Radiofrequency heating and magnetically induced displacement of dental magnetic attachments during 3.0 T MRI.

K Miyata1, M Hasegawa, Y Abe, T Tabuchi, T Namiki, T Ishigami.   

Abstract

OBJECTIVE: The aim of this study was to estimate the risk of injury from dental magnetic attachments due to their radiofrequency (RF) heating and magnetically induced displacement during 3.0 T MRI.
METHODS: To examine the magnetic attachments, we adopted the American Society for Testing and Materials F2182-02a and F2052-06 standards in two MRI systems (Achieva 3.0 T Nova Dual; Philips, Tokyo, Japan, and Signa HDxt 3.0 T; GE Healthcare, Milwaukee, WI). The temperature change was measured in a cylindrical keeper (GIGAUSS D600; GC, Tokyo, Japan) with coping of the casting alloy and a keeper with a dental implant at the maximum specific absorption rate (SAR) for 20 min. To measure the magnetically induced displacement force, three sizes of keepers (GIGAUSS D400, D600 and D1000) were used in deflection angle tests conducted at the point of the maximum magnetic field strength.
RESULTS: Temperature elevations of both coping and implant were higher in the Signa system than in the Achieva system. The highest temperature changes in the keeper with implant and keeper with coping were 0.6 °C and 0.8 °C in the Signa system, respectively. The temperature increase did not exceed 1.0 °C at any location. The deflection angle (α) was not measurable because it exceeded 90°. GIGAUSS D400 required an extra 3.0 g load to constrain the deflection angle to less than 45°; GIGAUSS D600 and D1000 required 5.0 and 9.0 g loads, respectively.
CONCLUSIONS: Dental magnetic attachments pose no risk due to RF heating and magnetically induced displacement at 3.0 T MRI. However, it is necessary to confirm that these keepers are securely attached to the prosthesis before imaging.

Mesh:

Substances:

Year:  2012        PMID: 22499128      PMCID: PMC3528202          DOI: 10.1259/dmfr/17778370

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  29 in total

1.  The direct and indirect techniques of making magnetically retained overdentures.

Authors:  N H Wang; W N von der Lehr
Journal:  J Prosthet Dent       Date:  1991-01       Impact factor: 3.426

2.  In vivo findings associated with heat generation during thermomechanical compaction of gutta-percha. 2. Histological response to temperature elevation on the external surface of the root.

Authors:  E M Saunders
Journal:  Int Endod J       Date:  1990-09       Impact factor: 5.264

3.  Neurostimulation systems: assessment of magnetic field interactions associated with 1.5- and 3-Tesla MR systems.

Authors:  Kenneth B Baker; John A Nyenhuis; Greg Hrdlicka; Ali R Rezai; Jean A Tkach; Frank G Shellock
Journal:  J Magn Reson Imaging       Date:  2005-01       Impact factor: 4.813

4.  Temperature evolution on human teeth root surface after diode laser assisted endodontic treatment.

Authors:  Norbert Gutknecht; Rene Franzen; Jörg Meister; Leon Vanweersch; Maziar Mir
Journal:  Lasers Med Sci       Date:  2005-07-09       Impact factor: 3.161

5.  Composite resin root coping with a keeper for magnetic attachment for replacing the missing coronal portion of a removable partial denture abutment.

Authors:  Yoshinobu Maeda; Katsuhiko Nakao; Kazutomo Yagi; Shisuke Matsuda
Journal:  J Prosthet Dent       Date:  2006-08       Impact factor: 3.426

6.  Variability in RF-induced heating of a deep brain stimulation implant across MR systems.

Authors:  Kenneth B Baker; Jean A Tkach; Micheal D Phillips; Ali R Rezai
Journal:  J Magn Reson Imaging       Date:  2006-12       Impact factor: 4.813

7.  Metal artifacts caused by gradient switching.

Authors:  Hansjörg Graf; Günter Steidle; Petros Martirosian; Ulrike A Lauer; Fritz Schick
Journal:  Magn Reson Med       Date:  2005-07       Impact factor: 4.668

Review 8.  Clinical advantages of 3.0 T MRI over 1.5 T.

Authors:  Winfried A Willinek; Hans H Schild
Journal:  Eur J Radiol       Date:  2007-12-26       Impact factor: 3.528

9.  Nonsusceptibility artifacts due to metallic objects in MR imaging.

Authors:  C R Camacho; D B Plewes; R M Henkelman
Journal:  J Magn Reson Imaging       Date:  1995 Jan-Feb       Impact factor: 4.813

10.  Thermal effects and antibacterial properties of energy levels required to sterilize stained root canals with an Nd:YAG laser.

Authors:  L O Ramsköld; C D Fong; T Strömberg
Journal:  J Endod       Date:  1997-02       Impact factor: 4.171

View more
  7 in total

1.  3-T MRI safety assessments of magnetic dental attachments and castable magnetic alloys.

Authors:  M Hasegawa; K Miyata; Y Abe; T Ishii; T Ishigami; K Ohtani; E Nagai; T Ohyama; Y Umekawa; S Nakabayashi
Journal:  Dentomaxillofac Radiol       Date:  2015-03-18       Impact factor: 2.419

2.  Predictability of magnetic susceptibility artifacts from metallic orthodontic appliances in magnetic resonance imaging.

Authors:  F Blankenstein; B T Truong; A Thomas; N Thieme; C Zachriat
Journal:  J Orofac Orthop       Date:  2014-11-26       Impact factor: 1.938

3.  Magnetization and demagnetization of magnetic dental attachments in a 3-T MRI system.

Authors:  Norio Hayashi; Akio Ogura; Toshio Tsuchihashi; Daisuke Takahashi; Tsuyoshi Matsuda; Shinya Seino; Tsukasa Doi
Journal:  Radiol Phys Technol       Date:  2017-04-27

4.  Radiofrequency heating of metallic dental devices during 3.0 T MRI.

Authors:  M Hasegawa; K Miyata; Y Abe; T Ishigami
Journal:  Dentomaxillofac Radiol       Date:  2013-03-21       Impact factor: 2.419

5.  Targeted endomyocardial biopsy guided by real-time cardiovascular magnetic resonance.

Authors:  Christina Unterberg-Buchwald; Christian Oliver Ritter; Verena Reupke; Robin Niklas Wilke; Christine Stadelmann; Michael Steinmetz; Andreas Schuster; Gerd Hasenfuß; Joachim Lotz; Martin Uecker
Journal:  J Cardiovasc Magn Reson       Date:  2017-04-19       Impact factor: 5.364

Review 6.  Imaging foreign bodies in head and neck trauma: a pictorial review.

Authors:  Jan Oliver Voss; Christoph Maier; Jonas Wüster; Benedicta Beck-Broichsitter; Tobias Ebker; Jana Vater; Steffen Dommerich; Jan D Raguse; Georg Böning; Nadine Thieme
Journal:  Insights Imaging       Date:  2021-02-15

7.  A Fully Implantable Miniaturized Liquid Crystal Polymer (LCP)-Based Spinal Cord Stimulator for Pain Control.

Authors:  Seunghyeon Yun; Chin Su Koh; Jungmin Seo; Shinyong Shim; Minkyung Park; Hyun Ho Jung; Kyungsik Eom; Jin Woo Chang; Sung June Kim
Journal:  Sensors (Basel)       Date:  2022-01-10       Impact factor: 3.576

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.