Rebound, applanation, and dynamic contour tonometry in pathologic corneas.

André Rosentreter1, Apostolos Athanasopoulos, Andrea M Schild, Alexandra Lappas, Claus Cursiefen, Thomas S Dietlein.   

Abstract

PURPOSE: To compare the practicability of using an Icare rebound tonometer (RT) versus a Goldmann applanation tonometer (GAT) or a Pascal dynamic contour tonometer (DCT) for measuring intraocular pressure (IOP) in patients with corneal abnormalities and, furthermore, to study the influence of central corneal thickness, corneal diameter, corneal radius, and axial length.
METHODS: One hundred seventy-one pathologic eyes with corneal abnormalities and 26 nonpathologic control eyes of 99 patients were included. Pathologic corneas were divided into subgroups: previous keratoplasty, keratoconus, corneal scars, corneal dystrophies, and bullous keratopathy.
RESULTS: Although IOP was successfully measured using the RT in all pathologic eyes, successful measurement of IOP was only possible in 98.2% when using the GAT and in 73.1% with the DCT. Mean IOP for all enrolled eyes was 12.7 ± 4.1 mm Hg for RT, 15.5 ± 4.4 mm Hg for GAT, and 16.3 ± 4.1 mm Hg for DCT. The mean difference between RT and GAT was ≤ 1 mm Hg (≤ 2 mm Hg) [≤ 3 mm Hg] in 23.4% (41.8%) [62.0%] of cases. Correlation analysis showed a moderate correlation between RT and GAT (r = 0.566; P < 0.001) and between RT and DCT (r = 0.364; P < 0.001). Bland-Altman analysis revealed a bias between RT and GAT and between RT and DCT of -2.8 and -3.8 mm Hg, with limits of agreement of -10.5 to 4.9 mm Hg and -12.2 to 4.6 mm Hg, respectively.
CONCLUSION: In pathologic corneas, IOP was difficult to obtain with GAT and DCT, whereas RT was able to determine IOP in all pathologic corneas. RT significantly underestimated IOP in all groups in relation to GAT and DCT. The agreement between the methods was clinically acceptable in corneal dystrophy and keratoconus but poor in eyes after keratoplasty.

Entities:  

Mesh:

Year:  2013        PMID: 22673854     DOI: 10.1097/ICO.0b013e318254a3fb

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  12 in total

1.  Impact of corneal parameters on intraocular pressure measurements in different tonometry methods.

Authors:  Aleksandra Zakrzewska; Marta P Wiącek; Anna Machalińska
Journal:  Int J Ophthalmol       Date:  2019-12-18       Impact factor: 1.779

2.  The influence of corneal geometrical and biomechanical properties on tonometry readings in keratoconic eyes.

Authors:  Mustafa Değer Bilgeç; Eray Atalay; Ömer Sözer; Hüseyin Gürsoy; Muzaffer Bilgin; Nilgün Yıldırım
Journal:  Int Ophthalmol       Date:  2019-12-02       Impact factor: 2.031

3.  Comparison of Diaton transpalpebral tonometer with applanation tonometry in keratoconus.

Authors:  Robert Pl Wisse; Natalie Peeters; Saskia M Imhof; Allegonda van der Lelij
Journal:  Int J Ophthalmol       Date:  2016-03-18       Impact factor: 1.779

4.  [Ease of handling of first and second generation rebound tonometers].

Authors:  N Mihailovic; J Termühlen; M Alnawaiseh; N Eter; T S Dietlein; A Rosentreter
Journal:  Ophthalmologe       Date:  2016-04       Impact factor: 1.059

5.  Twenty-four hours intraocular pressure in keratoconic eyes assessed by applanation tonometry and Tono-Pen AVIA.

Authors:  Rafael Vidal Merola; Sebastião Cronemberger; Artur William Veloso; Alberto Diniz-Filho
Journal:  Int J Ophthalmol       Date:  2022-01-18       Impact factor: 1.779

6.  Comprasion of ICare rebound tonometer and Goldmann applanation tonometer in high myopia.

Authors:  Halil Huseyin Cagatay; Metin Ekinci; Zeliha Yazar; Gokcen Gokce; Erdinc Ceylan
Journal:  ScientificWorldJournal       Date:  2014-10-14

7.  Evaluation of Goldmann applanation tonometry, rebound tonometry and dynamic contour tonometry in keratoconus.

Authors:  Fatih Özcura; Nilgün Yıldırım; Emre Tambova; Afsun Şahin
Journal:  J Optom       Date:  2016-07-09

8.  Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation.

Authors:  Mahmoud Rateb; Mahmoud Abdel-Radi; Zeiad Eldaly; Mohamed Nagy Elmohamady; Asaad Noor El Din
Journal:  J Ophthalmol       Date:  2019-05-09       Impact factor: 1.909

9.  A cross-sectional study to compare intraocular pressure measurement by sequential use of Goldman applanation tonometry, dynamic contour tonometry, ocular response analyzer, and Corvis ST.

Authors:  Sushma Tejwani; Shoruba Dinakaran; Anuja Joshi; Rohit Shetty; Abhijit Sinha Roy
Journal:  Indian J Ophthalmol       Date:  2015-11       Impact factor: 1.848

10.  Comparison of intraocular pressure measurements using Goldmann tonometer, I-care pro, Tonopen XL, and Schiotz tonometer in patients after Descemet stripping endothelial keratoplasty.

Authors:  Oded Ohana; David Varssano; Gabi Shemesh
Journal:  Indian J Ophthalmol       Date:  2017-07       Impact factor: 1.848

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.