a paper and set at the retinal position of our model
eye with its lens of 16-mm focal length. The paper is used as a reference plane in order to
provide sufficient backscattering for wavefront sensing.Phantom imaging was performed with a 0.8 degree (1024 A-line) scanning size, and 64 OCT
B-scans were obtained at the same location on the phantom. The lateral resolution was estimated
by measuring the bead size of the OCT B-scan images, as demonstrated by Leitgeb et
al. [40] or Ralston et al.
[51]. Applied SA was from –1.0 µm to 1.0 µm with a
0.1-µm step. For the defocus correction, we applied the sum of counter defocus and the defocus
to cancel residual LCA, where the counter defocus was estimated by the method which is
described in Subsection 3.3. In addition, we applied an additional 2.0-µm defocus. This was in
order to shift the minimum circle from the paper plane, which is the reference plane of the
wavefront sensing, to the inside of phantom. The applied SA and defocus are summarized in Table 1
.
Table1
Applied SA and defocus in phantom imaging
SA (µm)
–1.0
–0.9
–0.8
–0.7
–0.6
–0.5
–0.4
–0.3
–0.2
–0.1
Defocus (µm)
4.3
4.0
3.8
3.6
3.4
3.2
2.9
2.5
2.3
2.0
SA (µm)
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
Defocus (µm)
1.8
1.5
1.4
1.1
0.7
0.5
0
–0.2
–0.34
–0.6
–0.86
After taking OCT images, each micro-bead was selected from 64 OCT B-scan images which have a
maximum SNR of larger than 17 dB, and the lateral image size of the micro-bead was estimated.
By selecting the micro-bead with large SNR, we excluded the micro-beads which were illuminated
only at their periphery and were causing the error in the subsequent analysis. The lateral
profile of micro-bead was fitted to a Gaussian function, and 1/e2
width of the Gaussian is calculated for all 64 OCT B-scan images. The average of 64
1/e2 widths was utilized to define the lateral resolution. To
assess the DOF, the variation of the averaged width as a function of the depth was then fitted
to Eq. (6), which is parameterized by
, , andIn the numerical simulation, we calculated the PSFs by using Zemax. The method is similar to
Section 5. The model eye with its lens of 16-mm focal length was used. Then, the theoretical
DOF and the lateral resolution were estimated.
6.2. Results
The OCT images of the phantom with several SAs are shown in Fig. 7
. It was found that, the tightest focus appeared at the SA of +0.3 µm. The reason of this
positive shift is discussed in Subsection 8.1. Note that images with SA of from –0.5 µm to –1.0
µm were not correctly obtained because of vignetting at a spatial filter in the wavefront
sensing arm.
Fig. 7
Phantom images. The numbers in each images are the Zernike coefficients of applied SA.
Reference planes for wavefront sensing are shown in the bottom side of the images. The black
bars indicate 100 µm.
Phantom images. The numbers in each images are the Zernike coefficients of applied SA.
Reference planes for wavefront sensing are shown in the bottom side of the images. The black
bars indicate 100 µm.The variation of the lateral resolution and OCT signal intensity as functions of depth is
shown in Fig. 8(a)
(with –0.4-µm SA), (b) (without SA), (c) (with +0.3-µm SA), and (d) (with +0.6-µm SA).
In addition, the averaged intensity profile, micro beads size and the fitting curve is shown
for each SA. The variation of lateral resolution along the depth was the sharpest in the case
of 0.3-µm SA.
Fig. 8
Variation of bead size, fitting curves and averaged intensity profiles along the depth are
shown with their representative B-scan images in (a) with –0.4 µm-SA, (b) without SA, (c)
with +0.3 µm-SA, and (d) with +0.6 µm-SA, respectively. The black bars indicate 100 µm. The
blue dots indicate the size of micro-beads, the green dashed line indicate the fitting
curve. The red lines and red dashed lines indicate the averaged intensity profiles and their
envelopes. The gray dashed lines and arrows indicate the fitting parameter, ZR.
Variation of bead size, fitting curves and averaged intensity profiles along the depth are
shown with their representative B-scan images in (a) with –0.4 µm-SA, (b) without SA, (c)
with +0.3 µm-SA, and (d) with +0.6 µm-SA, respectively. The black bars indicate 100 µm. The
blue dots indicate the size of micro-beads, the green dashed line indicate the fitting
curve. The red lines and red dashed lines indicate the averaged intensity profiles and their
envelopes. The gray dashed lines and arrows indicate the fitting parameter, ZR.The Rayleigh length for each SA is shown in Fig. 9(a)
. The Rayleigh length was the shortest when +0.3-µm SA was applied.
was approximately 2 times longer with –0.4-µm SA
( = 182 µm) than with +0.3-µm SA
( = 90.5 µm). The minimum lateral resolutions
were reduced in 1.2 times with –0.4-µm SA
( = 4.9 µm) than with +0.3-µm SA
( = 4.2 µm).
