Dosimetric evaluation of a 320 detector row CT scanner unit.

Mario de Denaro1, Paola Bregant.   

Abstract

BACKGROUND: The technologic improvements in Multislice scanners include the increment in the X-ray beam width. Some new CT scanners are equipped with a 320 detector row which allows a longitudinal coverage of 160 mm and a total of 640 slices for a single rotation. When such parameters are used the length of the traditional pencil chamber (10 cm) is no more appropriate to measure the standard weighted computed tomography dose index (CTDI(w)) value. MATERIALS AND METHODS.: Dosimetric measurements were performed on a 640 slices Toshiba Aquilion One CT scanner using common instrumentation available in Medical Physics Departments.
RESULTS: For the measurements in air, two different ionization chambers were completely exposed to the beam. Dosimeters showed an acceptable agreement in the measurements. To evaluate the actual shape of the dose profile strips of Gafchromic XRQA film were used. Films were previously calibrated on site. From the graphic response of the scanned film it is possible to evaluate the full width at half maximum (FWHM) of the dose profile which represent the actual beam width.
CONCLUSIONS: Computed Tomography Dose Index (CTDI) and Dose Length Product (DLP) need to be changed when the beam width of the CT scanner is over 100 mm. To perform dose evaluation with the conventional instrumentation, two parameters should be considered: the average absorbed dose and the actual beam width. To measure the average absorbed dose, the conventional ionization chamber can be used. For the measurement of the width of the dose profile, Gafchromic XRQA film seemed to be suitable.

Entities:  

Keywords:  Gafchromic film; computed tomography dose index; dose profile; multislice CT; pencil chamber

Year:  2010        PMID: 22933937      PMCID: PMC3423722          DOI: 10.2478/v10019-010-0049-1

Source DB:  PubMed          Journal:  Radiol Oncol        ISSN: 1318-2099            Impact factor:   2.991


Introduction

The advent of multislice scanners technology1 leads to a continuous increment of the X-ray beam width along the cranio-caudal direction. The new CT scanner from Toshiba, Aquilion One, is equipped with the 320 detector rows each one 0.5 mm wide. This new technology allows a longitudinal coverage of 160 mm at rotation time of 0.35 seconds. The important change in dosimetric evaluation is needed to adapt the CT dosimetry metrics to the new standard.2–5 With a 160 mm beam width as an operating condition, the length of the traditional pencil chamber (10 cm) is no more appropriate to evaluate the standard computed tomography dose index (CTDI).6 Despite this problem the CTDI and the dose length product (DLP) values are still displayed on the scanner console. The problem for medical physicists, who must carry out quality assurance and dose optimization on the new CT equipment, is that suitable instrumentation is usually not available. In the present work we suggest some dosimetric measurements that can be carried out on a 320 row CT scanner unit, by means of conventional dosimetric instrumentation.

Material and methods

Some measurements on a 640 slices Toshiba Aquilion One CT scanner were carried out by means of common instrumentation typically employed at Medical Physics Departments. CT scanners providing 160 mm beam width allows to completely expose the 6cc ionization chamber of a Radcal Dosimeter, model 9015, commonly employed in the conventional radiodiagnostic measurement. To evaluate the average absorbed dose due to the primary beam, we compared the measurement carried out in air at the isocenter of the gantry, during a single rotation, exposing together the Radcal Dosimeter and the standard 10 cm long CT pencil chamber (WDCT 10, Wellhofer, Germany), connected to PMX-III electrometer. The average absorbed dose alone does not give information regarding the profile of the dose along the cranio-caudal direction (Z axis); it is a parameter to be considered to evaluate the shape and the actual width of the beam. To measure the dose profile free in air Gafchromic XRQA film was exposed to the primary beam during a single axial rotation. The film was previously calibrated exposing small pieces of Gafchromic (2 × 2 cm) to increasing doses; the Gafchromic films were put close to a reference ionization chamber. A calibration curve specific for the CT scanner beam quality was then obtained and it was used to convert the pixel values of the film into dose values (Figure 1).7
FIGURE 1.

