Oxygenator safety evaluation: a focus on connection grip strength and arterial temperature measurement accuracy.

Richard F Newland1, Robert A Baker, Andrew J Sanderson, Sigrid C Tuble, Phil J Tully.   

Abstract

This report describes the assessment of three specific safety-related specifications in the consideration of an alternate oxygenator; first the grip strength relationship between various oxygenator connectors and SMARxT tubing, second, the grip strength of various biopassive tubings and an isolated SMARxT connector, and finally, the accuracy of the arterial outlet temperature measurement. Grip strength experiments for the connections between the SMARxT tubing and the venous reservoir outlet and the oxygenator venous inlet and oxygenator arterial outlet of the Medtronic Affinity, Sorin Synthesis, Sorin Primox, and Terumo Capiox RX25 oxygenators were performed. In addition we compared the grip strength of polyvinyl chloride, Physio, Trillium, Carmeda, X-Coating, and SMARxT tubing. The accuracy of the integrated arterial outlet temperature probes was determined by comparing the temperatures measured by the integrated probe with a precision reference thermometer. Connector grip strength comparisons for the evaluation oxygenators with SMARxT tubing showed significant variation between oxygenators and connections (p = .02). Evaluation of the arterial outlet showed significant variation between evaluation oxygenators, while at the venous reservoir outlet and oxygenator inlet, there were no significant differences. Grip strength comparison data for the various tubing types demonstrated a main effect for tubing type F(5, 18) = 8.01, p = .002, eta(p)(2) = .77. Temperature accuracy measurements demonstrated that all oxygenators overread the arterial outlet temperature at 15 degrees C, whilst at temperatures > or = 25 degrees C, all oxygenators underread the arterial outlet temperature. The integrity of SMARxT tubing connection is influenced by the connector type, and may decline over time, highlighting the importance to not consider interchanging components of the bypass circuit as inconsequential.

Entities:  

Mesh:

Year:  2012        PMID: 22893983      PMCID: PMC4557451     

Source DB:  PubMed          Journal:  J Extra Corpor Technol        ISSN: 0022-1058


  15 in total

1.  The effect of oxygenator membranes on blood: a comparison of two oxygenators in open-heart surgery.

Authors:  Joseph Noora; Andre Lamy; Kelly M Smith; Rosanne Kent; Dianne Batt; John Fedoryshyn; Xiaoyin Wang
Journal:  Perfusion       Date:  2003-09       Impact factor: 1.972

2.  Disconnection of Cobe SMARxT tubing from the venous outlet of the Terumo Capiox SX25RX oxygenator during cardiopulmonary bypass.

Authors:  Jane Ottens; Robert A Baker; Andrew J Sanderson; Richard F Newland
Journal:  J Extra Corpor Technol       Date:  2010-06

3.  Accuracy of temperature measurement in the cardiopulmonary bypass circuit.

Authors:  Richard F Newland; Andrew J Sanderson; Robert A Baker
Journal:  J Extra Corpor Technol       Date:  2005-03

4.  In vitro validation of the Affinity NT oxygenator arterial outlet temperatures.

Authors:  Kieron C Potger; Darryl McMillan
Journal:  J Extra Corpor Technol       Date:  2005-06

Review 5.  An evidence-based review of the practice of cardiopulmonary bypass in adults: a focus on neurologic injury, glycemic control, hemodilution, and the inflammatory response.

Authors:  Kenneth G Shann; Donald S Likosky; John M Murkin; Robert A Baker; Yvon R Baribeau; Gordon R DeFoe; Timothy A Dickinson; Timothy J Gardner; Hilary P Grocott; Gerald T O'Connor; David J Rosinski; Frank W Sellke; Timothy W Willcox
Journal:  J Thorac Cardiovasc Surg       Date:  2006-08       Impact factor: 5.209

6.  A clinical evaluation of platelet function, haemolysis and oxygen transfer during cardiopulmonary bypass comparing the Quantum HF-6700 to the HF-5700 hollow fibre membrane oxygenator.

Authors:  A O Chukwuemeka; M R Turtle; U H Trivedi; G E Venn; D J Chambers
Journal:  Perfusion       Date:  2000-11       Impact factor: 1.972

7.  Clinical evaluation of nine hollow-fibre membrane oxygenators.

Authors:  P A Segers; J F Heida; I de Vries; C Maas; A J Boogaart; S Eilander
Journal:  Perfusion       Date:  2001-03       Impact factor: 1.972

8.  Clinical gaseous microemboli assessment of an oxygenator with integral arterial filter in the pediatric population.

Authors:  Thomas J Preston; Daniel Gomez; Vincent F Olshove; Alistair Phillips; Mark Galantowicz
Journal:  J Extra Corpor Technol       Date:  2009-12

9.  SMART tubing presents an increased risk of disconnection during extracorporeal circulation.

Authors:  Ross Newling; Richard Morris
Journal:  J Extra Corpor Technol       Date:  2005-12

10.  Comparison of a Duraflo II-coated cardiopulmonary bypass circuit and a trillium-coated oxygenator during open-heart surgery.

Authors:  Tom N Hoel; Vibeke Videm; Svein T Baksaas; Tom E Mollnes; Frank Brosstad; Jan L Svennevig
Journal:  Perfusion       Date:  2004-05       Impact factor: 1.972

View more
  4 in total

Review 1.  Development and hemocompatibility testing of nitric oxide releasing polymers using a rabbit model of thrombogenicity.

Authors:  Terry C Major; Hitesh Handa; Gail M Annich; Robert H Bartlett
Journal:  J Biomater Appl       Date:  2014-06-16       Impact factor: 2.646

2.  Commentary: Round and round the mesenchymal stromal cells go; where they stop, we hope to know.

Authors:  Gregory S Matte; James A DiNardo
Journal:  JTCVS Open       Date:  2021-02-23

3.  The Society of Thoracic Surgeons, The Society of Cardiovascular Anesthesiologists, and The American Society of ExtraCorporeal Technology: Clinical Practice Guidelines for Cardiopulmonary Bypass--Temperature Management during Cardiopulmonary Bypass.

Authors:  Richard Engelman; Robert A Baker; Donald S Likosky; Alina Grigore; Timothy A Dickinson; Linda Shore-Lesserson; John W Hammon
Journal:  J Extra Corpor Technol       Date:  2015-09

Review 4.  Control of Blood Coagulation by Hemocompatible Material Surfaces-A Review.

Authors:  Janna Kuchinka; Christian Willems; Dmitry V Telyshev; Thomas Groth
Journal:  Bioengineering (Basel)       Date:  2021-12-15
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.