| Literature DB >> 22718870 |
Jung Woo Choe1, Ömer Oralkan, Amin Nikoozadeh, Mustafa Gencel, Douglas N Stephens, Matthew O'Donnell, David J Sahn, Butrus T Khuri-Yakub.
Abstract
A ring array provides a very suitable geometry for forward-looking volumetric intracardiac and intravascular ultrasound imaging. We fabricated an annular 64-element capacitive micromachined ultrasonic transducer (CMUT) array featuring a 10-MHz operating frequency and a 1.27-mm outer radius. A custom software suite was developed to run on a PC-based imaging system for real-time imaging using this device. This paper presents simulated and experimental imaging results for the described CMUT ring array. Three different imaging methods--flash, classic phased array (CPA), and synthetic phased array (SPA)--were used in the study. For SPA imaging, two techniques to improve the image quality--Hadamard coding and aperture weighting--were also applied. The results show that SPA with Hadamard coding and aperture weighting is a good option for ring-array imaging. Compared with CPA, it achieves better image resolution and comparable signal-to-noise ratio at a much faster image acquisition rate. Using this method, a fast frame rate of up to 463 volumes per second is achievable if limited only by the ultrasound time of flight; with the described system we reconstructed three cross-sectional images in real-time at 10 frames per second, which was limited by the computation time in synthetic beamforming.Entities:
Mesh:
Year: 2012 PMID: 22718870 PMCID: PMC3415797 DOI: 10.1109/TUFFC.2012.2310
Source DB: PubMed Journal: IEEE Trans Ultrason Ferroelectr Freq Control ISSN: 0885-3010 Impact factor: 2.725