Fig. 9
ZR for each SA are shown in (a). RMS wavefront errors for each SA are shown in (b). The +
and × indicate the experimental and the simulation results, respectively.
ZR for each SA are shown in (a). RMS wavefront errors for each SA are shown in (b). The +
and × indicate the experimental and the simulation results, respectively.On the other hands, the numerical simulation results obtained by Zemax showed that the DOF,
which is defined by the FWHM of central peak intensity profile along the depth, was extended
3.5 times with ± 0.7-µm SAs. Specifically, the estimated DOFs were 148 µm, 42.7 µm and 146 µm
for SA = –0.7, 0.0, and +0.7 µm. The lateral resolution, which is defined by beam width, was
degraded 1.5 times with ± 0.7-µm SAs (6.4 µm for both SAs) from that without SA (4.2 µm).Figure 9 shows the RMS wavefront error during the
phantom measurement as a function of SA. Here the RMS error is defined as the RMS error between
the measured wavefront and the target wavefront which includes the controlled SA, the counter
defocus and LCA cancelation. It was found that the RMS error increases as SA increases as same
as Subsection 3.2. This is because of an inevitable property of our AO-control algorithm (see
Appendix A). Note that the RMS errors with SA varying from –0.5 µm to –1.0 µm are
unreliable.
7. Retinal imaging
According to the results of preceding sections, proper parameters for in vivo
eye imaging can be selected. In this section, we demonstrate high-resolution human retinal
imaging with extended DOF. We investigated the effect of SA by imaging several eyes and the
numerical simulation.
7.1. Method
7.1.1. Human retinal imaging for three healthy eyes
Three eyes of 3 normal subjects were involved in this study. The demographics of the
subjects are summarized in Table 2
. The imaging region was set to 5 degree superior to the fovea. The imaged region
on the retina was controlled by a fixation target placed on an optical conjugate of the
retina. For the stability of fixation, the dominant eye of each subject was measured. Before
the measurement, two drops of 0.5% tropicamide and 0.5% phenylephrine hydrochloride were
applied for pupil dilation.
Table 2
Subjects’ characteristics. Sph and Cyl: spherical and cylindrical powers in
diopters
Subject ID
Refractive error
Sph (D)
Cyl (D)
Eye (L/R)
Age
1
Myopia
–7.3
–0.3
R
25
2
Emmetropia
–1.7
–0.1
R
31
3
Myopia
–6.6
–0.4
R
24
In our case, the maximum applicable SA was ± 0.4 µm. In other words, AO-closed loop cannot
converge with a target aberration larger than this maximum. Because there is the inevitable
residual wavefront error discussed in Section 4 and Fig.
3. In this study, we applied SA of ± 0.4 µm and its associated counter defocus.
Optical power on the cornea was 1.28 mW for a probing beam and 96 µW for an AO beacon. These
optical powers respect the safety limit defined by the American National Standard Institute
[52]. Volumetric measurements were performed at
dimensions of 0.8 degree × 0.8 degree (128 × 128) A-lines. Nine sequential volumes were
acquired at a speed of 5.6 volumes/s in a single measurement session.For the image comparison, we generated en face averaged projection images
of the photoreceptor layer (PRL) with a projection thickness of 52 µm (21 pixels). The
representative volumes which appeared with less involuntary eye motion artifacts are selected
from the volumes taken at several measurement sessions. Because the most severe reason for
blurring image was the involuntary eye motion artifacts, the manual selection process was
necessary. For the quantitative assessment of the image quality, we calculated the RMS value
of en face projection image of PRL. The reason why we selected this value is
because of the higher correlation with the manual assessment than the others: sharpness
metric, information entropy and Haralick features.
7.1.2. Intra-subject variability
In order to assess the intra-subject variability of the method, we examined an eye of
subject-1 sequentially by altering the SA in order of –0.4-µm SA, 0.0-µm SA and +0.4-µm SA. In
total, ten measurement sessions (30 volumes) were obtained for each SA. Then, the RMS values
of the en face projection images of PRL, which were the mean signal intensity
of the image, were obtained.
7.1.3. Numerical simulation for human retinal imaging
We numerically simulated the PSFs for human retinal imaging by using Zemax. The method is
similar to Section 5. We used the human eye model by Liou and Brennan [53]. Then, the theoretical DOF and the lateral resolution were
estimated.
7.2. Results
7.2.1. Human retinal imaging for three healthy eyes
Figure 10
shows en face projection images of the photoreceptor layer (PRL). We
qualitatively observed better image quality with –0.4-µm SA for all subjects than 0-µm or
+0.4-µm SA. To more quantify the image, RMS of the image, i.e. the mean signal energy of the
image was obtained. Comparable or higher RMS values were found with –0.4-µm SA than without SA
as shown in Fig. 11(a)
. The image quality with +0.4-µm SA was worse than the others.
