SMART tubing presents an increased risk of disconnection during extracorporeal circulation.

Ross Newling1, Richard Morris.   

Abstract

A number of products exhibiting biocompatible features have been developed for use in extracorporeal blood circuits during cardiopulmonary bypass procedures. While attention has been focused on biocompatibility features of the blood-circuit interface, a number of issues applicable in clinical use of these circuits have arisen. Surface Modifying Additive Technology (SMART; Cobe Cardiovascular, Arvarda, CO) is one such technology. In this product, the structure of normal polyvinylchloride (PVC) tubing is altered through the blending of two copolymers to give a more biocompatible blood to plastic interface. In this study, we examined the in vitro mechanical ability of random samples (n = 10) of SMART and standard PVC tubing to withstand axial tension when the tubing was placed over a single barb of a connector. The tension required to remove the SMART tubing from the connector (83.3 +/- 7.3 [SD] N), was significantly less than standard PVC tubing (115.6 +/- 15.9 N; p < .0001, unpaired t test). The SMART tubing exhibited a 28% reduction in tubing to connector adhesion, which may have a significant effect on extracorporeal circuit disconnection and overall patient safety.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16524161      PMCID: PMC4680835     

Source DB:  PubMed          Journal:  J Extra Corpor Technol        ISSN: 0022-1058


  4 in total

1.  SMA circuits reduce platelet consumption and platelet factor release during cardiac surgery.

Authors:  J O Defraigne; J Pincemail; G Dekoster; R Larbuisson; M Dujardin; F Blaffart; J L David; R Limet
Journal:  Ann Thorac Surg       Date:  2000-12       Impact factor: 4.330

2.  Cardiopulmonary bypass circuit treated with surface-modifying additives: a clinical evaluation of blood compatibility.

Authors:  Y J Gu; P W Boonstra; A A Rijnsburger; J Haan; W van Oeveren
Journal:  Ann Thorac Surg       Date:  1998-05       Impact factor: 4.330

3.  Hematologic evaluation of cardiopulmonary bypass circuits prepared with a novel block copolymer.

Authors:  F D Rubens; R S Labow; G R Lavallée; M I Watson; J A Robblee; M E Voorhees; H J Nathan
Journal:  Ann Thorac Surg       Date:  1999-03       Impact factor: 4.330

4.  Surface modifying additives for improved device-blood compatibility.

Authors:  C C Tsai; R M Deppisch; L J Forrestal; G H Ritzau; A D Oram; H J Göhl; M E Voorhees
Journal:  ASAIO J       Date:  1994 Jul-Sep       Impact factor: 2.872

  4 in total
  2 in total

1.  Disconnection of Cobe SMARxT tubing from the venous outlet of the Terumo Capiox SX25RX oxygenator during cardiopulmonary bypass.

Authors:  Jane Ottens; Robert A Baker; Andrew J Sanderson; Richard F Newland
Journal:  J Extra Corpor Technol       Date:  2010-06

2.  Oxygenator safety evaluation: a focus on connection grip strength and arterial temperature measurement accuracy.

Authors:  Richard F Newland; Robert A Baker; Andrew J Sanderson; Sigrid C Tuble; Phil J Tully
Journal:  J Extra Corpor Technol       Date:  2012-06
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.