Purpose. To evaluate macular thickness, agreement, and intraclass repeatability in three optical coherence tomography (OCT) devices: the time domain (TD) Stratus OCT and two spectral domain (SD) OCTs, Spectralis and Cirrus SD-OCT, in eyes with macular edema secondary to diabetic retinopathy (DR) and retinal vein occlusion (VO). Methods. In a prospective observational study at a university-based retina practice, retinal thickness tomography was performed simultaneously for fifty-eight patients (91 eyes) with DR and VO employing a time domain and two spectral domain OCTs. Agreement in macular measurements was assessed by constructing Bland-Altman plots. Intraclass repeatability was assessed by intraclass correlation coefficients (ICCs). Results. Based on the Bland-Altman plots for central macular thickness, there was low agreement between the measurements of Cirrus SD-OCT and Stratus OCT, Spectralis OCT and Stratus OCT, as well as Spectralis OCT and Cirrus SD-OCT among DR and RVO patients. All three devices demonstrated high intraclass repeatability, with ICC of 98% for Stratus OCT, 97% for Cirrus SD-OCT, and 100% for Spectralis OCT among DR patients. The ICC was 97% for Stratus OCT, 79% for Cirrus SD-OCT, and 91% for Spectralis OCT among RVO patients. Conclusion. There are low agreements among interdevice measurements. However, intraclass repeatability is high in both TD and SD-OCT devices.Entities:
Year: 2012 PMID: 22888403 PMCID: PMC3410350 DOI: 10.1155/2012/354783
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Mean ± standard deviation of nine standard subfields of three OCT devices.
| Diabetic macular edema | Vein occlusion | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Before correction | After correction | Before correction | After correction | |||||||
| Stratus OCT | Cirrus SD-OCT | Spectralis OCT | Cirrus SD-OCT | Spectralis OCT | Stratus OCT | Cirrus SD-OCT | Spectralis OCT | Cirrus SD-OCT | Spectralis OCT | |
| Central | 282 ± 99 | 323 ± 108 | 344 ± 116 | 323 ± 108 | 340 ± 108 | 283 ± 149 | 315 ± 156 | 359 ± 198 | 353 ± 241 | 376 ± 245 |
| Inner superior | 299 ± 65 | 340 ± 60 | 363 ± 66 | 339 ± 60 | 358 ± 67 | 302 ± 100 | 357 ± 111 | 389 ± 161 | 377 ± 162 | 390 ± 168 |
| Inner nasal | 290 ± 58 | 336 ± 63 | 357 ± 68 | 336 ± 62 | 359 ± 72 | 319 ± 118 | 348 ± 87 | 389 ± 139 | 380 ± 167 | 397 ± 158 |
| Inner inferior | 280 ± 73 | 325 ± 80 | 346 ± 83 | 324 ± 80 | 343 ± 77 | 307 ± 110 | 327 ± 61 | 360 ± 101 | 366 ± 147 | 378 ± 144 |
| Inner temporal | 295 ± 83 | 343 ± 74 | 364 ± 82 | 343 ± 73 | 351 ± 76 | 300 ± 117 | 340 ± 97 | 364 ± 146 | 363 ± 175 | 381 ± 178 |
| Outer superior | 251 ± 36 | 289 ± 44 | 319 ± 42 | 289 ± 44 | 317 ± 40 | 258 ± 76 | 299 ± 69 | 345 ± 104 | 311 ± 79 | 331 ± 83 |
| Outer nasal | 261 ± 57 | 305 ± 66 | 319 ± 50 | 304 ± 66 | 323 ± 68 | 275 ± 60 | 301 ± 58 | 336 ± 67 | 327 ± 86 | 341 ± 82 |
| Outer inferior | 239 ± 84 | 283 ± 97 | 295 ± 34 | 325 ± 80 | 305 ± 87 | 258 ± 88 | 272 ± 51 | 329 ± 109 | 291 ± 63 | 326 ± 97 |
| Outer temporal | 253 ± 55 | 290 ± 53 | 314 ± 61 | 290 ± 53 | 306 ± 61 | 242 ± 74 | 277 ± 44 | 300 ± 76 | 287 ± 78 | 298 ± 74 |
Mean difference between two OCT devices and 95% confidence interval of nine standard subfields.