Fig. 10
En face projection images on PRL of subject-1 are shown in (1-a) with
+0.4 µm SA, (1-b) without SA and (1-c) with –0.4-µm SA. Those of Subject-2 are shown in
(2-a) with +0.4-µm SA, (2-b) without SA and (2-c) with –0.4-µm SA. Those of Subject-3 are
shown in (3-a) with +0.4-µm SA, (3-b) without SA and (3-c) with –0.4-µm SA. Field of view
of cropped images was 0.64 degree × 0.64 degree (102 pixels × 102 pixels). The white bar
indicates 50 µm on the retina.
Fig. 11
RMS values of en face projection images for SA = +0.4, 0.0 and –0.4 µm
are shown in (a). Residual RMS wavefront errors for = +0.4, 0.0 and –0.4 µm are shown in
(b). The +, ×, and * indicate subject-1, −2, and-3, respectively.
En face projection images on PRL of subject-1 are shown in (1-a) with
+0.4 µm SA, (1-b) without SA and (1-c) with –0.4-µm SA. Those of Subject-2 are shown in
(2-a) with +0.4-µm SA, (2-b) without SA and (2-c) with –0.4-µm SA. Those of Subject-3 are
shown in (3-a) with +0.4-µm SA, (3-b) without SA and (3-c) with –0.4-µm SA. Field of view
of cropped images was 0.64 degree × 0.64 degree (102 pixels × 102 pixels). The white bar
indicates 50 µm on the retina.RMS values of en face projection images for SA = +0.4, 0.0 and –0.4 µm
are shown in (a). Residual RMS wavefront errors for = +0.4, 0.0 and –0.4 µm are shown in
(b). The +, ×, and * indicate subject-1, −2, and-3, respectively.The residual RMS wavefront errors for each SA are shown in Fig. 11(b). Without SA, RMS wavefront errors were less than 0.1 µm for all subjects.
Introduction of SAincreases the RMS error up to around 0.16 µm (results ranged from 0.1 µm to
0.16 µm). Despite this increasing RMS error, we also found that the image quality for all
subjects was better or comparable with –0.4-µm SA than without SA.In order to see the more details, representative B-scan images of subject-1 for each SA are
shown in Fig. 12
. The nerve fiber layer (NFL) was more brightly appeared with –0.4-µm SA than other SAs.
However, the evident improved signal strength of NFL at –0.4-µm such as the case of Fig. 12 was found not for all cases, and there were large
fluctuations of the OCT signal intensity. It was difficult to discriminate between the effects
of SA from the other factors. More details analysis of the effect of SA apart from other
factors based on larger population would be a future study. The details are discussed in
Subsection 8.2.
Fig. 12
Representative B-scan images for SA = +0.4, 0.0 and –0.4 µm. The color bar indicates the
SNR range in which the averaged intensity at the vitreous is set to be 0 dB. The black bar
indicates 100 µm.
Representative B-scan images for SA = +0.4, 0.0 and –0.4 µm. The color bar indicates the
SNR range in which the averaged intensity at the vitreous is set to be 0 dB. The black bar
indicates 100 µm.
7.2.2. Intra-subject variability
To assess the inter-session variability of the method, 30 volumes are obtained with each of
–0.4-µm, 0-µm, and +0.4-µm SAs. The RMS values of en face projection images
were obtained as a measure of image quality. Nine volumes were excluded from the analysis
because of evident eye motion. The mean values of RMS were 16.7, 16.4, and 16.2 for –0.4-µm
SA, 0.0-µm SA and +0.4-µm SA, respectively. The median values of RMS were 16.9, 16.4 and 16.2
for –0.4-µm SA, 0.0-µm SA and +0.4-µm SA, respectively. The largest mean and median values
were found with –0.4-µm SA. This indicates the better image quality with –0.4-µm SA than the
others.The interquartile ranges (IQRs) of the RMS values are 1.5, 1.9, and 1.7 for –0.4-µm SA,
0.0-µm SA and +0.4-µm SA, respectively. The largest IQR was found with 0.0-µm SA. Since larger
IQR value indicates higher variability of the image quality and lower stability of the
imaging, –0.4-µm SA shows the highest stability and 0.0-µm SA shows the lowest stability.
7.2.3. Numerical simulation for human retinal imaging
The numerical simulation results obtained by Zemax showed that the DOF was extended 3 times
with ± 0.4-µm SA. Specifically, the estimated DOFs were 156 µm, 52.7 µm, and 157 µm for SA =
–0.7, 0.0, and +0.7 µm. The lateral resolution was degraded in 1.4 times with ± 0.4-µm SA.
