Literature DB >> 22529696

Swab culture monitoring of automated endoscope reprocessors after high-level disinfection.

Lung-Sheng Lu1, Keng-Liang Wu, Yi-Chun Chiu, Ming-Tzung Lin, Tsung-Hui Hu, King-Wah Chiu.   

Abstract

AIM: To conduct a bacterial culture study for monitoring decontamination of automated endoscope reprocessors (AERs) after high-level disinfection (HLD).
METHODS: From February 2006 to January 2011, authors conducted randomized consecutive sampling each month for 7 AERs. Authors collected a total of 420 swab cultures, including 300 cultures from 5 gastroscope AERs, and 120 cultures from 2 colonoscope AERs. Swab cultures were obtained from the residual water from the AERs after a full reprocessing cycle. Samples were cultured to test for aerobic bacteria, anaerobic bacteria, and mycobacterium tuberculosis.
RESULTS: The positive culture rate of the AERs was 2.0% (6/300) for gastroscope AERs and 0.8% (1/120) for colonoscope AERs. All the positive cultures, including 6 from gastroscope and 1 from colonoscope AERs, showed monofloral colonization. Of the gastroscope AER samples, 50% (3/6) were colonized by aerobic bacterial and 50% (3/6) by fungal contaminations.
CONCLUSION: A full reprocessing cycle of an AER with HLD is adequate for disinfection of the machine. Swab culture is a useful method for monitoring AER decontamination after each reprocessing cycle. Fungal contamination of AERs after reprocessing should also be kept in mind.

Entities:  

Keywords:  Automated endoscope reprocessor; Decontamination; Gastrointestinal scope; High-level disinfection; Monitoring; Swab culture

Mesh:

Year:  2012        PMID: 22529696      PMCID: PMC3325533          DOI: 10.3748/wjg.v18.i14.1660

Source DB:  PubMed          Journal:  World J Gastroenterol        ISSN: 1007-9327            Impact factor:   5.742


  10 in total

1.  Guidelines on cleaning and disinfection in GI endoscopy. Update 1999. The European Society of Gastrointestinal Endoscopy.

Authors: 
Journal:  Endoscopy       Date:  2000-01       Impact factor: 10.093

Review 2.  Variant Creutzfeldt-Jakob disease (vCJD) and gastrointestinal endoscopy.

Authors:  A T Axon; U Beilenhoff; M G Bramble; S Ghosh; A Kruse; G E McDonnell; C Neumann; J F Rey; K Spencer
Journal:  Endoscopy       Date:  2001-12       Impact factor: 10.093

3.  Surveillance cultures to monitor quality of gastrointestinal endoscope reprocessing.

Authors:  Frank M Moses; Jennifer Lee
Journal:  Am J Gastroenterol       Date:  2003-01       Impact factor: 10.864

4.  Surveillance culture monitoring of double-balloon enteroscopy reprocessing with high-level disinfection.

Authors:  King-Wah Chiu; Long-Sheng Lu; Keng-Liang Wu; Ming-Tzung Lin; Ming-Luen Hu; Wei-Chen Tai; Yi-Chun Chiu; Seng-Kee Chuah; Tsung-Hui Hu
Journal:  Eur J Clin Invest       Date:  2011-09-23       Impact factor: 4.686

5.  Surveillance culture of endoscope to monitor the quality of high-level disinfection of gastrointestinal reprocessing.

Authors:  King-Wah Chiu; Tze-Vun Fong; Keng-Liang Wu; Yi-Chun Chiu; Yeh-Pin Chou; Chung-Mou Kuo; Seng-Kee Chuah; Chung-Huang Kuo; Shue-Shien Chiou; Chi-Sin ChangChien
Journal:  Hepatogastroenterology       Date:  2010 May-Jun

6.  Endoscope drying/storage cabinet: interest and efficacy.

Authors:  L Pineau; E Villard; D L Duc; B Marchetti
Journal:  J Hosp Infect       Date:  2007-12-04       Impact factor: 3.926

7.  Infection control during GI endoscopy.

Authors:  Subhas Banerjee; Bo Shen; Douglas B Nelson; David R Lichtenstein; Todd H Baron; Michelle A Anderson; Jason A Dominitz; S Ian Gan; M Edwyn Harrison; Steven O Ikenberry; Sanjay B Jagannath; Robert D Fanelli; Ken Lee; Trina van Guilder; Leslie E Stewart
Journal:  Gastrointest Endosc       Date:  2008-03-19       Impact factor: 9.427

8.  Reprocessing of flexible gastrointestinal endoscopes. American Society for Gastrointestinal Endoscopy.

Authors: 
Journal:  Gastrointest Endosc       Date:  1996-05       Impact factor: 9.427

9.  A prospective study of the efficacy of routine decontamination for gastrointestinal endoscopes and the risk factors for failure.

Authors:  Linda Bisset; Yvonne E Cossart; Warwick Selby; Richard West; Denise Catterson; Kate O'hara; Karen Vickery
Journal:  Am J Infect Control       Date:  2006-06       Impact factor: 2.918

10.  Cleaning and disinfection of equipment for gastrointestinal endoscopy. Report of a Working Party of the British Society of Gastroenterology Endoscopy Committee.

Authors: 
Journal:  Gut       Date:  1998-04       Impact factor: 23.059

  10 in total
  5 in total

Review 1.  High-level disinfection of gastrointestinal endoscope reprocessing.

Authors:  King-Wah Chiu; Lung-Sheng Lu; Shue-Shian Chiou
Journal:  World J Exp Med       Date:  2015-02-20

2.  High pressure air jet in the endoscopic preparation room: risk of noise exposure on occupational health.

Authors:  King-Wah Chiu; Lung-Sheng Lu; Cheng-Kun Wu
Journal:  Biomed Res Int       Date:  2015-02-01       Impact factor: 3.411

3.  Surveillance cultures of samples obtained from biopsy channels and automated endoscope reprocessors after high-level disinfection of gastrointestinal endoscopes.

Authors:  King-Wah Chiu; Ming-Chao Tsai; Keng-Liang Wu; Yi-Chun Chiu; Ming-Tzung Lin; Tsung-Hui Hu
Journal:  BMC Gastroenterol       Date:  2012-09-03       Impact factor: 3.067

4.  Recent update of gastrointestinal endoscope reprocessing.

Authors:  Kyong Hee Hong; Yun Jeong Lim
Journal:  Clin Endosc       Date:  2013-05-31

5.  Endoscopic biopsy as quality assurance for endoscopic services.

Authors:  King-Wah Chiu; Shue-Shian Chiou
Journal:  PLoS One       Date:  2013-11-12       Impact factor: 3.240

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.