Huiqiong Xu1, Jiansheng Liang1, Yimei Wang1, Bin Wang1, Tianbao Zhang2, Xiaoli Liu1, Lin Gong1.
Abstract
The ATP bioluminescence method has been increasingly employed as a rapid, on-site detection method in nosocomial infections control. In this study, we used a paired design of monitoring methods, the colony counting method (C) and the ATP bioluminescence method, to evaluate environmental surfaces after disinfection. The ATP bioluminescence method included three detector types (B, P, and N). Every surface after disinfection was performed by combining two types of monitoring methods or detectors. There was no statistically significant difference in theATP content per surface siteamong samples from intensive care units (ICUs)and internal medicine wards using B (p = 0.435) and P (p = 0.260). According to the Spearman's rank correlation coefficients, with the exception of the correlation between the ATP content values detected by B and P, the correlation between the values generated by the remaining methods/detectors was weak or lacking, whereasthe differences between the detectors were statistically significant. Therefore, there are differences between the ATP bioluminescence method and the colony counting method, also between different detectors.Entities:
Mesh:
Substances:
Year: 2019 PMID: 31490948 PMCID: PMC6730931 DOI: 10.1371/journal.pone.0221665
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
The standard curve and linear correlation coefficient (r) of each detectors.
| Detector | Standard curve | Detector | Standard curve | ||
|---|---|---|---|---|---|
| B (1) | y = 0.893x+0.990 | 0.987 | N (1) | y = 0.914x-0.785 | 0.996 |
| B (2) | y = 0.987x+1.378 | 0.999 | N (2) | y = 0.905x-0.761 | 0.996 |
| B (3) | y = 1.004x+0.862 | 0.998 | N (3) | y = 0.931x-0.852 | 0.998 |
| P (1) | y = 0.941x-1.714 | 0.990 | N (4) | y = 0.874x-0.409 | 0.995 |
| P (2) | y = 1.022x-0.241 | 0.996 | N (5) | y = 0.851x-0.356 | 0.994 |
| P (3) | y = 1.002x-1.992 | 0.998 | N (6) | y = 0.853x-0.388 | 0.994 |
a x and y were fitted with the base 10 logarithms of the ATP content (10−17 mol) and RLU.
Distribution of CFU and ATP content per surface site after disinfection.
| Method/Detector | Source | ||||||
|---|---|---|---|---|---|---|---|
| C (CFU) | ICU | 155 | 0.0 | 0.00~1.00 | 0.0 | 95.0 | <0.001 |
| Internal medicine wards | 170 | 0.0 | 0.00~3.00 | 0.0 | 1 900.0 | <0.001 | |
| B (10−17 mol) | ICU | 201 | 436.24 | 123.32~3280.43 | 8.78 | 331 911.00 | <0.001 |
| Internal medicine wards | 238 | 358.67 | 131.70~2033.11 | 8.78 | 98 142.00 | <0.001 | |
| P (10−17 mol) | ICU | 182 | 581.60 | 99.32~1941.85 | 50.00 | 72 200.00 | <0.001 |
| Internal medicine wards | 227 | 678.32 | 194.29~2995.08 | 66.29 | 120 400.00 | <0.001 | |
| N (10−17 mol) | ICU | 135 | 183.42 | 89.73~328.88 | 27.22 | 1 845.60 | <0.001 |
| Internal medicine wards | 170 | 251.04 | 108.07~558.39 | 27.22 | 5 440.99 | <0.001 |
Correlationbetween different method/detector to determine the CFU and ATP content (per surface site), using the Spearman rank correlation coefficients (r).
| Method/Detector | C (CFU) | B (10-17mol) | P (10−17 mol) | N (10−17 mol) | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| C (CFU) | - | - | - | 0.293 | 0.005 | 90 | 0.244 | 0.029 | 80 | 0.313 | 0.049 | 40 |
| B (10−17 mol) | 0.450 | <0.001 | 93 | - | - | - | 0.904 | <0.001 | 121 | 0.246 | 0.062 | 58 |
| P(10−17 mol) | 0.485 | <0.001 | 90 | 0.931 | <0.001 | 139 | - | - | - | 0.073 | 0.613 | 51 |
| N (10−17 mol) | 0.564 | <0.001 | 40 | 0.378 | <0.001 | 81 | 0.466 | <0.001 | 70 | - | - | - |
a It shows the result from the ICUs.
b It shows the result from the internal medicine wards.
Comparison of the ATP content values (10-17mol) measured using different detectors, using the Wilcoxon signed-rank test.
| Paired Detectors | Source | |||
|---|---|---|---|---|
| P&B | 121 | ICU | -6.512 | <0.001 |
| 139 | internal medicine wards | -3.951 | <0.001 | |
| 260 | - | -7.475 | <0.001 | |
| N&B | 58 | ICU | -3.132 | 0.002 |
| 81 | internal medicine wards | -3.515 | <0.001 | |
| 139 | - | -4.726 | <0.001 | |
| P&N | 51 | ICU | -1.790 | 0.073 |
| 70 | internal medicine wards | -2.917 | 0.004 | |
| 121 | - | -3.483 | <0.001 |