| Literature DB >> 22459364 |
Andreas Kyriacou1, Matthew E Li Kam Wa, Punam A Pabari, Beth Unsworth, Resham Baruah, Keith Willson, Nicholas S Peters, Prapa Kanagaratnam, Alun D Hughes, Jamil Mayet, Zachary I Whinnett, Darrel P Francis.
Abstract
BACKGROUND: In atrial fibrillation (AF), VV optimization of biventricular pacemakers can be examined in isolation. We used this approach to evaluate internal validity of three VV optimization methods by three criteria. METHODS ANDEntities:
Keywords: Biventricular pacemakers; Echocardiography; Electrocardiography; Pressure
Mesh:
Year: 2012 PMID: 22459364 PMCID: PMC3744806 DOI: 10.1016/j.ijcard.2012.03.086
Source DB: PubMed Journal: Int J Cardiol ISSN: 0167-5273 Impact factor: 4.164
Fig. 1Each circle shows 1 cardiac cycle in sinus rhythm, with electrical activity shown on the inside and mechanical activity on the outside. The E-wave is passive ventricular filling and the A-wave is active ventricular filling. For example, in a Medtronic pacemaker, selecting RV pre-excitation instead of simultaneous, while keeping programmed AV delay constant, increases the actual effective AV delay to left-sided pacing, because the programmed AV delay is to the “first lead to be activated” in this manufacturer's convention. The E-wave, which can only occur after the ventricle finishes ejecting, therefore occurs later (in relation to the A-wave) on the left side than it did before the VV delay change. Conversely, selecting LV pre-excitation has the mirror image effect on the right side of the heart. Other manufactures have different conventions for labelling the delays, but the constraint remains: the VV delay cannot be changed without changing the mechanical AV delay on one side of the heart or the other. All apparent VV optimizations in sinus rhythm therefore include an occult element of AV optimization.
Baseline patient characteristics.
| Demographics | Mean and SD or n (%) |
|---|---|
| Age | 75 SD 7 |
| Age range | 58–91 |
| Male | 16 (80%) |
| Aetiology | |
| Ischaemic | 10 (50%) |
| Dilated | 8 (40%) |
| Valvular | 2 (10%) |
| NYHA class | |
| II | 13 (65%) |
| III | 7 (35%) |
| LVEF (%) | 31 SD 13 |
| Medications | |
| ACEi/ARB | 19 (95%) |
| β-blocker | 17 (85%) |
| Diuretic | 17 (85%) |
| Spironolactone | 8 (40%) |
| Digoxin | 12 (60%) |
Comparison of optima singularity between modalities. Optimization by pressure provided a high degree of optima singularity (expressed in % above chance alone). QRS optimization had a reasonable degree of singularity, but flow optimization was the least convincing. The difference in degrees of singularity between modalities was significant (p < 0.005).
| Optimization method | Number of optimizations, out of 40, having a single optimal region | Number out of 80 | Degree of singularity, beyond expectation of chance | |
|---|---|---|---|---|
| Slow rate | Fast rate | Overall | ||
| Pressure | 36 | 38 | 74 | 85% |
| Flow | 30 | 28 | 58 | 45% |
| QRS | 35 | 30 | 65 | 63% |
| Difference between modalities: | Chi2: | 10.9 | ||
| p value: | < 0.005 | |||
Fig. 5By random chance alone, 50% of the parabolic curves fitted would be expected to be in the physiologically meaningful orientation. Therefore the raw percentage (x) of correctly orientated curves needs be transformed to 2(x − 50) to obtain the proportion in excess of chance. On this scale, Ø represents the average expectation by chance alone, and 100 represents perfect orientation. SBP had a significantly higher percentage of correctly orientated optimization curves than LVOT VTI and QRS, at both the slow and fast heart rates.
Fig. 6The reproducibility (standard deviation of the difference—SDD in ms) of LVOT VTI (A) is equally poor at the slow (34 ms) and fast (35 ms) heart rates. SBP reproducibility (B) was better than LVOT VTI at slow (10 ms, p < 0.01) and fast heart rates (9 ms, p < 0.01). The reproducibility of the QRS width (C) was also better at slow (8 ms, p < 0.01) and fast (6 ms, p < 0.01) rates.
