A method of dose reconstruction for moving targets compatible with dynamic treatments.

Per Rugaard Poulsen1, Mai Lykkegaard Schmidt, Paul Keall, Esben Schjodt Worm, Walther Fledelius, Lone Hoffmann.   

Abstract

PURPOSE: To develop a method that allows a commercial treatment planning system (TPS) to perform accurate dose reconstruction for rigidly moving targets and to validate the method in phantom measurements for a range of treatments including intensity modulated radiation therapy (IMRT), volumetric arc therapy (VMAT), and dynamic multileaf collimator (DMLC) tracking.
METHODS: An in-house computer program was developed to manipulate Dicom treatment plans exported from a TPS (Eclipse, Varian Medical Systems) such that target motion during treatment delivery was incorporated into the plans. For each treatment, a motion including plan was generated by dividing the intratreatment target motion into 1 mm position bins and construct sub-beams that represented the parts of the treatment that were delivered, while the target was located within each position bin. For each sub-beam, the target shift was modeled by a corresponding isocenter shift. The motion incorporating Dicom plans were reimported into the TPS, where dose calculation resulted in motion including target dose distributions. For experimental validation of the dose reconstruction a thorax phantom with a moveable lung equivalent rod with a tumor insert of solid water was first CT scanned. The tumor insert was delineated as a gross tumor volume (GTV), and a planning target volume (PTV) was formed by adding margins. A conformal plan, two IMRT plans (step-and-shoot and sliding windows), and a VMAT plan were generated giving minimum target doses of 95% (GTV) and 67% (PTV) of the prescription dose (3 Gy). Two conformal fields with MLC leaves perpendicular and parallel to the tumor motion, respectively, were generated for DMLC tracking. All treatment plans were delivered to the thorax phantom without tumor motion and with a sinusoidal tumor motion. The two conformal fields were delivered with and without portal image guided DMLC tracking based on an embedded gold marker. The target dose distribution was measured with a radiochromic film in the moving rod and compared with the reconstructed doses using gamma tests.
RESULTS: Considerable interplay effects between machine motion and target motion were observed for the treatments without tracking. For nontracking experiments, the mean 2 mm∕2% gamma pass rate over all investigated scenarios was 99.6% between calculated and measured doses. For tracking experiments, the mean gamma pass rate was 99.4%.
CONCLUSIONS: A method for accurate dose reconstruction for moving targets with dynamic treatments was developed and experimentally validated in a variety of delivery scenarios. The method is suitable for integration into TPSs, e.g., for reconstruction of the dose delivered to moving tumors or calculation of target doses delivered with DMLC tracking.

Entities:  

Mesh:

Year:  2012        PMID: 23039659      PMCID: PMC3470610          DOI: 10.1118/1.4754297

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  22 in total

1.  A method for incorporating organ motion due to breathing into 3D dose calculations.

Authors:  A E Lujan; E W Larsen; J M Balter; R K Ten Haken
Journal:  Med Phys       Date:  1999-05       Impact factor: 4.071

2.  Validation of dynamic MLC-controller log files using a two-dimensional diode array.

Authors:  Jonathan G Li; James F Dempsey; Li Ding; Chihray Liu; Jatinder R Palta
Journal:  Med Phys       Date:  2003-05       Impact factor: 4.071

3.  Effects of motion on the total dose distribution.

Authors:  Thomas Bortfeld; Steve B Jiang; Eike Rietzel
Journal:  Semin Radiat Oncol       Date:  2004-01       Impact factor: 5.934

4.  A fluence convolution method to account for respiratory motion in three-dimensional dose calculations of the liver: a Monte Carlo study.

Authors:  Indrin J Chetty; Mihaela Rosu; Neelam Tyagi; Lon H Marsh; Daniel L McShan; James M Balter; Benedick A Fraass; Randall K Ten Haken
Journal:  Med Phys       Date:  2003-07       Impact factor: 4.071

5.  A stochastic convolution/superposition method with isocenter sampling to evaluate intrafraction motion effects in IMRT.

Authors:  Shahid A Naqvi; Warren D D'Souza
Journal:  Med Phys       Date:  2005-04       Impact factor: 4.071

6.  Dosimetric validation of the anisotropic analytical algorithm for photon dose calculation: fundamental characterization in water.

