Abstract

UNLABELLED: Present attenuation-correction algorithms in whole-body PET/MRI do not consider variations in lung density, either within or between patients; this may adversely affect accurate quantification. In this work, a technique to incorporate patient-specific lung density information into MRI-based attenuation maps is developed and compared with an approach that assumes uniform lung density.
METHODS: Five beagles were scanned with (18)F-FDG PET/CT and MRI. The relationship between MRI and CT signal in the lungs was established, allowing the prediction of attenuation coefficients from MRI. MR images were segmented into air, lung, and soft tissue and converted into attenuation maps, some with constant lung density and some with patient-specific lung densities. The resulting PET images were compared by both global metrics of quantitative fidelity (accuracy, precision, and root mean squared error) and locally with relative error in volumes of interest.
RESULTS: A linear relationship was established between MRI and CT signal in the lungs. Constant lung density attenuation maps did not perform as well as patient-specific lung density attenuation maps, regardless of what constant density was chosen. In particular, when attenuation maps with patient-specific lung density were used, precision, accuracy, and root mean square error improved in lung tissue. In volumes of interest placed in the lungs, relative error was significantly reduced from a minimum of 12% to less than 5%. The benefit extended to tissues adjacent to the lungs but became less important as distance from the lungs increased.
CONCLUSION: A means of using MRI to infer patient-specific attenuation coefficients in the lungs was developed and applied to augment whole-body MRI-based attenuation maps. This technique has been shown to improve the quantitative fidelity of PET images in the lungs and nearby tissues, compared with an approach that assumes uniform lung density.

Entities:  

Mesh:

Year:  2012        PMID: 22566600     DOI: 10.2967/jnumed.111.098350

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  17 in total

Review 1.  Challenges and current methods for attenuation correction in PET/MR.

Authors:  Vincent Keereman; Pieter Mollet; Yannick Berker; Volkmar Schulz; Stefaan Vandenberghe
Journal:  MAGMA       Date:  2012-08-09       Impact factor: 2.310

2.  A comparison of CT- and MR-based attenuation correction in neurological PET.

Authors:  John C Dickson; Celia O'Meara; Anna Barnes
Journal:  Eur J Nucl Med Mol Imaging       Date:  2014-01-15       Impact factor: 9.236

Review 3.  Advances in PET/MR instrumentation and image reconstruction.

Authors:  Jorge Cabello; Sibylle I Ziegler
Journal:  Br J Radiol       Date:  2016-07-22       Impact factor: 3.039

4.  The foundation layer of quantitative cardiac PET/MRI: Attenuation correction. Again.

Authors:  Stephan G Nekolla; Jorge Cabello
Journal:  J Nucl Cardiol       Date:  2016-02-23       Impact factor: 5.952

Review 5.  PET/MR: Yet another Tesla?

Authors:  Markus Schwaiger; Karl Kunze; Christoph Rischpler; Stephan G Nekolla
Journal:  J Nucl Cardiol       Date:  2016-09-22       Impact factor: 5.952

6.  Arterial CO2 as a Potent Coronary Vasodilator: A Preclinical PET/MR Validation Study with Implications for Cardiac Stress Testing.

Authors:  Hsin-Jung Yang; Damini Dey; Jane Sykes; Michael Klein; John Butler; Michael S Kovacs; Olivia Sobczyk; Behzad Sharif; Xiaoming Bi; Avinash Kali; Ivan Cokic; Richard Tang; Roya Yumul; Antonio H Conte; Sotirios A Tsaftaris; Mourad Tighiouart; Debiao Li; Piotr J Slomka; Daniel S Berman; Frank S Prato; Joseph A Fisher; Rohan Dharmakumar
Journal:  J Nucl Med       Date:  2017-03-02       Impact factor: 10.057

Review 7.  MR Imaging-Guided Attenuation Correction of PET Data in PET/MR Imaging.

Authors:  David Izquierdo-Garcia; Ciprian Catana
Journal:  PET Clin       Date:  2016-01-26

Review 8.  Morphology supporting function: attenuation correction for SPECT/CT, PET/CT, and PET/MR imaging.

Authors:  Tzu C Lee; Adam M Alessio; Robert M Miyaoka; Paul E Kinahan
Journal:  Q J Nucl Med Mol Imaging       Date:  2015-11-17       Impact factor: 2.346

9.  Bias atlases for segmentation-based PET attenuation correction using PET-CT and MR.

Authors:  Jinsong Ouyang; Se Young Chun; Yoann Petibon; Ali A Bonab; Nathaniel Alpert; Georges El Fakhri
Journal:  IEEE Trans Nucl Sci       Date:  2013-10-01       Impact factor: 1.679

10.  Generation of PET Attenuation Map for Whole-Body Time-of-Flight 18F-FDG PET/MRI Using a Deep Neural Network Trained with Simultaneously Reconstructed Activity and Attenuation Maps.

Authors:  Donghwi Hwang; Seung Kwan Kang; Kyeong Yun Kim; Seongho Seo; Jin Chul Paeng; Dong Soo Lee; Jae Sung Lee
Journal:  J Nucl Med       Date:  2019-01-25       Impact factor: 10.057

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.