Corrosion at the neck-stem junction as a cause of metal ion release and pseudotumour formation.

I P S Gill1, J Webb, K Sloan, R J Beaver.   

Abstract

We present a series of 35 patients (19 men and 16 women) with a mean age of 64 years (36.7 to 75.9), who underwent total hip replacement using the ESKA dual-modular short stem with metal on-polyethylene bearing surfaces. This implant has a modular neck section in addition to the modular head. Of these patients, three presented with increasing post-operative pain due to pseudotumour formation that resulted from corrosion at the modular neck-stem junction. These patients underwent further surgery and aseptic lymphocytic vaculitis associated lesions were demonstrated on histological analysis. Retrieval analysis of two modular necks showed corrosion at the neck-stem taper. Blood cobalt and chromium levels were measured at a mean of nine months (3 to 28) following surgery. These were compared with the levels in seven control patients (three men and four women) with a mean age of 53.4 years (32.1 to 64.1), who had an identical prosthesis and articulation but with a prosthesis that had no modularity at neck-stem junction. The mean blood levels of cobalt in the study group were raised at 50.75 nmol/l (5 to 145) compared with 5.6 nmol/l (2 to 13) in control patients. Corrosion at neck-stem tapers has been identified as an important source of metal ion release and pseudotumour formation requiring revision surgery. Finite element modelling of the dual modular stem demonstrated high stresses at the modular stem-neck junction. Dual modular cobalt-chrome hip prostheses should be used with caution due to these concerns.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22733942     DOI: 10.1302/0301-620X.94B7.29122

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  63 in total

1.  CORR Insights(®): Otto Aufranc Award: Large Heads Do Not Increase Damage at the Head-neck Taper of Metal-on-polyethylene Total Hip Arthroplasties.

Authors:  Philip C Noble
Journal:  Clin Orthop Relat Res       Date:  2015-08-21       Impact factor: 4.176

2.  Is increased modularity associated with increased fretting and corrosion damage in metal-on-metal total hip arthroplasty devices?: a retrieval study.

Authors:  Genymphas B Higgs; Josa A Hanzlik; Daniel W MacDonald; Jeremy L Gilbert; Clare M Rimnac; Steven M Kurtz
Journal:  J Arthroplasty       Date:  2013-08-01       Impact factor: 4.757

3.  Cementless modular neck stems: are they a safe option in primary total hip arthroplasty?

Authors:  José M Pelayo-de-Tomás; José L Rodrigo-Pérez; Carlos D Novoa-Parra; Alejandro Lizaur-Utrilla; María Morales-Suárez-Varela; José Antonio Blas-Dobón
Journal:  Eur J Orthop Surg Traumatol       Date:  2017-10-29

4.  Development and first validation of a simplified CT-based classification system of soft tissue changes in large-head metal-on-metal total hip replacement: intra- and interrater reliability and association with revision rates in a uniform cohort of 664 arthroplasties.

Authors:  Martijn F Boomsma; Mireille A Edens; Christiaan P Van Lingen; Niek Warringa; Harmen B Ettema; Cees C P M Verheyen; Mario Maas
Journal:  Skeletal Radiol       Date:  2015-05-06       Impact factor: 2.199

Review 5.  Metallic debris from metal-on-metal total hip arthroplasty regulates periprosthetic tissues.

Authors:  Christoph H Lohmann; Gurpal Singh; Hans-Georg Willert; Gottfried H Buchhorn
Journal:  World J Orthop       Date:  2014-11-18

6.  CORR Insights®: Exchangeable Femoral Neck (Dual-Modular) THA Prostheses Have Poorer Survivorship than Other Designs: A Nationwide Cohort of 324,108 Patients.

Authors:  H John Cooper
Journal:  Clin Orthop Relat Res       Date:  2017-02-24       Impact factor: 4.176

7.  In-situ Generated Tribomaterial in Metal/Metal Contacts: current understanding and future implications for implants.

Authors:  N Espallargas; A Fischer; A Igual Muñoz; S Mischler; M A Wimmer
Journal:  Biotribology (Oxf)       Date:  2017-05-12

8.  [Objectives and limits of test standards].

Authors:  C Kaddick; W Blömer
Journal:  Orthopade       Date:  2014-06       Impact factor: 1.087

Review 9.  Revision total hip arthroplasty for metal-on-metal failure.

Authors:  Justin S Chang; Fares S Haddad
Journal:  J Clin Orthop Trauma       Date:  2019-10-04

10.  Adverse local tissue reaction arising from corrosion at the femoral neck-body junction in a dual-taper stem with a cobalt-chromium modular neck.

Authors:  H John Cooper; Robert M Urban; Richard L Wixson; R Michael Meneghini; Joshua J Jacobs
Journal:  J Bone Joint Surg Am       Date:  2013-05-15       Impact factor: 5.284

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.