Accuracy of aortic annular measurements obtained from three-dimensional echocardiography, CT and MRI: human in vitro and in vivo studies.

Wendy Tsang1, Michael G Bateman, Lynn Weinert, Gian Pellegrini, Victor Mor-Avi, Lissa Sugeng, Hubert Yeung, Amit R Patel, Alexander J Hill, Paul A Iaizzo, Roberto M Lang.   

Abstract

OBJECTIVES: To determine the accuracy of calcium-containing rings measurements imaged by three-dimensional echocardiography (3DE), multi-slice CT (MSCT) and cardiac magnetic resonance (CMR) under ideal conditions against the true ring dimensions. To compare the accuracy of aortic annulus (AoA) measurements in ex vivo human hearts using 3DE, MSCT and CMR. To determine the accuracy of AoA measurements in an in vivo human model.
DESIGN: 3DE, MSCT and CMR imaging were performed on 30 calcium-containing rings and 28 explanted human hearts. Additionally, 15 human subjects with clinical indication for MSCT underwent 3DE. Two experts in each modality measured the images. MAIN OUTCOME MEASURES: Bias and intraclass correlation coefficient for accuracy of imaging measurements when compared with actual ring dimensions. Bias, intraclass correlation coefficient and variability were obtained: (1) when comparing explanted human heart AoA measurements from the two remaining imaging modalities with the most accurate one as determined from the ring measurements and (2) in in vivo human AoA measurements. Analysis was repeated on explanted heart subgroups divided by aortic valve Agatston score.
RESULTS: Against the known ring dimensions, CMR had the highest accuracy and the lowest variability. MSCT measurements had high accuracy but wider variability and 3DE had the lowest accuracy with the largest variability. When 3DE and MSCT were compared with CMR, 3DE underestimated and MSCT overestimated AoA dimensions, but inter-measurement variability of 3DE and MSCT were similar. When divided by Agatston score, both 3DE and MSCT measurements were larger and showed greater variability with increasing calcium burden. The in vivo study showed that the correlation between 3DE and MSCT measurements was high; however, 3DE measurements were smaller than those measured with MSCT.
CONCLUSIONS: In the in vitro model, CMR measurements were the most accurate for assessing the actual dimensions suggesting that further investigations on its role in AoA measurement in TAVR are needed. However from the in vivo model, MSCT and 3DE are reasonable alternatives with the understanding that they can slightly overestimate and underestimate annular dimensions, respectively.

Entities:  

Mesh:

Year:  2012        PMID: 22773684     DOI: 10.1136/heartjnl-2012-302074

Source DB:  PubMed          Journal:  Heart        ISSN: 1355-6037            Impact factor:   5.994


  24 in total

Review 1.  Guidance of transcatheter aortic valve replacement by echocardiography.

Authors:  Rebecca T Hahn
Journal:  Curr Cardiol Rep       Date:  2014-01       Impact factor: 2.931

2.  A non-contrast self-navigated 3-dimensional MR technique for aortic root and vascular access route assessment in the context of transcatheter aortic valve replacement: proof of concept.

Authors:  Matthias Renker; Akos Varga-Szemes; U Joseph Schoepf; Stefan Baumann; Davide Piccini; Michael O Zenge; Wolfgang G Rehwald; Edgar Müller; Jeremy D Rier; Helge Möllmann; Christian W Hamm; Daniel H Steinberg; Carlo N De Cecco
Journal:  Eur Radiol       Date:  2015-07-20       Impact factor: 5.315

3.  Multi-parametric approach to predict prosthetic valve size using CMR and clinical data: insights from SAVR.

Authors:  Federico E Mordini; Conor F Hynes; Richard L Amdur; Jeffrey Panting; Dominic A Emerson; Jason Morrissette; Erin Goheen-Thomas; Michael D Greenberg; Gregory D Trachiotis
Journal:  Int J Cardiovasc Imaging       Date:  2021-03-10       Impact factor: 2.357

4.  Iterative reconstruction in single-source dual-energy CT angiography: feasibility of low and ultra-low volume contrast medium protocols.

Authors:  Ping Hou; Xiangnan Feng; Jie Liu; Yue Zhou; Yaojun Jiang; Xiaochen Jiang; Jianbo Gao
Journal:  Br J Radiol       Date:  2017-06-23       Impact factor: 3.039

Review 5.  Paravalvular regurgitation following transcutaneous aortic valve replacement: predictors and clinical significance.

Authors:  Rebecca T Hahn; Susheel Kodali; Philippe Généreux; Martin Leon
Journal:  Curr Cardiol Rep       Date:  2014-05       Impact factor: 2.931

Review 6.  Aortic annulus sizing in stenotic bicommissural non-raphe-type bicuspid aortic valves: reconstructing a three-dimensional structure using only two hinge points.

Authors:  Antonio H Frangieh; Jonathan Michel; Oliver Deutsch; Michael Joner; Costanza Pellegrini; Tobias Rheude; Sabine Bleiziffer; Albert Markus Kasel
Journal:  Clin Res Cardiol       Date:  2018-06-14       Impact factor: 5.460

7.  Modern Use of Echocardiography in Transcatheter Aortic Valve Replacement: an Up-Date.

Authors:  Cristina Caldararu; Serban Balanescu
Journal:  Maedica (Bucur)       Date:  2016-12

Review 8.  Imaging in the context of replacement heart valve development: use of the Visible Heart(®) methodologies.

Authors:  Michael G Bateman; Paul A Iaizzo
Journal:  Cardiovasc Diagn Ther       Date:  2012-09

9.  The clinical anatomy and pathology of the human arterial valves: implications for repair or replacement.

Authors:  Michael G Bateman; Alexander J Hill; Jason L Quill; Paul A Iaizzo
Journal:  J Cardiovasc Transl Res       Date:  2013-01-17       Impact factor: 4.132

10.  Contrast volume reduction using third generation dual source computed tomography for the evaluation of patients prior to transcatheter aortic valve implantation.

Authors:  Daniel O Bittner; Martin Arnold; Lutz Klinghammer; Annika Schuhbaeck; Michaela M Hell; Gerd Muschiol; Soeren Gauss; Michael Lell; Michael Uder; Udo Hoffmann; Stephan Achenbach; Mohamed Marwan
Journal:  Eur Radiol       Date:  2016-03-19       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.