Central corneal thickness measurements using Orbscan II scanning slit topography, noncontact specular microscopy, and ultrasonic pachymetry in eyes with keratoconus.

Keisuke Kawana1, Kazunori Miyata, Tadatoshi Tokunaga, Takahiro Kiuchi, Takahiro Hiraoka, Tetsuro Oshika.   

Abstract

PURPOSE: To compare corneal thickness measurements using Orbscan II scanning slit topography, Topcon SP-2000P noncontact specular microscopy, and ultrasonic pachymetry in eyes with keratoconus.
METHODS: Central corneal thickness was measured in 22 eyes with keratoconus. Eyes with apparent corneal opacity were excluded. Scanning slit topography, noncontact specular microscopy, and ultrasonic pachymetry were used in this sequence. The acoustic equivalent correlation factor (0.92) was used for Orbscan readings.
RESULTS: Three devices gave significantly different corneal thickness readings (P < 0.001, repeated-measure analysis of variance). Measurements with Orbscan scanning slit topography (449.5 +/- 43.2 [SD] mum) were significantly smaller than those of ultrasonic pachymetry (485.0 +/- 29.3 microm; P < 0.001, Tukey multiple comparison) and SP-2000P noncontact specular microscopy (476.7 +/- 28.3 microm; P = 0.002). There were significant linear correlations between ultrasonic pachymetry and scanning slit topography (Pearson correlation coefficient r = 0.741, P < 0.001), between scanning slit topography and noncontact specular microscopy (r = 0.880, P < 0.001), and between noncontact specular microscopy and ultrasonic pachymetry (r = 0.811, P < 0.001).
CONCLUSION: In eyes with keratoconus, Orbscan II scanning slit topography system gave significantly smaller corneal thickness readings than the other 2 devices. Measurements taken by noncontact specular microscopy and ultrasonic pachymetry were comparable. Three devices showed significant linear correlations with one another.

Entities:  

Mesh:

Year:  2005        PMID: 16227843     DOI: 10.1097/01.ico.0000159733.37554.ba

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  8 in total

1.  Assessment of total corneal power after myopic corneal refractive surgery in Chinese eyes.

Authors:  Yanjun Hua; Chao Pan; Qinmei Wang
Journal:  Int Ophthalmol       Date:  2019-03-01       Impact factor: 2.031

2.  Epithelial, stromal, and total corneal thickness in keratoconus: three-dimensional display with artemis very-high frequency digital ultrasound.

Authors:  Dan Z Reinstein; Marine Gobbe; Timothy J Archer; Ronald H Silverman; D Jackson Coleman
Journal:  J Refract Surg       Date:  2010-04-07       Impact factor: 3.573

3.  Central corneal thickness measured by the Orbscan II system, contact ultrasound pachymetry, and the Artemis 2 system.

Authors:  Tania Paul; Mira Lim; Christopher E Starr; Harriet O Lloyd; D Jackson Coleman; Ronald H Silverman
Journal:  J Cataract Refract Surg       Date:  2008-11       Impact factor: 3.351

4.  Repeatability and interobserver reproducibility of Artemis-2 high-frequency ultrasound in determination of human corneal thickness.

Authors:  Kelechi C Ogbuehi; Uchechukwu L Osuagwu
Journal:  Clin Ophthalmol       Date:  2012-05-21

5.  Measurement of Central Corneal Thickness Using Ultrasound Pachymetry and Orbscan II in Normal Eyes.

Authors:  Mohammad Mehdi Sadoughi; Bahram Einollahi; Neda Einollahi; Javad Rezaei; Danial Roshandel; Sepehr Feizi
Journal:  J Ophthalmic Vis Res       Date:  2015 Jan-Mar

6.  A contralateral eye study comparing characteristics of corneal endothelial cells in bilateral keratoconus patients with unilateral corneal Vogt's striae.

Authors:  Hadi Ostadi-Moghaddam; Mohamad-Reza Sedaghat; Tahereh Rakhshandadi; Sattar Rajabi; Foroozan Narooie-Noori; Farshad Askarizadeh
Journal:  J Curr Ophthalmol       Date:  2018-03-03

7.  Quantitative OCT-based corneal topography in keratoconus with intracorneal ring segments.

Authors:  Sergio Ortiz; Pablo Pérez-Merino; Nicolas Alejandre; E Gambra; I Jimenez-Alfaro; Susana Marcos
Journal:  Biomed Opt Express       Date:  2012-04-02       Impact factor: 3.732

8.  Central Corneal Thickness Measurement Using Ultrasonic Pachymetry, Rotating Scheimpflug Camera, and Scanning-slit Topography Exclusively in Thin Non-keratoconic Corneas.

Authors:  Mehrdad Mohammadpour; Kazem Mohammad; Nasser Karimi
Journal:  J Ophthalmic Vis Res       Date:  2016 Jul-Sep
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.