Josef Veselka1, Pavla Cadová, Theodor Adla, David Zemánek.
Abstract
INTRODUCTION: The aim of this prospective study was to evaluate the diagnostic accuracy of dual-source computed tomography coronary angiography (CTCA) compared to intravascular ultrasound (IVUS) for the detection of restenosis in patients who underwent coronary stenting for bifurcation left main (LM) stenosis.Entities:
Keywords: computed tomography; intervention; left main stenosis; stent
Year: 2012 PMID: 22852000 PMCID: PMC3400902 DOI: 10.5114/aoms.2012.29220
Source DB: PubMed Journal: Arch Med Sci ISSN: 1734-1922 Impact factor: 3.318
Figure 1Assessment of a coronary artery stent by CTCA. A– Longitudinal view, B– minimal luminal area measurement
Baseline and procedural characteristics (n=24)
| Variable | Value |
|---|---|
| Age [years] (range) | 66±11 (46-88) |
| Men [%] | 58 |
| BMI [kg/m2] | 28±5 |
| Previous myocardial infarction [%] | 38 |
| Current smoker [%] | 21 |
| Hypertension [%] | 71 |
| Total plasma cholesterol [mmol/l; mg/dl] | 4.4 ±1.4; 170.1 ±54.1 |
| LDL cholesterol [mmol/l; mg/dl] | 2.3 ±1.0; 88.9 ±38.7 |
| HDL cholesterol [mmol/l; mg/dl] | 1.2 ±0.4; 46.4 ±15.5 |
| Plasma triglyceride [mmol/l; mg/dl] | 1.8 ±1.6; 159.4 ±141.7 |
| Diabetes mellitus [%] | 42 |
| β-Blockers [%] | 71 |
| Calcium channel blockers [%] | 25 |
| Angiotensin-converting inhibitors [%] | 92 |
| Clopidogrel [%] | 88 |
Diagnostic performance of CTCA in detecting significant stenosis
| Parameter | LM ( | LAD ( | LCx ( | All segments ( | All patients ( |
|---|---|---|---|---|---|
| Sensitivity (95% CI) | 100% (29.2–100) | 100% (15.8–100) | 75% (19.4–99.4) | 89% (51.8–99.7) | 86% (42.1–99.6) |
| Specificity (95% CI) | 74% (52.8–91.8) | 68% (43.5–87.4) | 54% (25.1–80.8) | 68% (53.7–80.1) | 41% (18.4–67.1) |
| Positive predictive value (95% CI) | 38% (8.5–75.5) | 25% (3.2–65.1) | 33% (7.5–70.1) | 32% (14.9–53.5) | 38% (15.2–64.6) |
| Negative predictive value (95% CI) | 100% (79.4–100) | 100% (75.3–100) | 88% (47.4–99.7) | 97% (85.8–99.9) | 88% (47.4–99.7) |
| Accuracy (95% CI) | 79% (62.7–95.3) | 71% (50.5–91.5) | 59% (35.6–82.4) | 71% (59.7–82.3) | 54% (34.1–73.9) |
| Likelihood ratio | 4.2 | 3.2 | 1.6 | 2.8 | 1.5 |
| Cut-off value [mm2] | <6.19 | <4.78 | <4.97 | <5.29 | |
| SE 46% | SE 67% | SE 75% | SE 64% | ||
| SP 88% | SP 90% | SP 65% | SP 77% | ||
| Area under curve | 0.6788 | 0.8204 | 0.7243 | 0.7293 |
SE – sensitivity, SP – specificity
Figure 2A– Correlation between IVUS and CTCA measurement of minimal luminal area in the stented left main coronary artery. B– Bland-Altman analysis of differences between the minimal luminal area measured by CTCA and IVUS
Figure 5A– Correlation between IVUS and CTCA measurement of minimal luminal area in all measured segments. B– Bland-Altman analysis of differences between the minimal luminal area measured by CTCA and IVUS
Figure 6Diagnostic value of CTCA measurement for assessment significant restenosis in all segments expressed by ROC curve