Alex S Willoughby1, Stephanie J Chiu1, Rachel K Silverman1, Sina Farsiu1, Clare Bailey2, Henry E Wiley3, Frederick L Ferris3, Glenn J Jaffe1.
Abstract
PURPOSE: We determine whether the automated segmentation software, Duke Optical Coherence Tomography Retinal Analysis Program (DOCTRAP), can measure, in a platform-independent manner, retinal thickness on Cirrus and Spectralis spectral domain optical coherence tomography (SD-OCT) images in eyes with diabetic macular edema (DME) under treatment in a clinical trial.Entities:
Keywords: diabetic macular edema; diabetic retinopathy; optical coherence tomography
Year: 2017 PMID: 28180033 PMCID: PMC5295813 DOI: 10.1167/tvst.6.1.9
Source DB: PubMed Journal: Transl Vis Sci Technol ISSN: 2164-2591 Impact factor: 3.283
Figure 1Example of DOCTRAP segmentation of ILM (white), inner RPE (pink), and BM (blue) boundaries in study eye with diabetic macular edema.
Figure 2ETDRS sector map of OD macula (mirror image for OS).
Figure 3Upper left: Spectralis software segmentation; automatic segmentation lines are placed on the ILM (top red line) and BM (bottom red line). Upper right: DOCTRAP software segmentation using Spectralis dataset; automatic segmentation lines are placed on the ILM (white line), inner one-third of RPE (pink line), and BM (blue line). Lower Left: Cirrus software segmentation; automatic segmentation are lines placed on the ILM (white line) and the inner one-third of the RPE (black line). Lower Right: DOCTRAP software segmentation using Cirrus dataset; automatic segmentation lines are placed on the ILM (white line), inner 1/3 of RPE (pink line), and BM (blue line).
Comparison of DOCTRAP and Commercial Software Segmentation
Figure 4Above: Cirrus software segmentation showing an uncorrected segmentation error in the inner one-third of the RPE (traced in red). Below: The segmentation of the inner 1/3 of the RPE (traced in red) after manual correction.
Retina Thickness Comparisons
Mean Paired Difference in Retinal Thickness as Measured from ILM to BM
Figure 5Bland-Altman plots of the differences versus the mean retinal thickness. Solid reference line indicates line of best fit (N = 30).
Mean Paired Difference in Retinal Thickness as Measured from ILM to Inner RPE Layer