Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of >10 years.

Thomas Kleemann1, Torsten Becker, Klaus Doenges, Margit Vater, Jochen Senges, Steffen Schneider, Werner Saggau, Udo Weisse, Karlheinz Seidl.   

Abstract

BACKGROUND: The number of patients with longer follow-up after implantation of an implantable cardioverter-defibrillator is increasing continuously. Defibrillation lead failure is a typical long-term complication. Therefore, the long-term reliability of implantable cardioverter-defibrillator leads has become an increasing concern. The aim of the present study was to assess the annual rate of transvenous defibrillation lead defects related to follow-up time after lead implantation. METHODS AND
RESULTS: A total of 990 consecutive patients who underwent first implantation of an implantable cardioverter-defibrillator between 1992 and May 2005 were analyzed. Median follow-up time was 934 days (interquartile range, 368 to 1870). Overall, 148 defibrillation leads (15%) failed during the follow-up. The estimated lead survival rates at 5 and 8 years after implantation were 85% and 60%, respectively. The annual failure rate increased progressively with time after implantation and reached 20% in 10-year-old leads (P<0.001). Lead defects affected newer as well as older models. Patients with lead defects were 3 years younger at implantation and more often female. Multiple lead implantation was associated with a trend to a higher rate of defibrillation lead defects (P=0.06). The major lead complications were insulation defects (56%), lead fractures (12%), loss of ventricular capture (11%), abnormal lead impedance (10%), and sensing failure (10%).
CONCLUSIONS: An increasing annual lead failure rate is noted primarily during long-term follow-up and reached 20% in 10-year-old leads. Patients with lead defects are younger and more often female.

Entities:  

Mesh:

Year:  2007        PMID: 17470696     DOI: 10.1161/CIRCULATIONAHA.106.663807

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  105 in total

Review 1.  Dual- versus single-coil implantable defibrillator leads: review of the literature.

Authors:  Jörg Neuzner; Jörg Carlsson
Journal:  Clin Res Cardiol       Date:  2012-01-10       Impact factor: 5.460

2.  Axillary sub-pectoral pulse generator pocket for lowering defibrillation threshold.

Authors:  Ernest W Lau
Journal:  J Interv Card Electrophysiol       Date:  2012-02-23       Impact factor: 1.900

3.  Device therapy: a subcutaneous ICD--preliminary results.

Authors:  Paolo Spirito; Giuseppe Boriani
Journal:  Nat Rev Cardiol       Date:  2010-10       Impact factor: 32.419

Review 4.  Initial experience of subcutaneous implantable cardioverter defibrillators in Singapore: a case series and review of the literature.

Authors:  Tien Siang Eric Lim; Boon Yew Tan; Kah Leng Ho; Chuh Yih Paul Lim; Wee Siong Teo; Chi-Keong Ching
Journal:  Singapore Med J       Date:  2015-10       Impact factor: 1.858

Review 5.  Considerations for cardiac device lead extraction.

Authors:  Oussama Wazni; Bruce L Wilkoff
Journal:  Nat Rev Cardiol       Date:  2016-01-29       Impact factor: 32.419

6.  Use and Abuse of Internal Cardioverter Defibrillators for Primary Prevention.

Authors:  Joshua R Silverstein; Demosthenes G Katritsis; Mark E Josephson
Journal:  Arrhythm Electrophysiol Rev       Date:  2012-09

7.  The Entirely Subcutaneous Defibrillator - A New Generation and Future Expectations.

Authors:  Hussam Ali; Pierpaolo Lupo; Riccardo Cappato
Journal:  Arrhythm Electrophysiol Rev       Date:  2015-08

8.  Remote monitoring in implantable defibrillator therapy.

Authors:  D A M J Theuns; L S Jordaens
Journal:  Neth Heart J       Date:  2008-02       Impact factor: 2.380

Review 9.  [ICD lead defects: diagnosis and therapeutical options].

Authors:  Christian G Wollmann; Dirk Böcker; Andreas Löher; Hans H Scheld; Günter Breithardt; Rainer Gradaus
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2009-02-11

10.  Long-term performance of the St Jude Riata 1580-1582 ICD lead family.

Authors:  S D A Valk; D A M J Theuns; L Jordaens
Journal:  Neth Heart J       Date:  2013-03       Impact factor: 2.380

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.