Fitting curve for the calibration of Gafchromic XRQA film.

The exposed film was then scanned by a regular scanner (Epson 1680 pro) using red-green-blue (RGB) modality at 48 bits of depth and spatial resolution of 72 dots per inch. Only the red component of the image was considered. The dose profile was obtained correcting the pixel values by the calibration curve (Figure 2).
FIGURE 2.

Dose profile measured by Gafchromic XRQA film.

From the graph the full width at half maximum (FWHM) of the dose profile can be evaluated. FWHM is a parameter which is related to the Z-axis geometric efficiency. CTDIair,300 (CTDI value using an integration interval of 300 mm) and CTDIair,100 (CTDI value using an integration interval of 100 mm) were evaluated and compared (Figure 3).2
FIGURE 3.

(A) Computed tomography dose index (CTDI)air,300. (B) CTDIair, 100.

Finally, we put a strip of Gafchromic film in the holes of a standard CT dosimetric PMMA phantom (Figure 4) to evaluate the average absorbed dose inside the phantom (Figure 5).
FIGURE 4.

Strip of Gafchromic film inserted inside the hole of a poliymethyl methacrylate (PMMA) phantom.

FIGURE 5.

Average dose evaluation by Gafchromic film into the poliymethyl methacrylate (PMMA) phantom.

Results

The measurement performed by the Radcal dosimeter showed an acceptable agreement with the values obtained by the Wellhofer pencil chamber (Table 1). The comparison was meaningful because the 6 cc chamber of the Radcal dosimeter is completely exposed by the beam width.
TABLE 1.

Absorbed dose evaluated with two different dosimeters

DosimeterDose FOV Head (mGy/mAs)Dose FOV Body (mGy/mAs)
Wellhöfer pencil chamber (WDCT 10), with PMX-III0.2570.225
Radcal model 9015with 6cc chamber 10X5–60.2720.243

FOV, field-of-view; WDCT 10, Wellhöfer CT pencil chamber 10 cm long

To maintain the meaning of the CTDI, the integration interval for 160 mm beam width must be extended to a value of at least 300 mm.2 For the measurement of the weighted computed tomography dose index (CTDIw) a PMMA phantom is required, but at the moment it is very difficult to find and manage a PMMA phantom with a length of 300 mm. Moreover, also for the measurement of CTDI in air the length of the conventional pencil chamber (100 mm) is inappropriate. A way to perform dosimetric measurements using conventional instrumentation is to use the ionization chamber for the measurement of the average absorbed dose and Gafchromic film for the evaluation of the actual shape of the dose profile along z-axis. In Figure 1 the shape of the dose profile is plotted; an asymmetry appears evident in the profile due to the anode heel effect. Analysing the profile FWHM was calculated to be 171.8 mm pointing out an overbeaming of 11.8 mm. Z-axis geometric efficiency was also evaluated; it is defined by the International Electrotechnical Commission (IEC)8 as the ratio between the integral over the range subtended by detectors and total integral of the dose profile. The measured value of 95.4% showed for this parameter a low contribution of the penumbra effect. In Figure 3 two different intervals of integration to evaluate the CTDIair are presented; the value of CTDIair,300 was 36.7 mGy and the value of CTDIair,100 was 34.5 mGy. The result was as expected because the longer interval takes into account also the contribution due to the tails of the dose profile. To obtain an estimation of the absorbed dose inside the PMMA phantom, a strip of Gafchromic film was inserted in the phantom holes. The resulting value of 74.7 mGy was compared with the value of 81.4 mGy displayed at the CT console as CTDIvol. The comparison showed a significant underestimation of the measured value probably due to the length of the PMMA phantom (15 cm) which is not long enough to simulate completely the effect of the scattered radiation.

