BACKGROUND AND OBJECTIVES: Within recent years, the insufflation technique for laparoscopy has become more important with high flow insufflators (> or = 30 L/min) and high gas turn over (> or = 800 L/procedure). Increased amounts of carbon dioxide (CO2) gas used can lead to laparoscopic hypothermia. We studied the insufflator with versus insufflators without internal gas heating (inside insufflator) as a sufficient method of hypothermia prevention at different flow rates. METHODS: With a computer-based data acquisition model, different standard insufflators with internal gas heating (Snowden Pencer) vs. without (Storz Endoflator, Storz Laparoflator, Richard Wolf, and BEI Medical) were compared regarding CO2 gas temperature at different points in the insufflation system (insufflator exit, insufflation hose end). RESULTS: Gas temperature of the Snowden Pencer insufflator, which is flow-rate dependent, increases at the exit (max. 35.4 degrees C). However, gas temperature is back to room temperature (-0.22 to +1.10 degrees C) at the end of the insufflation hose (10 ft or 3 m) for all 5 insufflators studied. Even at high gas flow rates (< or = 20 L/min), CO2 gas is at room temperature when it reaches the patient. DISCUSSION: No difference was noted regarding gas temperature between the insufflators compared. Insufflator internal gas heating, such as the Snowden Pencer insufflator, can not have a clinically significant effect because it is too far away from the patient to raise the gas temperature in the abdomen. Purchasers are misled because the gas-heating device has no measurable benefit for the patient.
BACKGROUND AND OBJECTIVES: Within recent years, the insufflation technique for laparoscopy has become more important with high flow insufflators (> or = 30 L/min) and high gas turn over (> or = 800 L/procedure). Increased amounts of carbon dioxide (CO2) gas used can lead to laparoscopic hypothermia. We studied the insufflator with versus insufflators without internal gas heating (inside insufflator) as a sufficient method of hypothermia prevention at different flow rates. METHODS: With a computer-based data acquisition model, different standard insufflators with internal gas heating (Snowden Pencer) vs. without (Storz Endoflator, Storz Laparoflator, Richard Wolf, and BEI Medical) were compared regarding CO2 gas temperature at different points in the insufflation system (insufflator exit, insufflation hose end). RESULTS: Gas temperature of the Snowden Pencer insufflator, which is flow-rate dependent, increases at the exit (max. 35.4 degrees C). However, gas temperature is back to room temperature (-0.22 to +1.10 degrees C) at the end of the insufflation hose (10 ft or 3 m) for all 5 insufflators studied. Even at high gas flow rates (< or = 20 L/min), CO2 gas is at room temperature when it reaches the patient. DISCUSSION: No difference was noted regarding gas temperature between the insufflators compared. Insufflator internal gas heating, such as the Snowden Pencer insufflator, can not have a clinically significant effect because it is too far away from the patient to raise the gas temperature in the abdomen. Purchasers are misled because the gas-heating device has no measurable benefit for the patient.
Although it had already been shown in 1870 that intraabdominal insufflation of CO2 gas can lead to hypothermia in a rabbit model[1] and a simple gas heating and hydration device for laparoscopy was used 4 decades ago (,[2] only within recent years has growing attention been given to the problem. With increasing amounts of up to several hundred liters of CO2 gas being used,[3] — our experience was a maximum of 801 L in 1 procedure—and high gas flow rates of ≥30 L/min, the effect and prevention of laparopelviscopic hypothermia needs to be addressed.[4]A simple gas heating and hydration device for laparoscopy named “Feuchte Kammer“ (wet chamber) with wet filter paper inside and warming light bulb above used by Siede and Schneider from Darmstadt, Germany, before 1962.[2]Intraoperative systemic hypothermia is linked to a higher intra- and postoperative complication rate (discomfort, pain, coagulopathy, morbid cardiac events, and other things). CO2 gas warming has been proven to reduce pain;[2,5,6] therefore, a closer look at the prevention methods offered is necessary.In general, 2 different methods are currently in use to prevent laparopelviscopic hypothermia: changing the gas state (heating, hydration, eg, Insuflow; Lexion Medical Inc., St. Paul, MN) before it reaches the patient, or prevention of systemic hypothermia with patient surface warming equipment (eg, a warm air fan like Bair Hugger, Arizant Healthcare Inc., Eden Prairie, MN; a water mattress like Blanketrol, Cincinnati Sub-Zero Products Inc., Cincinnati, OH, or fluid warmer, warming blankets, and others).Another method offered is an insufflator built-in CO2 gas heating device that heats up CO2 gas to body temperature, eg, inside insufflators like the Snowden Pencer (Snowden Pencer, Tucker, GA) or in the NuMo insufflator (distributed by EndoVentions International Inc., Des Plaines, IL).It is clinically obvious that the patient benefits from hypothermia prevention methods, but which technique shall the physician rely on? The physician is finally responsible for all equipment used on or around the patient, so from this background arise 2 questions: How efficient is insufflator internal gas heating, and do high gas flow rates affect the CO2 gas temperature? In the following, we compare both types of insufflators— one with vs. 4 without internal gas heating—at increasing flow rates in a laboratory measurement model.
