| Literature DB >> 22915789 |
Robert G Hauser1, Raed H Abdelhadi, Deepa M McGriff, Linda Kallinen Retel.
Abstract
AIM: The purpose of this study was to determine if Optim™, a unique copolymer of silicone and polyurethane, protects Riata ST Optim and Durata implantable cardioverter-defibrillator (ICD) leads (SJM, St Jude Medical Inc., Sylmar, CA, USA) from abrasions that cause lead failure. METHODS ANDEntities:
Mesh:
Substances:
Year: 2012 PMID: 22915789 PMCID: PMC3549219 DOI: 10.1093/europace/eus245
Source DB: PubMed Journal: Europace ISSN: 1099-5129 Impact factor: 5.214
Riata ST Optim leads
| Case no. | Model | MDR event key | Lead age | Sign | Analysis |
|---|---|---|---|---|---|
| 1 | 7020/65 | 2486749 | 44 | Low impedance | Can abrasion |
| 2 | 7022/65 | 2409042 | 48 | Low R-wave | Outer insulation abrasion RV cable fracture |
| 3 | 7020/65 | 2332700 | 49 | High capture threshold | Can abrasion |
| 4 | 7020/65 | 2332774 | 34 | Noise | Can abrasion |
| Aborted therapy | |||||
| 5 | 7022/65 | 2242754 | 33 | Inappropriate shocks | Can abrasion |
| Arc on can | RV cable melted; HV failure | ||||
| 6 | 7020/60 | 1788235 | 37 | Therapy truncated | Abrasion 52 cm from connector due to friction with another device; HV failure |
| Patient died | |||||
| 7 | 7020/65 | 1627573 | 32 | Oversensing; loss of RV coil EGM | Internal insulation abrasion 6. 7 cm from the distal tip. ETFE damaged |
| 8 | 7020/60 | 1573108 | 30 | Not specified | Can abrasion. Electrically intact |
| 9 | 7020/60 | 1573109 | 28 | Inappropriate shocks | Abrasion 50 cm from connector due to friction with another device. |
| P/S cables exposed | |||||
| 10 | 7021/65 | 1524192 | 17 | Inappropriate shocks | Can abrasion |
| P/S cable melted; HV failure | |||||
| 11 | 7022/65 | 1411014 | 21 | Low impedance | Can abrasion |
| P/S cable exposed | |||||
| 12 | 7070/65 | 1411029 | 16 | High impedance | Pacing coil fracture; clavicular crush |
| Incidental connector insulation abrasion | |||||
| 13 | 7020/60 | 1337721 | 20 | HV short during VF | Outer insulation abrasion; HV failure |
| HV RV cables and SVC coil were melted | |||||
| 14 | 7021/65 | 1337731 | 19 | Noise | Can abrasion |
| P/S cable exposed | |||||
| 15 | 7022/60 | 1010619 | 9 | Inappropriate shocks | Abrasion 52 cm from connector due to friction with another device |
RV, right ventricle; SVC, superior vena cava; HV, high-voltage; P/S, pace/sense; EGM, electrogram; VF, ventricular fibrillation.
Durata leads
| Case no. | Model | MDR event key | Lead age | Sign | Analysis |
|---|---|---|---|---|---|
| 1 | 7120/65 | 2487350 | 48 | Low impedance | Internal abrasion 23 cm from distal tip. RV conductor abraded: ‘This could cause the reported low impedance when in contact with the SVC shock coil’. |
| 2 | 7120/65 | 2332798 | 38 | Low impedance | Insulation abrasion near connector pin |
| 3 | 7120/65 | 2409021 | 36 | Insulation break near yolk | Analysis found insulation abrasion in unspecified location |
| 4 | 7120/65 | 2463178 | 35 | Intermittent loss of capture | Abrasion adjacent to proximal end of RV coil; friction with annuloplasty ring |
| 5 | 7120/65 | 2409091 | 35 | Low impedance | Can abrasion |
| RV cables exposed | |||||
| 6 | 7122/65 | 2409053 | 33 | Oversensing | Analysis found insulation abrasion in unspecified location |
| 7 | 7170/65 | 2332939 | 33 | Oversensing | Analysis found insulation abrasion in unspecified location |
| Inappropriate shocks | |||||
| 8 | 7120/65 | 2084498 | 33 | Low impedance | Can abrasion |
| RV cables exposed; ETFE compromised | |||||
| 9 | 7121/65 | 2084194 | 31 | Oversensing Inappropriate shocks | Can abrasion |
| SVC cable abraded | |||||
| 10 | 7170/65 | 2084197 | 30 | Oversensing Inappropriate shocks | Can abrasion |
| Sensing cable exposed | |||||
| 11 | 7120/65 | 2333118 | 29 | Inappropriate shocks | Analysis found insulation abrasion in unspecified location |