Fig. 7LVOT VTI (A) and QRS (C) optima distributions contain a high proportion of optima with significant (− 40 ms) RV pre-excitation. In contrast, the SBP (B) optima distribution is more central, around VV 0 ms, and most optima are LV pre-excited.
Agreement of optima between modalities at the slow heart rate. Values on the principal diagonal of the table indicate reproducibility (in bold) of the optimum using the same modality twice; other values indicate the agreement between modalities. All agreements are quantified as the standard deviation of difference (SDD, ms) between the two optima obtained, in the 20 subjects.
| LVOT VTI | QRS to LVOT VTI | LVOT ejection time | MV VTI | MV VTI to QRS | MV ejection time | MPI | QRS | SBP | |
|---|---|---|---|---|---|---|---|---|---|
| Echocardiography | |||||||||
| LVOT VTI | 41.1 | 64.7 | 34.3 | 43.6 | 56.9 | 31.4 | 44.8 | 37.0 | |
| QRS to LVOT VTI | 41.1 | 62.6 | 43.1 | 51.6 | 49.5 | 35.8 | 23.8 | 36.1 | |
| LVOT ejection time | 64.7 | 62.6 | 57.9 | 67.9 | 51.5 | 67.5 | 60.6 | 61.7 | |
| MV VTI | 34.3 | 43.1 | 57.9 | 46.4 | 59.2 | 38.7 | 42.0 | 44.6 | |
| MV VTI to QRS | 43.6 | 51.6 | 67.9 | 46.4 | 52.9 | 51.4 | 49.0 | 47.2 | |
| MV ejection time | 56.9 | 49.5 | 51.5 | 59.2 | 52.9 | 62.8 | 57.7 | 53.0 | |
| MPI | 31.4 | 35.8 | 67.5 | 38.7 | 51.4 | 62.8 | 39.0 | 33.2 | |
| Electrocardiography | |||||||||
| QRS width | 44.8 | 23.8 | 60.6 | 42.0 | 49.0 | 57.7 | 39.0 | 33.3 | |
| Finometer | |||||||||
| SBP | 37.0 | 36.1 | 61.7 | 44.6 | 47.2 | 53.0 | 33.2 | 33.3 |
LVOT indicates left ventricular outflow tract; MV, mitral valve; VTI, velocity time integral; MPI, myocardial performance index; SBP, acute change in systolic blood pressure.
Agreement of optima between modalities at the fast heart rate. Values on the principal diagonal of the table indicate reproducibility (in bold) of the optimum using the same modality twice; other values indicate the agreement between modalities. All agreements are quantified as the as standard deviation of difference (SDD, ms) between the two optima obtained, in the 20 subjects.
| LVOT VTI | QRS to LVOT VTI | LVOT ejection time | MV VTI | MV VTI to QRS | MV ejection time | MPI | QRS | SBP | |
|---|---|---|---|---|---|---|---|---|---|
| Echocardiography | |||||||||
| LVOT VTI | 46.9 | 43.1 | 49.7 | 51.9 | 52.1 | 47.8 | 41.4 | 45.3 | |
| QRS to LVOT VTI | 46.9 | 46.6 | 47.3 | 69.5 | 59.4 | 47.8 | 43.6 | 53.2 | |
| LVOT ejection time | 43.1 | 46.6 | 58.7 | 68.7 | 54.1 | 56.2 | 48.4 | 54.6 | |
| MV VTI | 49.7 | 47.3 | 58.7 | 56.6 | 64.6 | 46.1 | 33.1 | 38.0 | |
| MV VTI to QRS | 51.9 | 69.5 | 68.7 | 56.6 | 58.2 | 57.6 | 51.1 | 40.6 | |
| MV ejection time | 52.1 | 59.4 | 54.1 | 64.6 | 58.2 | 69.2 | 59.9 | 52.7 | |
| MPI | 47.8 | 47.8 | 56.2 | 46.1 | 57.6 | 69.2 | 48.6 | 42.9 | |
| Electrocardiography | |||||||||
| QRS width | 41.4 | 43.6 | 48.4 | 33.1 | 51.1 | 59.9 | 48.6 | 43.9 | |
| Finometer | |||||||||
| SBP | 45.3 | 53.2 | 54.6 | 38.0 | 40.6 | 52.7 | 42.9 | 43.9 |
LVOT indicates left ventricular outflow tract; MV, mitral valve; VTI, velocity time integral; MPI, myocardial performance index; SBP, acute change in systolic blood pressure.