Discussion

Multislice CT scanners lead to a progressive increment in the X-ray beam width along z-axis. CTDI appears to be no more suitable to represent the main dosimetric quantity in CT dosimetry. Recent publications try to find new dose parameters more representative of the technological state of art.9 The conventional quality assurance and the dose assessment would require an upgrade in the conventional instrumentation, but new suitable dosimeters and phantoms are not yet available for most Medical Physics Department. Nevertheless, dosimetric parameters for a CT scanner with a beam width of 16 cm can be estimated by means of commonly available dosimetric instrumentation. The average absorbed dose related to a single axial rotation can be measured by conventional ionization chambers, while the geometric dose distribution along a profile in the cranio-caudal direction (Z-axis) can be evaluated exposing Gafchromic film to the primary beam.
  5 in total

1.  Evaluation of image quality and dose in renal colic: comparison of different spiral-CT protocols.

Authors:  A Rimondini; R Pozzi Mucelli; M De Denaro; P Bregant; L Dalla Palma
Journal:  Eur Radiol       Date:  2001       Impact factor: 5.315

2.  Is it time to retire the CTDI for CT quality assurance and dose optimization?

Authors:  David J Brenner
Journal:  Med Phys       Date:  2005-10       Impact factor: 4.071

3.  In vivo dosimetry for estimation of effective doses in multislice CT coronary angiography.

Authors:  M de Denaro; P Bregant; M Severgnini; F de Guarrini
Journal:  Med Phys       Date:  2007-10       Impact factor: 4.071

4.  Computed tomography dose assessment for a 160 mm wide, 320 detector row, cone beam CT scanner.

Authors:  J Geleijns; M Salvadó Artells; P W de Bruin; R Matter; Y Muramatsu; M F McNitt-Gray
Journal:  Phys Med Biol       Date:  2009-05-06       Impact factor: 3.609

5.  Diagnostic imaging of traumatic pseudoaneurysm of the thoracic aorta.

Authors:  Serif Beslic; Nermina Beslic; Selma Beslic; Amela Sofic; Muris Ibralic; Jasmina Karovic
Journal:  Radiol Oncol       Date:  2010-09-09       Impact factor: 2.991

  5 in total
  6 in total

1.  The importance of accurate treatment planning, delivery, and dose verification.

Authors:  Julian Malicki
Journal:  Rep Pract Oncol Radiother       Date:  2012-03-06

2.  Influence of difference in cross-sectional dose profile in a CTDI phantom on X-ray CT dose estimation: a Monte Carlo study.

Authors:  Tomonobu Haba; Shuji Koyama; Yoshihiro Ida
Journal:  Radiol Phys Technol       Date:  2013-11-24

3.  Dosimetric evaluation of a 256-slice computed tomography scanner.

Authors:  Anna Radice; Claudio Ielasi; Gabriele D'Andrea; Nicoletta Paruccini; Andrea Crespi
Journal:  Radiol Med       Date:  2014-03-21       Impact factor: 3.469

4.  Does body mass index outperform body weight as a surrogate parameter in the calculation of size-specific dose estimates in adult body CT?

Authors:  Johannes Boos; Rotem S Lanzman; Philipp Heusch; Joel Aissa; Christoph Schleich; Christoph Thomas; Lino M Sawicki; Gerald Antoch; Patric Kröpil
Journal:  Br J Radiol       Date:  2015-12-23       Impact factor: 3.039

5.  Application of a color scanner for (60)Co high dose rate brachytherapy dosimetry with EBT radiochromic film.

Authors:  Mahdi Ghorbani; Mohammad Taghi Bahreyni Toossi; Ali Asghar Mowlavi; Shahram Bayani Roodi; Ali Soleimani Meigooni
Journal:  Radiol Oncol       Date:  2012-11-09       Impact factor: 2.991

6.  Estimation of primary radiation output for wide-beam computed tomography scanner.

Authors:  Atsushi Fukuda; Pei-Jan P Lin; Nao Ichikawa; Kosuke Matsubara
Journal:  J Appl Clin Med Phys       Date:  2019-05-02       Impact factor: 2.102

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.