| Diabetic macular edema | Vein occlusion | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Before correction | After correction | Before correction | After correction | |||||||||
| Spectralis OCT versus Cirrus SD-OCT | Cirrus SD-OCT versus Stratus OCT | Spectralis OCT versus Stratus OCT | Spectralis OCT versus Cirrus SD-OCT | Cirrus SD-OCT versus Stratus OCT | Spectralis OCT versus Stratus OCT | Spectralis OCT versus Cirrus SD-OCT | Cirrus SD-OCT versus Stratus OCT | Spectralis OCT versus Stratus OCT | Spectralis OCT versus Cirrus SD-OCT | Cirrus SD-OCT versus Stratus OCT | Spectralis OCT versus Stratus OCT | |
| Central | 21 [9, 33] | 41 [29,53] | 61 [51,72] | 17 [5,28] | 41 [29,53] | 58 [45,70] | 47 [7,86] | 31 [1,64] | 76 [48,104] | 26 [13,38] | 68 [20,117] | 93 [45,142] |
| Inner superior | 25 [11, 39] | 40 [26,54] | 65 [58,72] | 18 [5,32] | 40 [26,54] | 59 [48,69] | 31 [14,76] | 55 [41,70] | 84 [50,118] | 15 [3,33] | 75 [44,107] | 90 [53,125] |
| Inner nasal | 21 [9, 33] | 46 [39,54] | 67 [57,77] | 23 [11,34] | 46 [38,54] | 69 [59,79] | 44 [7,81] | 26 [4,56] | 70 [59,80] | 20 [3,38] | 58 [36, 81] | 78 [58,97] |
| Inner inferior | 21 [14, 28] | 45 [36,53] | 66 [58,74] | 18 [12,24] | 45 [36,53] | 63 [54,72] | 34 [6,75] | 18 [22,58] | 53 [43,63] | 15 [3,26] | 56 [22,90] | 71 [42,99] |
| Inner temporal | 21 [13, 29] | 47 [36,59] | 68 [60,77] | 8 [3,19] | 47 [36,59] | 56 [40,71] | 25 [12,61] | 41 [20,61] | 65 [46,83] | 19 [8,31] | 63 [35,93] | 82 [50,114] |
| Outer superior | 26 [14, 38] | 38 [29,47] | 68 [60,75] | 26 [15,37] | 38 [29,47] | 64 [57,70] | 43 [5,81] | 41 [12,69] | 79 [50,107] | 22 [7,35] | 52 [40,65] | 72 [61,83] |
| Outer nasal | 13 [6, 19] | 43 [39,48] | 56 [52,60] | 18 [10,26] | 43 [39,48] | 61 [53,68] | 34 [1,67] | 25 [6,56] | 59 [54,64] | 15 [4,26] | 51 [40,63] | 66 [55,77] |
| Outer inferior | 21 [12, 30] | 44 [37,52] | 64 [58,69] | 26 [10, 42] | 44 [36,51] | 61 [56,66] | 43 [12,97] | 25 [12,61] | 59 [48,70] | 19 [13,25] | 44 [37, 50] | 62 [53,72] |
| Outer temporal | 21 [13, 30] | 38 [30,46] | 63 [55,70] | 11 [2, 20] | 86 [70,101] | 51 [40,62] | 19 [0,37] | 38 [16,60] | 58 [47,69] | 11 [6,16] | 48 [37,58] | 56 [50,62] |
Figure 1Bland-Altman plots of central subfield macular thickness between each device for the values before manual correction of algorithms (a, c, and e) and after manual correction (b, d, and f) in eyes with diabetic retinopathy. Solid lines: average mean difference; dotted lines: 95% confidence limits of agreement.
