Comparison of left ventricular stroke volume assessment by two- and three-dimensional echocardiography in a swine model of acute myocardial infarction validated by thermodilution method.

Yuichi J Shimada1, Kiyotake Ishikawa, Yoshiaki Kawase, Dennis Ladage, Lisa Tilemann, Takahiro Shiota, Roger J Hajjar.   

Abstract

BACKGROUND: Accurate left ventricular stroke volume (LVSV) measurement is clinically important in patients presenting with acute myocardial infarction. Three-dimensional echocardiography (3DE) is expected to overcome limitations of two-dimensional echocardiography (2DE). However, inaccuracy in volumetry by 3DE has often been reported hindering further clinical application. This study aimed at comparing agreement and correlation with the thermodilution method (TDM) between 2DE and 3DE measurement of LVSV.
METHODS: Swine model of myocardial infarction was created and LVSV was measured by 3DE by subtracting end-systolic from end-diastolic volume (3DE-method). Pulsed Doppler ultrasound and left ventricular outlet tract area were used to measure LVSV by 2DE (2DE-method). TDM was performed by the Swan-Ganz catheter. Bland-Altman analysis followed by assessment of intraclass correlation coefficient (ICC) were performed between 2DE-method and TDM as well as 3DE-method and TDM.
RESULTS: A total of 25 comparisons revealed a significant overestimation of LVSV by the 2DE-method (bias = 6.5 mL; 95% confidence interval [CI], 3.9-9.0 mL; P < 0.0001), whereas there was no significant bias by the 3DE-method (bias =-1.6; 95% CI, -4.3 to 1.1 mL; P = 0.22). The ICC between 2DE and TDM was 0.49 (95% CI, 0.14-0.74) whereas ICC between 3DE and TDM was 0.75 (95% CI, 0.51-0.88).
CONCLUSIONS: This study elucidated that LVSV is better estimated by 3DE-method compared to the conventional 2DE-method. This investigation will provide a more accurate, quick and noninvasive way of LVSV and cardiac output assessment at bedside by further application of 3DE.
© 2012, Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2012        PMID: 22672366     DOI: 10.1111/j.1540-8175.2012.01753.x

Source DB:  PubMed          Journal:  Echocardiography        ISSN: 0742-2822            Impact factor:   1.724


  5 in total

1.  Accuracy of the ClearSight™ system in patients undergoing abdominal aortic aneurysm surgery.

Authors:  Miho Sumiyoshi; Takuma Maeda; Erika Miyazaki; Naoshi Hotta; Hitoshi Sato; Eisuke Hamaguchi; Hiroko Kanazawa; Yoshihiko Ohnishi; Masataka Kamei
Journal:  J Anesth       Date:  2019-03-23       Impact factor: 2.078

2.  Accuracy and trending ability of the fourth-generation FloTrac/Vigileo System™ in patients undergoing abdominal aortic aneurysm surgery.

Authors:  Takuma Maeda; Kohshi Hattori; Miho Sumiyoshi; Hiroko Kanazawa; Yoshihiko Ohnishi
Journal:  J Anesth       Date:  2018-04-03       Impact factor: 2.078

3.  Passive Self Resonant Skin Patch Sensor to Monitor Cardiac Intraventricular Stroke Volume Using Electromagnetic Properties of Blood.

Authors:  Fayez Alruwaili; Kim Cluff; Jacob Griffith; Hussam Farhoud
Journal:  IEEE J Transl Eng Health Med       Date:  2018-09-26       Impact factor: 3.316

4.  Left Ventricular Unloading Using an Impella CP Improves Coronary Flow and Infarct Zone Perfusion in Ischemic Heart Failure.

Authors:  Shin Watanabe; Kenneth Fish; Jason C Kovacic; Olympia Bikou; Lauren Leonardson; Koichi Nomoto; Jaume Aguero; Navin K Kapur; Roger J Hajjar; Kiyotake Ishikawa
Journal:  J Am Heart Assoc       Date:  2018-03-07       Impact factor: 5.501

5.  Transesophageal echocardiography in swine: evaluation of left and right ventricular structure, function and myocardial work.

Authors:  Sebastian Billig; Rashad Zayat; Heike Schnöring; Matthias Derwall; Andreas Ebeling; Henning Steffen; Christoph Nix; Nima Hatam
Journal:  Int J Cardiovasc Imaging       Date:  2020-10-13       Impact factor: 2.357

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.