Authors:  Antonella Fogliata; Giorgia Nicolini; Eugenio Vanetti; Alessandro Clivio; Luca Cozzi
Journal:  Phys Med Biol       Date:  2006-02-21       Impact factor: 3.609

7.  The effects of intra-fraction organ motion on the delivery of dynamic intensity modulation.

Authors:  C X Yu; D A Jaffray; J W Wong
Journal:  Phys Med Biol       Date:  1998-01       Impact factor: 3.609

8.  Dosimetric consequences of intrafraction prostate motion.

Authors:  Haisen S Li; Indrin J Chetty; Charles A Enke; Ryan D Foster; Twyla R Willoughby; Patrick A Kupellian; Timothy D Solberg
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008-01-30       Impact factor: 7.038

9.  Limitations of a convolution method for modeling geometric uncertainties in radiation therapy. I. The effect of shift invariance.

Authors:  Tim Craig; Jerry Battista; Jake Van Dyk
Journal:  Med Phys       Date:  2003-08       Impact factor: 4.071

10.  Verification of dynamic and segmental IMRT delivery by dynamic log file analysis.

Authors:  Dale W Litzenberg; Jean M Moran; Benedick A Fraass
Journal:  J Appl Clin Med Phys       Date:  2002       Impact factor: 2.102

View more
  13 in total

1.  The first clinical implementation of electromagnetic transponder-guided MLC tracking.

Authors:  Paul J Keall; Emma Colvill; Ricky O'Brien; Jin Aun Ng; Per Rugaard Poulsen; Thomas Eade; Andrew Kneebone; Jeremy T Booth
Journal:  Med Phys       Date:  2014-02       Impact factor: 4.071

Review 2.  Review of Real-Time 3-Dimensional Image Guided Radiation Therapy on Standard-Equipped Cancer Radiation Therapy Systems: Are We at the Tipping Point for the Era of Real-Time Radiation Therapy?

Authors:  Paul J Keall; Doan Trang Nguyen; Ricky O'Brien; Pengpeng Zhang; Laura Happersett; Jenny Bertholet; Per R Poulsen
Journal:  Int J Radiat Oncol Biol Phys       Date:  2018-04-14       Impact factor: 7.038

3.  Dosimetric benefit of DMLC tracking for conventional and sub-volume boosted prostate intensity-modulated arc radiotherapy.

Authors:  Tobias Pommer; Marianne Falk; Per R Poulsen; Paul J Keall; Ricky T O'Brien; Peter Meidahl Petersen; Per Munck af Rosenschöld
Journal:  Phys Med Biol       Date:  2013-03-14       Impact factor: 3.609

4.  Assessing the dosimetric impact of real-time prostate motion during volumetric modulated arc therapy.

Authors:  Juan Diego Azcona; Lei Xing; Xin Chen; Karl Bush; Ruijiang Li
Journal:  Int J Radiat Oncol Biol Phys       Date:  2014-04-01       Impact factor: 7.038

5.  4D dose simulation in volumetric arc therapy: Accuracy and affecting parameters.

Authors:  Thilo Sothmann; Tobias Gauer; René Werner
Journal:  PLoS One       Date:  2017-02-23       Impact factor: 3.240

6.  Real-time 4D dose reconstruction for tracked dynamic MLC deliveries for lung SBRT.

Authors:  Cornelis Ph Kamerling; Martin F Fast; Peter Ziegenhein; Martin J Menten; Simeon Nill; Uwe Oelfke
Journal:  Med Phys       Date:  2016-11       Impact factor: 4.071

7.  A novel probabilistic approach to generating PTV with partial voxel contributions.

Authors:  H S Tsang; C P Kamerling; P Ziegenhein; S Nill; U Oelfke
Journal:  Phys Med Biol       Date:  2017-04-05       Impact factor: 3.609

8.  Evaluation of delivered dose to a moving target by 4D dose reconstruction in gated volumetric modulated arc therapy.

Authors:  Hyekyun Chung; Jinhong Jung; Chiyoung Jeong; Jungwon Kwak; Jin-Hong Park; Su Ssan Kim; Sang Min Yoon; Si Yeol Song; Jong Hoon Kim; Eun Kyung Choi; Seungryong Cho; Byungchul Cho
Journal:  PLoS One       Date:  2018-09-07       Impact factor: 3.240

9.  Impact of organ motion on volumetric and dosimetric parameters in stomach lymphomas treated with intensity-modulated radiotherapy.

Authors:  Yusuke Uchinami; Ryusuke Suzuki; Norio Katoh; Hiroshi Taguchi; Koichi Yasuda; Naoki Miyamoto; Yoichi M Ito; Shinichi Shimizu; Hiroki Shirato
Journal:  J Appl Clin Med Phys       Date:  2019-08-10       Impact factor: 2.102

10.  Assessment of MLC tracking performance during hypofractionated prostate radiotherapy using real-time dose reconstruction.

Authors:  M F Fast; C P Kamerling; P Ziegenhein; M J Menten; J L Bedford; S Nill; U Oelfke
Journal:  Phys Med Biol       Date:  2016-01-27       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.