METHODS
A notebook PC (TECRA 510 CDT; Toshiba Inc., Irvine, CA) and a PCMCIA data acquisition board (PCI-460-P1; Intelligent Instrumentation Inc., Tucson, AZ) were connected to electronic temperature meters (THERM 2280-1 with T-430-2R (K probes, NiCr-Ni); Ahlborn, Holzkirchen, Germany). Measurement accuracy was ≤±0.1°C. At 2 different points in the insufflation system (at the insufflator exit and at end of insufflation hose), CO2 gas temperature was measured with different standard insufflators with a maximum flow rate of 10 L/min to 20 L/min (BEI, Snowden Pencer, 2 x Storz, Richard Wolf) (. Gas flow was measured with a laminar flow element LFE type 1 (≤60 L/min) and an electronic differential meter Digima premo 720 (both Special Instruments, Noerdlingen, Germany). The measurement scheme ( was graphically created with an adjustable, multipurpose measurement program (Visual Designer 3.0; Intelligent Instrumentation Inc., Tucson, AZ) and transferred into the final program. Results are displayed with a scientific graphic program (Origin 4.1; Microcal Inc., Northampton, MA).Insufflators evaluated (top to bottom): Storz Endoflator; Storz Laparoflator; Snowden Pencer; BEI Medical; Richard Wolf.Measurement scheme for gas temperature evaluation of laparoscopic insufflators.
RESULTS
The results of laboratory measurements are described under 3 aspects:(1) Gas temperature at the exit of the insufflator (patient distal): The Snowden Pencer insufflator showed a gas flow rate-dependent increase in CO2 gas temperature at the exit of the insufflator up to a maximum of 35.4°C, which is reached only at the maximum gas flow rate of 15 L/min. At lower flow rates, the CO2 gas temperature is less, eg, 29.3°C at 2 L/min and 31.4 at 4 L/min. The measurements showed that body temperature of 37.0°C is neither reached nor exceeded (Figure 4).
Figure 4.
Gas flow-dependent CO2 temperature of the Snowden Pencer insufflator at insufflator exit (top curve) and at insufflation hose end (bottom curve).
Gas flow-dependent CO2 temperature of the Snowden Pencer insufflator at insufflator exit (top curve) and at insufflation hose end (bottom curve).(2) Gas temperature at the end of the insufflation hose (patient proximal): CO2 gas temperature is around OR room temperature (∼ 24°C) at the insufflation hose end when it reaches the patient. At the specific maximum gas flow rate, the range of temperature change from initial room temperature is only -0.24°C to + 1.10°C for all insufflators. Although the Snowden Pencer insufflator shows the largest increase of + 1.10°C above room temperature, an absolute value of 24.45°C is far from reaching body temperature (Figure 5). Minimal changes in operating room temperature caused by the air conditioning system can also be reflected in the CO2 gas temperature in the insufflation hose.
Figure 5.
Gas flow-dependent CO2 temperature for all insufflators at the insufflation hose end.
Gas flow-dependent CO2 temperature for all insufflators at the insufflation hose end.(3) CO2 temperature at high gas flow rates at the insufflation hose end (patient proximal): Despite increasing gas flow rates, the temperature measurement curves are almost linear for all insufflators (. Even at high gas flow rates with up to 20 L/min, the insufflation gas stays at room temperature at the hose end for all insufflators. Although only minor flow-dependent temperature changes are measurable, all insufflators have a gas temperature trend. Three insufflators show a CO2 gas temperature increase (Snowden Pencer + 1.10°C, BEI Medical + 0.86°C, and Richard Wolf + 0.22°C), and 2 show a minor decrease in gas temperature (Storz Laparoflator -0.24°C and Storz Endoflator -0.09°C) (. However, these effects are not clinically relevant.
Purchasers of insufflators with internal gas heating should be aware that such insufflators do not increase the intra-abdominal gas temperature sufficiently. It has been proven that these insufflators have no significant effect and therefore cannot and should not be relied on to prevent laparoscopic hypothermia.Manufacturers and developing industry are our partners in performing laparoscopy. But physicians have to carefully follow and critically evaluate new technology and instruments for actually available function for a real benefit for the patient.[18] This study proves that quality control studies for all laparoscopic equipment are necessary.[19]