Specifically, 5.8 µm, 4.3 µm, and 6.4 µm for SA = –0.7, 0.0, and +0.7 µm. If we achieved 3
times longer DOF by shrinking the beam diameter on the cornea, the lateral resolution is
degraded in 1.7 times. Namely, 1.2 times better resolution is obtained with 0.4-µm SA than by
simply reducing the beam diameter with the identical DOF.
8. Discussion
8.1. Sample induced spherical aberrations
In Section 5, the tightest focus was found with SA = +0.3 µm not with SA = 0.0 µm. The sample
induced aberrations could explain this issue. Specifically, the refractive index mismatch
between the air and the irregular gel surface induces depth dependent spherical aberrations
[54,55]. We
infer that the sample induced SA was canceled by the above mentioned SA of +0.3 µm. This
assumption is derived from the following reasons. First, the system induced spherical
aberrations is significantly small as mentioned in Section 4. Second, the most probable source
of such amount of aberration is the irregular shape of gel surface. Because if we assumed the
3-mm thickness of a homogenous sample and the gel surface was flat, the expected sample induced
spherical aberrations is approximately –0.03 µm, which is much smaller than the counter +0.3-µm
SA.
8.2. Extension of DOF
According to the B-scan images of a phantom shown in Fig.
7, negative SA (less than +0.3-µm SA) elongates the focal range and shifts it farther
away from the lens. This property is consistent with the theoretical prediction. However, in
the case of positive SA (more than +0.3-µm SA), no evident directionality was found in the
phantom imaging. This is because of the increase of RMS wavefront error as shown in Fig. 9(b). According to Fig.
9(b), the RMS wavefront errors were kept relatively low when negative SA was applied,
while they monotonically increased when positive SA was applied. Therefore, the phantom imaging
also indicates the extension of DOF and their directionality as confirmed in Section 5. In
addition, the averaged intensity profiles were useful to find the OCT signal intensity
enhancement with SA, and it was round 1–2 dB.For the retinal imaging, the slight differences of image contrast were observed among several
SA values as mentioned in Subsections 7.2.1 and 7.2.2. Through the measurement, we
qualitatively and quantitatively observed the equivalent or higher contrast of cone mosaic with
–0.4-µm SA than without SA. However, there were no statistically significant differences. The
reasons would be explained by the larger fluctuation of the OCT signal intensity than the
effect of SA. For instance, the OCT signal intensity gain with SA was found only 1–2 dB in the
phantom imaging. However, for the human retinal imaging, there are large fluctuations from the
other factors: involuntary motion artifacts, structural changes, and OCT signal decay.
8.3. Other benefits for retinal imaging
For the retinal imaging, the involuntary eye motion artifacts were the most severe problem.
Secondly, proper alignment of the depth position of the focus is essential for high contrast
imaging. There are several uncontrollable factors in this alignment. For instance, multi-layer
structures in the retina cause a relatively large depth scattered field of a wavefront sensor
images, hence the exact depth position of wavefront sensing is not predictable. In addition,
micro-fluctuations of the accommodation also result in fluctuations in the depth position of
the reference of the wavefront measurement [16,30]. Involuntary eye motion could also be a cause of
turbulence. In AO retinal imaging, the depth position of the focus is aligned based on an
en face image. This maneuver is easy and practical for AOSLO and a flood
illumination AO fundus camera. However, it is not totally compatible with AOOCT because AOOCT,
with the exception of T-scan time domain AOOCT, required a full volumetric acquisition to
obtain an en face image. In this circumstance, a long focal range obtained by
intentionally induced SA becomes beneficial. This benefit might be resulted in the high
contrast photoreceptor imaging with –0.4-µm SA in Subsections 7.2.1 and 7.2.2, and it also
results in smaller IQRs with SAs in Subsection 7.2.2.
9. Conclusion
In conclusion, an extended DOF system using SA was implemented to extend focal range while
preserving lateral resolution. The extension of the focal range was theoretically, numerically,
and experimentally confirmed. Finally, we demonstrated in vivo human retinal
imaging by altering SAs.
Authors: Barry Cense; Weihua Gao; Jeffrey M Brown; Steven M Jones; Ravi S Jonnal; Mircea Mujat; B Hyle Park; Johannes F de Boer; Donald T Miller Journal: Opt Express Date: 2009-11-23 Impact factor: 3.894
Authors: Robert J Zawadzki; Arlie G Capps; Dae Yu Kim; Athanasios Panorgias; Scott B Stevenson; Bernd Hamann; John S Werner Journal: IEEE J Sel Top Quantum Electron Date: 2014-03 Impact factor: 4.544
Authors: Conor J Sheil; Andreas Wartak; Graham L C Spicer; Guillermo J Tearney Journal: J Opt Soc Am A Opt Image Sci Vis Date: 2022-04-01 Impact factor: 2.104