Comparison of a non-invasive arterial pulse contour technique and echo Doppler aorta velocity-time integral on stroke volume changes in optimization of cardiac resynchronization therapy.

Irene E van Geldorp1, Tammo Delhaas, Ben Hermans, Kevin Vernooy, Bernard Broers, Julija Klimusina, Francois Regoli, Francesco F Faletra, Tiziano Moccetti, Bart Gerritse, Richard Cornelussen, Jos J Settels, Harry J G M Crijns, Angelo Auricchio, Frits W Prinzen.   

Abstract

AIMS: We investigated the accuracy and feasibility of a non-invasive arterial pulse contour technique for continuous measurement of stroke volume (SV) in optimization of atrioventricular (AV) delay in cardiac resynchronization therapy (CRT), by comparing SV changes assessed by Nexfin CO-Trek® (Nexfin) and echo Doppler aortic velocity-time integral (VTIao). Furthermore, we investigated whether AV-delay optimization increases the effect of CRT when compared with a default AV delay (120 ms). METHODS AND
RESULTS: In 23 CRT patients, biventricular pacing (BiVP) was applied at various AV delays, while recording 10 beats preceding BiVP (baseline) and the first 10 BiVP beats, for both methods in parallel. Agreement between Nexfin and VTIao measurements was evaluated (Bland-Altman) on beat-to-beat changes in SV, as well as on effects of BiVP (averaged over 8 beats) at various AV delays. Individual optimal AV delays, for Nexfin (AVopt-n) and VTIao (AVopt-ao), were derived from the second-order polynomial fitted to the effect measurements of 20 patients. In 252 episodes assessed, the difference between measurements (= Nexfin - VTIao) was -0.6 ± 8.1% for beat-to-beat SV changes and -1.3 ± 7.3% for effects of BiVP. Optimal AV delays for Nexfin were well related to AVopt-ao (R(2) = 0.69). The effect (%) of BiVP at the optimal AV delay was significantly larger than at the default AV delay: median difference (range) being +6.3% (0.1-14.4%; P < 0.001) for VTIao and +4.7% (0.0-14.0%; P < 0.001) for Nexfin.
CONCLUSION: Individual AV optimization increases the effect of CRT. Nexfin is a promising tool in individual CRT optimization, as Nexfin agrees with VTIao on measuring beat-to-beat SV changes and on assessing relative effects of BiVP on SV at various AV delays.

Entities:  

Mesh:

Year:  2010        PMID: 20880954     DOI: 10.1093/europace/euq348

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  13 in total

1.  Determinants of vascular and cardiac baroreflex sensitivity values in a random population sample.

Authors:  Anke J Borgers; Bert-Jan H van den Born; Anneke Alkemade; Daan W Eeftinck Schattenkerk; Johannes J van Lieshout; Karel H Wesseling; Peter H Bisschop; Berend E Westerhof
Journal:  Med Biol Eng Comput       Date:  2013-10-19       Impact factor: 2.602

Review 2.  Strategies to improve cardiac resynchronization therapy.

Authors:  Kevin Vernooy; Caroline J M van Deursen; Marc Strik; Frits W Prinzen
Journal:  Nat Rev Cardiol       Date:  2014-05-20       Impact factor: 32.419

3.  The acute effects of changes to AV delay on BP and stroke volume: potential implications for design of pacemaker optimization protocols.

Authors:  Charlotte H Manisty; Ali Al-Hussaini; Beth Unsworth; Resham Baruah; Punam A Pabari; Jamil Mayet; Alun D Hughes; Zachary I Whinnett; Darrel P Francis
Journal:  Circ Arrhythm Electrophysiol       Date:  2011-11-17

Review 4.  When is an optimization not an optimization? Evaluation of clinical implications of information content (signal-to-noise ratio) in optimization of cardiac resynchronization therapy, and how to measure and maximize it.

Authors:  Punam A Pabari; Keith Willson; Berthold Stegemann; Irene E van Geldorp; Andreas Kyriacou; Michela Moraldo; Jamil Mayet; Alun D Hughes; Darrel P Francis
Journal:  Heart Fail Rev       Date:  2011-05       Impact factor: 4.214

5.  Optimisation of cardiac resynchronization therapy in clinical practice during exercise.

Authors:  M M D Molenaar; B Oude Velthuis; M F Scholten; J Y Stevenhagen; W A Wesselink; J M van Opstal
Journal:  Neth Heart J       Date:  2013-10       Impact factor: 2.380

Review 6.  Atrioventricular and interventricular delay optimization in cardiac resynchronization therapy: physiological principles and overview of available methods.

Authors:  Patrick Houthuizen; Frank A L E Bracke; Berry M van Gelder
Journal:  Heart Fail Rev       Date:  2011-05       Impact factor: 4.214

Review 7.  The "missing" link between acute hemodynamic effect and clinical response.

Authors:  Frits W Prinzen; Angelo Auricchio
Journal:  J Cardiovasc Transl Res       Date:  2011-11-17       Impact factor: 4.132

Review 8.  Noninvasive continuous hemodynamic monitoring.

Authors:  Jasper Truijen; Johannes J van Lieshout; Wilbert A Wesselink; Berend E Westerhof
Journal:  J Clin Monit Comput       Date:  2012-06-14       Impact factor: 2.502

9.  The estimation of cardiac output by the Nexfin device is of poor reliability for tracking the effects of a fluid challenge.

Authors:  Xavier Monnet; Fabien Picard; Elsa Lidzborski; Malcie Mesnil; Jacques Duranteau; Christian Richard; Jean-Louis Teboul
Journal:  Crit Care       Date:  2012-10-29       Impact factor: 9.097

10.  Optimisation of cardiac resynchronisation therapy during exercise.

Authors:  A Ghani; A R Ramdat Misier; A Elvan; P P H M Delnoy
Journal:  Neth Heart J       Date:  2013-10       Impact factor: 2.380

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.