Abstract
Monte Carlo simulation with PENELOPE (version 2003) is applied to calculate Leksell Gamma Knife dose distributions for heterogeneous phantoms. The usual spherical water phantom is modified with a spherical bone shell simulating the skull and an air-filled cube simulating the frontal or maxillary sinuses. Different simulations of the 201 source configuration of the Gamma Knife have been carried out with a simplified model of the geometry of the source channel of the Gamma Knife recently tested for both single source and multisource configurations. The dose distributions determined for heterogeneous phantoms including the bone- and/or air-tissue interfaces show non-negligible differences with respect to those calculated for a homogeneous one, mainly when the Gamma Knife isocentre approaches the separation surfaces. Our findings confirm an important underdosage (approximately 10%) nearby the air-tissue interface, in accordance with previous results obtained with the PENELOPE code with a procedure different from ours. On the other hand, the presence of the spherical shell simulating the skull produces a few per cent underdosage at the isocentre wherever it is situated.Entities:
Mesh:
Substances:
Year: 2005 PMID: 16306660 DOI: 10.1088/0031-9155/50/23/018
Source DB: PubMed Journal: Phys Med Biol ISSN: 0031-9155 Impact factor: 3.609