Literature DB >> 22833262

Benchtop evaluation of pressure barrier insufflator and standard insufflator systems.

Kenneth G Nepple1, Dorina Kallogjeri, Sam B Bhayani.   

Abstract

BACKGROUND: Previous experimental research has reported minimal differences in pressure maintenance between different versions of standard insufflators (SI). However, a recent report identified potential clinical benefits with a valveless pressure barrier insufflator (PBI). We sought to perform a benchtop objective evaluation of SI and PBI systems.
METHODS: A rigid box system with continuous pressure manometry was used to evaluate a PBI (Surgiquest Airseal) and two SIs (SI1 = Stryker PneumoSure High Flow Insufflator and SI2 = Storz SCB Thermoflator). Pressure maintenance of 15 mmHg was evaluated during experimental conditions of leakage from a 5 mm port site, leakage from a 12 mm port site, and continuous suction.
RESULTS: With leakage from the 5 mm port site, the PBI maintained pressure of >13 mmHg whereas the pressures dropped moderately with the SI1 (7-13 mmHg) and SI2 insufflators (3-7 mmHg) and did not regain goal pressure until leakage was stopped. With leakage from 12 mm port site, the PBI pressure decreased to 9-11 mmHg, whereas the SI1 and SI2 lost insufflation pressures completely. The PBI maintained pressure of >11 mmHg during continuous suction while the SI1 and SI2 lost pressure entirely, and actually showed negative pressure from air suction into the rigid box system. When evaluated statistically with the mixed model repeated measures ANOVA, the SI1 and SI2 performed similarly while the PBI maintained increased pressure.
CONCLUSIONS: In the experimental rigid box system, the PBI more successfully maintained pressure in response to leakage and suction than SIs.

Entities:  

Mesh:

Year:  2012        PMID: 22833262     DOI: 10.1007/s00464-012-2434-x

Source DB:  PubMed          Journal:  Surg Endosc        ISSN: 0930-2794            Impact factor:   4.584


  13 in total

1.  New technology and health care costs--the case of robot-assisted surgery.

Authors:  Gabriel I Barbash; Sherry A Glied
Journal:  N Engl J Med       Date:  2010-08-19       Impact factor: 91.245

Review 2.  Laparoscopic lens fogging: a review of etiology and methods to maintain a clear visual field.

Authors:  Nathan Lawrentschuk; Neil E Fleshner; Damien M Bolton
Journal:  J Endourol       Date:  2010-06       Impact factor: 2.942

3.  Twenty-five simple ways to increase insufflation performance and patient safety in laparoscopy.

Authors:  Volker R Jacobs; John E Morrison; Marion Kiechle
Journal:  J Am Assoc Gynecol Laparosc       Date:  2004-08

4.  Problems with technical equipment during laparoscopic surgery. An observational study.

Authors:  E G G Verdaasdonk; L P S Stassen; M van der Elst; T M Karsten; J Dankelman
Journal:  Surg Endosc       Date:  2006-11-21       Impact factor: 4.584

5.  Three-dimensional model for gas flow, resistance, and leakage-dependent nominal pressure maintenance of different laparoscopic insufflators.

Authors:  Volker R Jacobs; John E Morrison; Stefan Paepke; Thorsten Fischer; Marion Kiechle
Journal:  J Minim Invasive Gynecol       Date:  2006 May-Jun       Impact factor: 4.137

6.  Comparison of five different abdominal access trocar systems: analysis of insertion force, removal force, and defect size.

Authors:  D M Shafer; Y Khajanchee; J Wong; L L Swanström
Journal:  Surg Innov       Date:  2006-09       Impact factor: 2.058

7.  Use of the valveless trocar system reduces carbon dioxide absorption during laparoscopy when compared with standard trocars.

Authors:  Amin S Herati; Sero Andonian; Soroush Rais-Bahrami; Mohamed A Atalla; Arun K Srinivasan; Lee Richstone; Louis R Kavoussi
Journal:  Urology       Date:  2010-10-02       Impact factor: 2.649

Review 8.  Heated CO(2) with or without humidification for minimally invasive abdominal surgery.

Authors:  Daniel W Birch; Namdar Manouchehri; Xinzhe Shi; Ghassan Hadi; Shahzeer Karmali
Journal:  Cochrane Database Syst Rev       Date:  2011-01-19

9.  A new valve-less trocar for urologic laparoscopy: initial evaluation.

Authors:  Amin S Herati; Mohamed A Atalla; Soroush Rais-Bahrami; Sero Andonian; Manish A Vira; Louis R Kavoussi
Journal:  J Endourol       Date:  2009-09       Impact factor: 2.942

10.  Carbon dioxide gas heating inside laparoscopic insufflators has no effect.

Authors:  Volker R Jacobs; Marion Kiechle; John E Morrison
Journal:  JSLS       Date:  2005 Apr-Jun       Impact factor: 2.172

View more
  8 in total

1.  AirSeal system insufflator to maintain a stable pneumorectum during TAMIS.

Authors:  G Bislenghi; A M Wolthuis; A de Buck van Overstraeten; A D'Hoore
Journal:  Tech Coloproctol       Date:  2014-11-26       Impact factor: 3.781

Review 2.  Advances in laparoscopic surgery in urology.

Authors:  Jens J Rassweiler; Dogu Teber
Journal:  Nat Rev Urol       Date:  2016-05-24       Impact factor: 14.432

Review 3.  Anatomical Considerations and Procedure-Specific Aspects Important in Preventing Operative Morbidity during Transanal Total Mesorectal Excision.

Authors:  Sam Atallah
Journal:  Clin Colon Rectal Surg       Date:  2020-04-28

Review 4.  Clinical and Organizational Impact of the AIRSEAL® Insufflation System During Laparoscopic Surgery: A Systematic Review.

Authors:  David Balayssac; Marie Selvy; Anthony Martelin; Caroline Giroudon; Delphine Cabelguenne; Xavier Armoiry
Journal:  World J Surg       Date:  2020-11-30       Impact factor: 3.352

5.  Robotic partial nephrectomy performed with Airseal versus a standard CO2 pressure pneumoperitoneum insufflator: a prospective comparative study.

Authors:  Filippo Annino; Luca Topazio; Domenico Autieri; Tiziano Verdacchi; Michele De Angelis; Anastasios D Asimakopoulos
Journal:  Surg Endosc       Date:  2016-08-05       Impact factor: 4.584

6.  Operative vectors, anatomic distortion, fluid dynamics and the inherent effects of pneumatic insufflation encountered during transanal total mesorectal excision.

Authors:  S Atallah; P Gonzalez; S Chadi; R Hompes; J Knol
Journal:  Tech Coloproctol       Date:  2017-10-10       Impact factor: 3.781

Review 7.  Laparoscopy during the COVID-19 Pandemic: Absence of Evidence is not Evidence of Absence.

Authors:  Prabudh Goel; Ashoke K Basu
Journal:  J Indian Assoc Pediatr Surg       Date:  2020-06-24

8.  Intraoperative Measures to Reduce the Risk of COVID-19 Transmission During Minimally Invasive Procedures: A Systematic Review and Critical Appraisal of Societies' Recommendations.

Authors:  Giovanni A Tommaselli; Philippe Grange; Crystal D Ricketts; Jeffrey W Clymer; Raymond S Fryrear
Journal:  Surg Laparosc Endosc Percutan Tech       Date:  2021-07-28       Impact factor: 1.719

  8 in total