External evaluation of the Dimension Vista 1500® intelligent lab system.

Arnaud Bruneel1, Monique Dehoux, Anne Barnier, Anne Boutten.   

Abstract

BACKGROUND: Dimension Vista® analyzer combines four technologies (photometry, nephelometry, V-LYTE® integrated multisensor potentiometry, and LOCI® chemiluminescence) into one high-throughput system.
METHODS: We assessed analytical performance of assays routinely performed in our emergency laboratory according to the VALTEC protocol, and practicability.
RESULTS: Precision was good for most parameters. Analytical domain was large and suitable for undiluted analysis in most clinical settings encountered in our hospital. Data were comparable and correlated to our routine analyzers (Roche Modular DP®, Abbott AXSYM®, Siemens Dimension® RxL, and BN ProSpec®). Performance of nephelometric and LOCI modules was excellent. Functional sensitivity of high-sensitivity C-reactive protein and cardiac troponin I were 0.165 mg/l and 0.03 ng/ml, respectively (coefficient of variation; CV < 10%). The influence of interfering substances (i.e., hemoglobin, bilirubin, or lipids) was moderate, and Dimension Vista® specifically alerted for interference according to HIL (hemolysis, icterus, lipemia) indices. Good instrument performance and full functionality (no reagent or sample carryover in the conditions evaluated, effective sample-volume detection, and clot detection) were confirmed. Simulated routine testing demonstrated excellent practicability, throughput, ease of use of software and security.
CONCLUSION: Performance and practicability of Dimension Vista® are highly suitable for both routine and emergency use. Since no volume detection and thus no warning is available on limited sample racks, pediatric samples require special caution to the Siemens protocol to be analyzed in secured conditions. Our experience in routine practice is also discussed, i.e., the impact of daily workload, "manual" steps resulting from dilutions and pediatric samples, maintenances, flex hydration on instrument's performance on throughput and turnaround time.
© 2012 Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23001985      PMCID: PMC6807579          DOI: 10.1002/jcla.21539

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


  16 in total

1.  Evaluation of imprecision for cardiac troponin assays at low-range concentrations.

Authors:  Mauro Panteghini; Franca Pagani; Kiang-Teck J Yeo; Fred S Apple; Robert H Christenson; Francesco Dati; Johannes Mair; Jan Ravkilde; Alan H B Wu
Journal:  Clin Chem       Date:  2003-12-04       Impact factor: 8.327

2.  The effect of hemolysis on current troponin assays--a confounding preanalytical variable?

Authors:  Chris Florkowski; John Wallace; Trevor Walmsley; Peter George
Journal:  Clin Chem       Date:  2010-04-22       Impact factor: 8.327

3.  The effect of sample hemolysis on cardiac troponin I and T assays.

Authors:  Renze Bais
Journal:  Clin Chem       Date:  2010-06-04       Impact factor: 8.327

4.  Luminescent oxygen channeling assay (LOCI): sensitive, broadly applicable homogeneous immunoassay method.

Authors:  E F Ullman; H Kirakossian; A C Switchenko; J Ishkanian; M Ericson; C A Wartchow; M Pirio; J Pease; B R Irvin; S Singh; R Singh; R Patel; A Dafforn; D Davalian; C Skold; N Kurn; D B Wagner
Journal:  Clin Chem       Date:  1996-09       Impact factor: 8.327

5.  Graphical comparisons of interferences in clinical chemistry instrumentation.

Authors:  M R Glick; K W Ryder; S A Jackson
Journal:  Clin Chem       Date:  1986-03       Impact factor: 8.327

6.  French Society of Clinical Biology. Commission for Validation of Methods.

Authors: 
Journal:  Ann Biol Clin (Paris)       Date:  1986       Impact factor: 0.459

7.  Evaluation of the Technicon Dax 48: a multiparametric biochemical analyzer.

Authors:  N Ben Naoui; C Periou; C Harault; G Le Moel
Journal:  Ann Biol Clin (Paris)       Date:  1993       Impact factor: 0.459

Review 8.  Analytical characteristics of high-sensitivity cardiac troponin assays.

Authors:  Fred S Apple; Paul O Collinson
Journal:  Clin Chem       Date:  2011-09-30       Impact factor: 8.327

9.  [Evaluation of the analytical performance characteristics of NT-proBNP immunoassay on dimension RxL-HM (Dade Behring)].

Authors:  M-O Benoit; F Dautezac; M Taccoen; J-L Paul
Journal:  Ann Biol Clin (Paris)       Date:  2005 Jul-Aug       Impact factor: 0.459

10.  Cardiovascular disease testing on the Dimension Vista system: biomarkers of acute coronary syndromes.

Authors:  Walter E Kelley; Christina M Lockwood; Denise R Cervelli; Jamie Sterner; Mitchell G Scott; Show-Hong Duh; Robert H Christenson
Journal:  Clin Biochem       Date:  2009-06-10       Impact factor: 3.281

View more
  3 in total

1.  Galectin 3: association to neurohumoral activity, echocardiographic parameters and renal function in outpatients with heart failure.

Authors:  Freja Stoltze Gaborit; Helle Bosselmann; Caroline Kistorp; Kasper Iversen; Thomas Kumler; Finn Gustafsson; Jens P Goetze; György Sölétormos; Niels Tønder; Morten Schou
Journal:  BMC Cardiovasc Disord       Date:  2016-05-31       Impact factor: 2.298

2.  Prevalence and risk factors of preconception anemia: A community based cross sectional study of rural women of reproductive age in northeastern Tanzania.

Authors:  Omari A Msemo; Ib C Bygbjerg; Sofie L Møller; Birgitte B Nielsen; Lars Ødum; Kathrine Perslev; John P A Lusingu; Reginald A Kavishe; Daniel T R Minja; Christentze Schmiegelow
Journal:  PLoS One       Date:  2018-12-18       Impact factor: 3.240

3.  Association between left ventricular global longitudinal strain and natriuretic peptides in outpatients with chronic systolic heart failure.

Authors:  F Gaborit; H Bosselmann; N Tønder; K Iversen; T Kümler; C Kistorp; G Sölétormos; J P Goetze; M Schou
Journal:  BMC Cardiovasc Disord       Date:  2015-08-20       Impact factor: 2.298

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.