Literature DB >> 19523464

Cardiovascular disease testing on the Dimension Vista system: biomarkers of acute coronary syndromes.

Walter E Kelley1, Christina M Lockwood, Denise R Cervelli, Jamie Sterner, Mitchell G Scott, Show-Hong Duh, Robert H Christenson.   

Abstract

OBJECTIVES: Performance characteristics of the LOCI cTnI, CK-MB, MYO, NTproBNP and hsCRP methods on the Dimension Vista System were evaluated. DESIGN AND METHODS: Imprecision (following CLSI EP05-A2 guidelines), limit of quantitation (cTnI), limit of blank, linearity on dilution, serum versus plasma matrix studies (cTnI), and method comparison studies were conducted.
RESULTS: Method imprecision of 1.8 to 9.7% (cTnI), 1.8 to 5.7% (CK-MB), 2.1 to 2.2% (MYO), 1.6 to 3.3% (NTproBNP), and 3.5 to 4.2% (hsCRP) were demonstrated. The manufacturer's claimed imprecision, detection limits and upper measurement limits were met. Limit of Quantitation was 0.040 ng/mL for the cTnI assay. Agreement of serum and plasma values for cTnI (r=0.99) was shown. Method comparison study results were acceptable.
CONCLUSIONS: The Dimension Vista cTnI, CK-MB, MYO, NTproBNP, and hsCRP methods demonstrate acceptable performance characteristics for use as an aid in the diagnosis and risk assessment of patients presenting with suspected acute coronary syndromes.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19523464     DOI: 10.1016/j.clinbiochem.2009.05.020

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  5 in total

1.  External evaluation of the Dimension Vista 1500® intelligent lab system.

Authors:  Arnaud Bruneel; Monique Dehoux; Anne Barnier; Anne Boutten
Journal:  J Clin Lab Anal       Date:  2012-09       Impact factor: 2.352

2.  Serial Troponin for Early Detection of Nivolumab Cardiotoxicity in Advanced Non-Small Cell Lung Cancer Patients.

Authors:  Matteo Sarocchi; Francesco Grossi; Eleonora Arboscello; Andrea Bellodi; Carlo Genova; Maria Giovanna Dal Bello; Erika Rijavec; Giulia Barletta; Giovanni Rossi; Federica Biello; Giorgio Ghigliotti; Marco Canepa; Michele Mussap; Claudio Brunelli; Paolo Spallarossa
Journal:  Oncologist       Date:  2018-03-22

3.  Performance characteristics of loci method for measuring cardiac troponin I on the dimension EXL.

Authors:  José L Martín Calderón; Julia M Varona Pérez; Fernando Bustos; Juan C Sánchez Gómez
Journal:  Pract Lab Med       Date:  2015-04-08

4.  An agomir of miR-144-3p accelerates plaque formation through impairing reverse cholesterol transport and promoting pro-inflammatory cytokine production.

Authors:  Yan-Wei Hu; Ya-Rong Hu; Jia-Yi Zhao; Shu-Fen Li; Xin Ma; Shao-Guo Wu; Jing-Bo Lu; Yu-Rong Qiu; Yan-Hua Sha; Yan-Chao Wang; Ji-Juan Gao; Lei Zheng; Qian Wang
Journal:  PLoS One       Date:  2014-04-14       Impact factor: 3.240

5.  The importance of having a flexible scope ISO 15189 accreditation and quality specifications based on biological variation--the case of validation of the biochemistry analyzer Dimension Vista.

Authors:  Pilar Fernandez-Calle; Sandra Pelaz; Paloma Oliver; Maria Jose Alcaide; Ruben Gomez-Rioja; Antonio Buno; Jose Manuel Iturzaeta
Journal:  Biochem Med (Zagreb)       Date:  2013       Impact factor: 2.313

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.