Intraclass correlation coefficient and 95% confidence interval for repeated measures of each subfield by each OCT device.
| Diabetic macular edema | Vein occlusion | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Before correction | Stratus OCT | After correction | Before correction | Stratus OCT | After correction | |||||
| Cirrus SD-OCT | Spectralis OCT | Cirrus SD-OCT | Spectralis OCT | Cirrus SD-OCT | Spectralis OCT | Cirrus SD-OCT | Spectralis OCT | |||
| ICCa (%) [95% CIb] | ICC (%) [95% CI] | ICC (%) [95% CI] | ICC (%) [95% CI] | ICC (%) [95% CI] | ICC (%) [95% CI] | ICC (%) [95% CI] | ICC (%) [95% CI] | ICC (%) [95% CI] | ICC (%) [95% CI] | |
| Central | 97 | 100 | 98 | 97 | 99 | 79 | 91 | 97 | 99 | 100 |
| Inner superior | 82 | 100 | 93 | 82 | 100 | 93 | 100 | 97 | 99 | 99 |
| Inner nasal | 98 | 100 | 95 | 98 | 99 | 89 | 83 | 99 | 98 | 100 |
| Inner inferior | 96 | 100 | 97 | 96 | 100 | 93 | 95 | 97 | 99 | 99 |
| Inner temporal | 94 | 100 | 98 | 94 | 98 | 97 | 98 | 96 | 99 | 100 |
| Outer superior | 81 | 99 | 90 | 80 | 93 | 98 | 87 | 96 | 99 | 99 |
| Outer nasal | 99 | 100 | 97 | 99 | 95 | 98 | 98 | 94 | 97 | 100 |
| Outer inferior | 98 | 98 | 98 | 98 | 100 | 99 | 99 | 97 | 99 | 99 |
| Outer temporal | 90 | 97 | 88 | 90 | 89 | 91 | 100 | 91 | 98 | 100 |
aICC: intraclass correlation coefficient.
bCI: confidence interval.
Coefficient of repeatability (μm) for each subfield in each OCT device.
| Diabetic macular edema | Vein occlusion | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Cirrus SD-OCT | Stratus OCT | Spectralis OCT | Cirrus SD-OCT | Stratus OCT | Spectralis OCT | |||||
| Before correction | After correction | Before correction | After correction | Before correction | After correction | Before correction | After correction | |||
| Central | 36 | 36 | 27 | 5 | 18 | 146 | 30 | 54 | 119 | 7 |
| Inner superior | 52 | 52 | 33 | 9 | 8 | 54 | 30 | 33 | 18 | 28 |
| Inner nasal | 18 | 18 | 26 | 6 | 14 | 60 | 49 | 19 | 110 | 11 |
| Inner inferior | 31 | 30 | 24 | 8 | 7 | 30 | 31 | 37 | 48 | 24 |
| Inner temporal | 35 | 35 | 22 | 7 | 20 | 33 | 25 | 50 | 41 | 20 |
| Outer superior | 36 | 36 | 23 | 6 | 20 | 20 | 15 | 27 | 64 | 19 |
| Outer nasal | 12 | 11 | 21 | 4 | 29 | 16 | 30 | 34 | 19 | 8 |
| Outer inferior | 24 | 23 | 26 | 10 | 8 | 10 | 10 | 31 | 15 | 11 |
| Outer temporal | 34 | 34 | 40 | 20 | 39 | 32 | 19 | 55 | 9 | 11 |
Figure 2Bland-Altman plots of central subfield macular thickness between each device for the values before manual correction of algorithms (a, c, and e) and after manual correction (b, d, and f) in eyes with vein occlusion. Solid lines: average mean difference; dotted lines: 95% confidence limits of agreement.