Literature DB >> 22512374

Accuracy of intraocular lens power calculation using partial coherence interferometry in patients with high myopia.

Gernot F Roessler1, Thomas S Dietlein, Niklas Plange, Anna-Katharina Roepke, Sven Dinslage, Peter Walter, Babac A Mazinani.   

Abstract

PURPOSE: Ultrasound-A-scan-biometry intraocular lens power calculation for cataract surgery sometimes shows lack of accuracy in patients with high myopia. The purpose of this retrospective study was to assess the accuracy of lens power calculation with optical biometry using the Zeiss IOLMaster across a large range of myopia levels.
METHODS: We included 37 consecutive, myopic eyes with an axial length >26.5mm (31 patients, 62±13years old, average preoperative refraction of -14.46±6.61D, range -3.5 to -32.0D which underwent phacoemulsification and implantation of an intraocular lens following biometry using the IOLMaster. For lens power calculation, the Haigis formula was used in all cases. For comparison, refraction was back-calculated using the SRK/T and Holladay I formulae.
RESULTS: The preoperative mean axial length was 29.37±2.44 mm with a range of 26.50-35.52mm. Thirty eyes (81.1%) showed a postoperative spherical equivalent which differed 1.00D or less from the predicted value, in 20 cases (54.1%) the postoperative refractive error was within±0.50D. The mean absolute error (MAE) was 0.70±0.59D (Holladay I, 0.85±0.68; SRK/T, 1.01±0.61D).
CONCLUSIONS: Optical biometry for intraocular lens power calculation seems to deliver reliable results for cataract surgery in patients with high myopia, although our data describe an increasing lack of accuracy beyond an axial length of 30mm. The Haigis formula provided the best predictability of postoperative refractive outcome for myopic eyes in general. Ophthalmic & Physiological Optics
© 2012 The College of Optometrists.

Entities:  

Mesh:

Year:  2012        PMID: 22512374     DOI: 10.1111/j.1475-1313.2012.00903.x

Source DB:  PubMed          Journal:  Ophthalmic Physiol Opt        ISSN: 0275-5408            Impact factor:   3.117


  17 in total

1.  Comparison of formulas and methods for high myopia patients requiring intraocular lens powers less than six diopters.

Authors:  Harry S Geggel
Journal:  Int Ophthalmol       Date:  2017-06-20       Impact factor: 2.031

2.  A Comparative Study to Assess the Predictability of Different IOL Power Calculation Formulas in Eyes of Short and Long Axial Length.

Authors:  Dharmil Doshi; Purvi Limdi; Nilesh Parekh; Neepa Gohil
Journal:  J Clin Diagn Res       Date:  2017-01-01

3.  The noninvasive retro-mode imaging of confocal scanning laser ophthalmoscopy in myopic maculopathy: a prospective observational study.

Authors:  Y Su; X Zhang; K Wu; Y Ji; C Zuo; M Li; F Wen
Journal:  Eye (Lond)       Date:  2014-06-13       Impact factor: 3.775

4.  Accuracy of axial length measurements from immersion B-scan ultrasonography in highly myopic eyes.

Authors:  Qing-Hua Yang; Bing Chen; Guang-Hua Peng; Zhao-Hui Li; Yi-Fei Huang
Journal:  Int J Ophthalmol       Date:  2014-06-18       Impact factor: 1.779

5.  Optical biometry intraocular lens power calculation using different formulas in patients with different axial lengths.

Authors:  Jia-Kang Wang; Shu-Wen Chang
Journal:  Int J Ophthalmol       Date:  2013-04-18       Impact factor: 1.779

6.  Partial Coherence Laser Interferometry in Highly Myopic versus Emmetropic Eyes.

Authors:  Gernot F Roessler; Yassin Djalali Talab; Thomas S Dietlein; Sven Dinslage; Niklas Plange; Peter Walter; Babac Ae Mazinani
Journal:  J Ophthalmic Vis Res       Date:  2014-04

7.  Intrapersonal comparison of initial axial length, keratometric readings, and intraocular lens power over a 6-month interval using an IOLMaster device.

Authors:  Bharkbhum Khambhiphant; Adisai Varadisai; Nat Visalvate
Journal:  Clin Ophthalmol       Date:  2014-12-22

8.  Influence of the prediction error of the first eye undergoing cataract surgery on the refractive outcome of the fellow eye.

Authors:  Ludmilla Gorodezky; Babac Ae Mazinani; Niklas Plange; Peter Walter; Martin Wenzel; Gernot Roessler
Journal:  Clin Ophthalmol       Date:  2014-10-31

9.  Accuracy of Intraocular Lens Power Calculation Formulas for Highly Myopic Eyes.

Authors:  Yichi Zhang; Xiao Ying Liang; Shu Liu; Jacky W Y Lee; Srinivasan Bhaskar; Dennis S C Lam
Journal:  J Ophthalmol       Date:  2016-03-29       Impact factor: 1.909

10.  Visual Performance after Bilateral Implantation of a Four-Haptic Diffractive Toric Multifocal Intraocular Lens in High Myopes.

Authors:  John S M Chang; Vincent K C Chan; Jack C M Ng; Antony K P Law
Journal:  J Ophthalmol       Date:  2016-08-02       Impact factor: 1.909

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.