Literature DB >> 17328324

Parallel beamforming using synthetic transmit beams.

Torbjørn Hergum1, Tore Bjåstad, Kjell Kristoffersen, Hans Torp.   

Abstract

Parallel beamforming is frequently used to increase the acquisition rate of medical ultrasound imaging. However, such imaging systems will not be spatially shift invariant due to significant variation across adjacent beams. This paper investigates a few methods of parallel beam-forming that aims at eliminating this flaw and restoring the shift invariance property. The beam-to-beam variations occur because the transmit and receive beams are not aligned. The underlying idea of the main method presented here is to generate additional synthetic transmit beams (STB) through interpolation of the received, unfocused signal at each array element prior to beamforming. Now each of the parallel receive beams can be aligned perfectly with a transmit beam--synthetic or real--thus eliminating the distortion caused by misalignment. The proposed method was compared to the other compensation methods through a simulation study based on the ultrasound simulation software Field II. The results have been verified with in vitro experiments. The simulations were done with parameters similar to a standard cardiac examination with two parallel receive beams and a transmit-line spacing corresponding to the Rayleigh criterion, wavelength times f-number (lambda x f#). From the results presented, it is clear that straightforward parallel beamforming reduces the spatial shift invariance property of an ultrasound imaging system. The proposed method of using synthetic transmit beams seems to restore this important property, enabling higher acquisition rates without loss of image quality.

Mesh:

Year:  2007        PMID: 17328324     DOI: 10.1109/tuffc.2007.241

Source DB:  PubMed          Journal:  IEEE Trans Ultrason Ferroelectr Freq Control        ISSN: 0885-3010            Impact factor:   2.725


  1 in total

1.  Volumetric real-time imaging using a CMUT ring array.

Authors:  Jung Woo Choe; Ömer Oralkan; Amin Nikoozadeh; Mustafa Gencel; Douglas N Stephens; Matthew O'Donnell; David J Sahn; Butrus T Khuri-Yakub
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2012-06       Impact factor: 2.725

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.