Joseph T Rakowski1, Karen Chin, Sandeep Mittal.
Abstract
Calvarial reconstruction following resection of tumors involving the skull is often followed by stereotactic radiosurgery. Prior studies have addressed the effects of various cranioplasty materials on dose distributions in linac-based radiosurgery. We aim to determine the effects of titanium mesh implants on Gamma Knife dose. Radiation backscatter and transmission were measured for eight types of titanium mesh using film, ion chamber, and Theratron Co-60 teletherapy device. A single mesh was selected for Gamma Knife irradiation using a CaSO(4) skull filled with ballistics gel. Dose profiles for reconstructed and intact skulls were compared with the planning system prediction at 2.5 and 5.5 cm depth. Titanium contact backscatter and transmission dose perturbations ranged from -18% to 23%. Radiation dose measured at 1.5 cm below the calvarial implant increased by 0.5% to 3.3% relative to bone. Measured Gamma Knife dose profile diameters agreed with expected profiles. Maximum dose within the intact phantom was 3% less than planned due to skull attenuation. Maximum dose within the reconstructed phantom was between the intact phantom and planned doses. Titanium mesh implants and hydroxyapatite cranioplasty result in minimal alteration (< 3%) in the delivered Gamma Knife dose.Entities:
Mesh:
Substances:
Year: 2012 PMID: 22955648 PMCID: PMC5718236 DOI: 10.1120/jacmp.v13i5.3833
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1The eight different meshes evaluated in this study. Dimensions: 1. ; 2. ; 3. ; 4. ; 5. d: 10.1 cm t: 0.6 mm; 6: d: 7.35 cm t: 0.4 mm; 7. ; 8. .
Ion chamber transmission measurement ratios.
|
|
|
|
|---|---|---|
| 1 | 1.033 | 1.033 |
| 2 | 1.025 | 1.025 |
| 3 | 1.026 | 1.026 |
| 4 | 1.020 | 1.019 |
| 5 | 1.027 | 1.027 |
| 6 | 1.032 | 1.032 |
| 7 | 1.026 | 1.026 |
| 8 | 1.026 | 1.026 |
| 0.5 cm hydroxyapatite | 1.005 | 1.005 |
Figure 2Intact skull phantom in Leksell frame (a) and resected with mesh and gel (b).
Figure 3Mesh profiles ((a)‐(i)) immediately above and below the mesh compared to the open field for eight different titanium meshes and 0.5 cm hydroxyapatite.
Figure 4Treatment planning axial CT slices through intact skull for 5.5 cm depth (a) and 2.5 cm depth (b).
Figure 52.5 and 5 cm depths x‐ and z‐axis coronal plane dose profiles ((a)‐(d)) from Leksell Gamma Knife treatment plan, measured dose profile for unresected phantom, and measured dose profile for resected with mesh.