Literature DB >> 22472556

Long-term results of enterprise stent-assisted coiling of cerebral aneurysms.

Kyle M Fargen1, Brian L Hoh, Babu G Welch, G Lee Pride, Giuseppe Lanzino, Alan S Boulos, Jeffrey S Carpenter, Ansaar Rai, Erol Veznedaroglu, Andrew Ringer, Rafael Rodriguez-Mercado, Peter Kan, Adnan Siddiqui, Elad I Levy, J Mocco.   

Abstract

BACKGROUND: The Enterprise Vascular Reconstruction Device and Delivery System (Cordis; the Enterprise stent) was approved for use in conjunction with coiling of wide-necked aneurysms in 2007. No published long-term aneurysm occlusion or complication data exist for the Enterprise system.
OBJECTIVE: We compiled data on consecutive patients treated with Enterprise stent-assisted coiling of aneurysms from 9 high-volume neurointerventional centers.
METHODS: A 9 center registry was created to evaluate large volume data on the delayed safety and efficacy of the Enterprise stent system. Pooled data were compiled for consecutive patients undergoing Enterprise stent-assisted coiling at each institution prior to May 2009.
RESULTS: Two-hundred twenty-nine patients with 229 aneurysms, 32 of which were ruptured aneurysms, were included in the study. Mean clinical and angiographic follow-up was 619.6 ± 26.4 days and 655.7 ± 25.2 days, respectively. Mean aneurysm size was 9.2 ± 0.4 mm. Fifty-nine percent of patients demonstrated 100% coil obliteration and 81% had 90% or higher occlusion at last follow-up angiography. A total of 19 patients (8.3%) underwent retreatment of their aneurysms during the follow-up period. Angiographic in-stent stenosis was seen in 3.4% and thromboembolic events occurred in 4.4%. Overall, 90% of patients who underwent Enterprise-assisted coiling had a modified Rankin Scale score of 2 or less at last follow-up. A poor modified Rankin Scale score was strongly associated with rupture status (P < .001).
CONCLUSION: Although this study is limited by its retrospective nature, the Enterprise stent system appears to be an effective, safe, and durable treatment for intracranial aneurysms when used in conjunction with coiling.

Entities:  

Mesh:

Year:  2012        PMID: 22472556     DOI: 10.1227/NEU.0b013e3182571953

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  38 in total

1.  Does stent-assisted coiling still have a place in the management of intracranial aneurysms?

Authors:  L Pierot; C Cognard
Journal:  AJNR Am J Neuroradiol       Date:  2013-06-06       Impact factor: 3.825

Review 2.  Is flow diversion the death of cerebral bypass and coiling/stent-assisted coiling for giant cavernous aneurysms? A critical review on comparative outcomes and ongoing clinical trials.

Authors:  Tobias A Mattei; Andrew S Ferrell; Gavin W Britz
Journal:  Neurosurg Rev       Date:  2013-04-18       Impact factor: 3.042

3.  Stent-assisted coiling of intracranial aneurysms using LEO stents: long-term follow-up in 153 patients.

Authors:  Jacques Sedat; Yves Chau; Jean Gaudart; Marina Sachet; Stephanie Beuil; Michel Lonjon
Journal:  Neuroradiology       Date:  2017-12-23       Impact factor: 2.804

4.  Simultaneous stent expansion/balloon deflation technique to salvage failed balloon remodeling.

Authors:  Travis R Ladner; Lucy He; Brandon J Davis; Michael T Froehler; J Mocco
Journal:  BMJ Case Rep       Date:  2015-03-18

5.  Dual microcatheter coil embolization of acutely ruptured wide-necked intracranial aneurysms.

Authors:  Pyeong-Ho Yoon; Jae-Wook Lee; Yun-Ho Lee; Young-Sub Kwon; Kook-Hee Yang
Journal:  Interv Neuroradiol       Date:  2017-05-22       Impact factor: 1.610

6.  Is there an association between flow diverter fish mouthing and delayed-type hypersensitivity to metals?-a case-control study.

Authors:  Naci Kocer; Prabath Kumar Mondel; Elif Yamac; Ayse Kavak; Osman Kizilkilic; Civan Islak
Journal:  Neuroradiology       Date:  2017-09-05       Impact factor: 2.804

7.  Stent-assisted coiling in endovascular treatment of 500 consecutive cerebral aneurysms with long-term follow-up.

Authors:  S Geyik; K Yavuz; N Yurttutan; I Saatci; H S Cekirge
Journal:  AJNR Am J Neuroradiol       Date:  2013-07-25       Impact factor: 3.825

8.  Geometrical characteristics after Y-stenting of the basilar bifurcation.

Authors:  Muzaffer Sağlam; Osman Kızılkılıç; Vania Anagnostakou; Bülent Yıldız; Naci Koçer; Civan Işlak
Journal:  Diagn Interv Radiol       Date:  2015 Nov-Dec       Impact factor: 2.630

9.  A successful treatment of arterial ischemic stroke with stent insertion in a child with congenital heart disease.

Authors:  Hee Joon Yu; I-Seok Kang; Munhyang Lee; Keon-Ha Kim; Jeehun Lee
Journal:  Childs Nerv Syst       Date:  2012-07-31       Impact factor: 1.475

10.  First Experiences with the New Enterprise2® Stent.

Authors:  C Herweh; S Nagel; J Pfaff; C Ulfert; M Wolf; M Bendszus; M Möhlenbruch
Journal:  Clin Neuroradiol       Date:  2016-09-27       Impact factor: 3.649

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.