Continuous glucose monitoring--a study of the Enlite sensor during hypo- and hyperbaric conditions.

Peter Adolfsson1, Hans Örnhagen, Bengt M Eriksson, Ken Cooper, Johan Jendle.   

Abstract

BACKGROUND: The performance and accuracy of the Enlite(™) (Medtronic, Inc., Northridge, CA) sensor may be affected by microbubble formation at the electrode surface during hypo- and hyperbaric conditions. The effects of acute pressure changes and of prewetting of sensors were investigated.
MATERIALS AND METHODS: On Day 1, 24 sensors were inserted on the right side of the abdomen and back in one healthy individual; 12 were prewetted with saline solution, and 12 were inserted dry. On Day 2, this procedure was repeated on the left side. All sensors were attached to an iPro continuous glucose monitoring (CGM) recorder. Hypobaric and hyperbaric tests were conducted in a pressure chamber, with each test lasting 105 min. Plasma glucose values were obtained at 5-min intervals with a HemoCue(®) (Ängelholm, Sweden) model 201 glucose analyzer for comparison with sensor glucose values.
RESULTS: Ninety percent of the CGM systems operated during the tests. The mean absolute relative difference was lower during hyperbaric than hypobaric conditions (6.7% vs. 14.9%, P<0.001). Sensor sensitivity was slightly decreased (P<0.05) during hypobaric but not during hyperbaric conditions. Clarke Error Grid Analysis showed that 100% of the values were found in the A+B region. No differences were found between prewetted and dry sensors.
CONCLUSIONS: The Enlite sensor performed adequately during acute pressure changes and was more accurate during hyperbaric than hypobaric conditions. Prewetting the sensors did not improve accuracy. Further studies on type 1 diabetes subjects are needed under various pressure conditions.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22428621     DOI: 10.1089/dia.2011.0284

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  6 in total

1.  Performance evaluation of three continuous glucose monitoring systems: comparison of six sensors per subject in parallel.

Authors:  Guido Freckmann; Stefan Pleus; Manuela Link; Eva Zschornack; Hans-Martin Klötzer; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2013-07-01

2.  Pilots and Diabetes Technology.

Authors:  Johan Jendle; Kurt Rinnert; Anton Westman; Lutz Heinemann
Journal:  J Diabetes Sci Technol       Date:  2016-12-05

3.  Personal Insulin Pump With Predictive Low Glucose Management Technology at High Altitude.

Authors:  Bartłomiej Matejko; Teresa Benbenek-Klupa; Maciej T Malecki; Tomasz Klupa
Journal:  J Diabetes Sci Technol       Date:  2016-07-10

4.  In-vitro performance of the Enlite Sensor in various glucose concentrations during hypobaric and hyperbaric conditions.

Authors:  Peter Adolfsson; Hans Ornhagen; Bengt M Eriksson; Raghavendhar Gautham; Johan Jendle
Journal:  J Diabetes Sci Technol       Date:  2012-11-01

5.  Comparison of venous, capillary and interstitial blood glucose data measured during hyperbaric oxygen treatment from patients with diabetes mellitus.

Authors:  Carol Baines; Don Vicendese; David Cooper; William McGuiness; Charne Miller
Journal:  Diving Hyperb Med       Date:  2021-09-30       Impact factor: 1.228

6.  Performance evaluation of a continuous glucose monitoring system under conditions similar to daily life.

Authors:  Stefan Pleus; Christina Schmid; Manuela Link; Eva Zschornack; Hans-Martin Klötzer; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2013-07-01
  6 in total