| PG in back-up mode | |||||
| 12 | 7120/65 | 1954890 | 27 | Inappropriate shocks | Analysis found insulation abrasion in unspecified location |
| 13 | 7120/65 | 2015091 | 26 | Oversensing | Can abrasion or friction with another lead |
| Sensing coil flattened | |||||
| 14 | 7121/65 | 2487053 | 25 | Unspecified | Abrasion 27 cm from connector |
| Caused by friction with another device | |||||
| 15 | 7120/60 | 2014710 | 25 | Unspecified | Abrasion 53 cm from connector exposing HV RV cables due to friction with another device |
| 16 | 7170/65 | 2157205 | 24 | Low impedance | Internal abrasion between 7 and 8 cm from the lead tip causing short circuit between the sensing and pacing conductor |
| 17 | 7120/65 | 1955135 | 24 | Abnormal sensing | Can abrasion affecting sensing cable insulation |
| 18 | 7120/60 | 1679666 | 24 | Inappropriate shocks | Abrasion at proximal end of RV shock coil due to friction with another device |
| 19 | 7122/65 | 2487319 | 22 | Unspecified | Can abrasion |
| Coil melted | |||||
| 20 | 7120/65 | 1679680 | 22 | Low impedance | Abrasion under the proximal end of the SVC coil. Intermittent contact between the RV shock cables and SVC coil |
| 21 | 7122/65 | 1376803 | 22 | Noise | Abrasion was observed on the lead at 14. 8–15. 2 cm from the connector pin |
| 22 | 7122Q/65 | 2332898 | 20 | Break | Abrasion probably due to clavicular crush |
| 23 | 7122/65 | 2015063 | 20 | Unspecified | Abrasion at 57. 7–58 cm at the proximal end of the RV shock coil due to friction with another device |
| 24 | 7120/65 | 1895671 | 20 | Inappropriate shocks | Failure due to sensing coil fracture. Incidental connector abrasion |
| 25 | 7122Q/58 | 2242920 | 19 | Unspecified | Can abrasion. RV cables exposed. Electrically normal |
| 26 | 7122/65 | 1896015 | 19 | Low impedance | Can abrasion. RV cable exposed. |
| 27 | 7121/65 | 1679744 | 17 | High impedance | Can abrasion |
| Failure due to fracture. | |||||
| 28 | 7120/65 | 2332800 | 16 | Unspecified | Abrasion in unspecified location |
| 29 | 7121/65 | 1752115 | 15 | Low impedance | Partial lead surface abrasion |
| 30 | 7122/65 | 2409093 | 10 | Low impedance | Abrasion was found between 38. 8 and 40. 9 cm from connector pin |
| 31 | 7121/65 | 1830483 | 10 | High impedance | Failure due to fracture. |
| Oversensing | Incidental connector abrasion | ||||
| 32 | 7121Q/58 | 2084058 | 9 | Noise | Abrasion at 44. 1–45 cm from connector pin due to friction with another device |
| Low R-wave | |||||
| 33 | 7121Q/58 | 2015064 | 6 | Break | Abrasion at 22 cm from the connector pin. The ETFE insulation of the svc cable and proximal cable were abraded due to clavicular crush |
| 34 | 7122/60 | 1411082 | 6 | Noise | Can abrasion. Cables exposed. |
| 35 | 7122Q/65 | 2084562 | 5 | Inappropriate shocks | Abrasion caused by clavicular crush |
| Low impedance | |||||
| 36 | 7120Q/65 | 2242898 | 4 | Unspecified | Surface abrasion 22 cm from connector |
| 37 | 7120/60 | 1627605 | 3 | Unspecified | Can abrasion exposing the RV and SVC cables |
RV, right ventricle; SVC, superior vena cava; ETFE, ethylene-tetrafluoroethylene.
Signs of Riata ST Optim and Durata failure. A lead may have had more than one sign of failure
| Sign | PG can abrasion | Other device abrasion | Internal abrasion | Unspecified abrasion | Other abrasion |
|---|---|---|---|---|---|
| Death, truncated shock | 1 | ||||
| High-voltage short during VF | DF-1 cable and SVC coil melted | ||||
| Low impedance | 5 | 3 | 3 | Clavicle-2 | |
| Inappropriate shocks | 4 | 3 | 3 | Clavicle-1 | |
| Oversensing, noise | 6 | 1 | 2 | 4 | Fracture-1 |
| Low R-wave | 1 | 1 | 1 | ||
| ‘Break’ | 1 | Clavicle-1 | |||
| High impedance | Clavicle-1 | ||||
| Loss of capture, rising threshold | 1 | Annuloplasty ring-1 | |||
| Unspecified | 4 | 3 | 3 |