| Literature DB >> 22718400 |
Kentaro Hayashida1, Erik Bouvier, Thierry Lefèvre, Thomas Hovasse, Marie-Claude Morice, Bernard Chevalier, Mauro Romano, Philippe Garot, Arnaud Farge, Patrick Donzeau-Gouge, Bertrand Cormier.
Abstract
Although annulus rupture is one of the most severe complications of transcatheter aortic valve implantation (TAVI), the incidence and mechanism of this complication remain unclear. Out of 387 consecutive TAVI cases in our institution, the incidence of annulus rupture was 1.0% (4/387). The first two patients died because of hemodynamic collapse due to tamponade on day 0. Both surviving patients had undergone preprocedural multidetector computed tomography which revealed large calcifications in the epicardial fat part of the aortic annulus. In both cases, annulus rupture occurred after deployment of a balloon expandable valve suggesting that mechanical compression of this "vulnerable area" by calcification may cause annulus rupture.Entities:
Keywords: annular geometry; aortic stenosis; cardiac tamponade; computed tomography; transcatheter aortic valve
Mesh:
Year: 2013 PMID: 22718400 DOI: 10.1002/ccd.24524
Source DB: PubMed Journal: Catheter Cardiovasc Interv ISSN: 1522-1946 Impact factor: 2.692