| Literature DB >> 22766951 |
E Ghasroddashti1, W L Smith, S Quirk, C Kirkby.
Abstract
Changing pulse repetition frequency or dose rate used for IMRT treatments can alter the number of monitor units (MUs) and the time required to deliver a plan. This work was done to develop a practical picture of the magnitude of these changes. We used Varian's Eclipse Treatment Planning System to calculate the number of MUs and beam-on times for a total of 40 different treatment plans across an array of common IMRT sites including prostate/pelvis, prostate bed, head and neck, and central nervous system cancers using dose rates of 300, 400 and 600 MU/min. In general, we observed a 4%-7% increase in the number of MUs delivered and a 10-40 second decrease in the beam-on time for each 100 MU/min of dose rate increase. The increase in the number of MUs resulted in a reduction of the "beam-on time saved". The exact magnitude of the changes depended on treatment site and planning target volume. These changes can lead to minor, but not negligible, concerns with respect to radiation protection and treatment planning. Although the number of MUs increased more rapidly for more complex treatment plans, the absolute beam-on time savings was greater for these plans because of the higher total number of MUs required to deliver them. We estimate that increasing the IMRT dose rate from 300 to 600 MU/min has the potential to add up to two treatment slots per day for each IMRT linear accelerator. These results will be of value to anyone considering general changes to IMRT dose rates within their clinic.Entities:
Mesh:
Year: 2012 PMID: 22766951 PMCID: PMC5716514 DOI: 10.1120/jacmp.v13i4.3810
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Properties of the MLC system used in this work.
|
|
|
|---|---|
| MLC Model | Varian Millennium 120 |
| Maximum leaf span | 15.0 cm |
| Maximum leaf speed | 2.5 cm/s |
| Dosimetric leaf gap | 0.15 cm |
| Interleaf leakage | 1.5 % |
| Leaf transmission | 2.5 % |
| Controller Software Version | 7.4.1.6 |
A summary of the results across all sites for dose rates (in MU/min). Presented are the mean and standard deviation across the ten plans considered.
|
|
|
|
|
|
|
| |
|---|---|---|---|---|---|---|---|
| Prostate Pelvis Ph1 | mean | 5.38 | 5.76 | 6.58 | 3.41 | 2.74 | 2.08 |
| std | 1.01 | 1.09 | 1.31 | 0.58 | 0.47 | 0.38 | |
| Prostate Pelvis Ph2 | mean | 2.85 | 2.95 | 3.19 | 1.99 | 1.55 | 1.12 |
| std | 0.49 | 0.50 | 0.54 | 0.34 | 0.26 | 0.19 | |
| Prostate Bed | mean | 3.26 | 3.41 | 3.75 | 2.28 | 1.79 | 1.31 |
| std | 0.84 | 0.89 | 1.00 | 0.49 | 0.38 | 0.28 | |
| Head & Neck | mean | 4.92 | 5.29 | 5.94 | 3.70 | 2.98 | 2.23 |
| std | 0.75 | 0.82 | 0.91 | 0.56 | 0.46 | 0.34 | |
| Central Nervous | mean | 2.86 | 3.03 | 3.36 | 1.98 | 1.57 | 1.16 |
| std | 0.59 | 0.67 | 0.78 | 0.33 | 0.28 | 0.22 |
Figure 1The mean trends in relative number of MU (normalized to the doe rate) as a function of dose rate for all sites investigated. The maximum standard deviation in the rate was 0.04 for both the phase one prostate/ pelvis and the brain cases.
Figure 2The mean trends in beam‐on time saved (as a difference from the doe rate) as a function of dose rate for all sites investigated. The trend is not linear because, as the dose rate increases, more MUs are required to deliver the same fluence pattern.
Figure 3The percentage increase in MU per 100MU/min increase in dose rate as a function of PTV volume in cc for the prostate cases. The black line is the linear regression fit to the data. It increases at a rate of ~ 0.36% per 100 cc of volume increase.
Figure 4The measured pass rates (mean across all fields, error bars indicate one standard deviation) for a subset of plans meeting a 3%/ 3 mm acceptance criteria. A 95% pass rate is considered clinically acceptable. This demonstrates the plans are delivering dose maps clinically consistent with the treatment planning system prediction generated using a dose rate of 300MU/min.