| Literature DB >> 22403482 |
Timothy P Taylor1, Michael G Janech, Elizabeth H Slate, Evelyn C Lewis, John M Arthur, Jim C Oates.
Abstract
Using multiplex bead assays to measure urine proteins has a great potential for biomarker discovery, but substances in urine (the matrix) can interfere with assay measurements. By comparing the recovery of urine spiked with known quantities of several common analytes, this study demonstrated that the urine matrix variably interfered with the accurate measurement of low abundance proteins. Dilution of the urine permitted a more accurate measure of these proteins, equivalent to the standard dilution technique when the diluted analytes were above the limits of detection of the assay. Therefore, dilution can be used as an effective technique for over-coming urine matrix effects in urine immunoassays. These results may be applicable to other biological fluids in which matrix components interfere with assay performance.Entities:
Keywords: analysis/urine; assay validation; biomarkers; body fluids urine; standard addition
Year: 2012 PMID: 22403482 PMCID: PMC3290108 DOI: 10.4137/BMI.S8703
Source DB: PubMed Journal: Biomark Insights ISSN: 1177-2719
Figure 1Recovery of five proteins spiked into urine samples from patients with kidney disease. Four different urine samples from patients with acute tubular necrosis, prerenal azotemia, and lupus nephritis were spiked with known quantities of each of five analytes. The percent recovery of analyte was calculated by subtracting the measured concentration of unspiked urine from the measured concentration of urine spiked with standard then dividing by the expected concentration of that standard (([measured urine] − [measured urine])/[expected standard] × 100).
Note: Results were reported as means ± standard deviations.
Figure 2Recovery of analytes spiked into urine samples and after dilution. Urine samples from the same four patients in Figure 1 were diluted in sample buffer at either neat concentrations or 1:2, 1:5, 1:10 and 1:20. Samples were then spiked with known amounts of five different protein analytes. The percent recovery was calculated using the formula from Figure 1.
Notes: Results were reported as means ± standard deviations.
Figure 3Comparison of dilution and standard addition techniques for overcoming matrix effects. Analyte concentrations from the same four urine samples from Figures 1 and 2 were measured by diluting each sample 1:10 and by using the standard addition technique. Standard addition was performed by spiking several concentrations of analyte into neat urine. The inset is an example of how the standard addition technique is used to determine the concentration of an unknown sample. The analyte concentration is calculated as the x-intercept multiplied by −1. A line of identity demonstrates how closely (when the diluted concentration of analyte was above the limit of quantification) the dilution and standard addition techniques determine analyte concentration for each of the four urine samples.
Note: Results were reported as means ± standard deviations.
Diagnosis and renal function in patients 1–4.
| Patient # | Diagnosis | Renal function surrogate |
|---|---|---|
| 90 (1) | Acute tubular necrosis | Serum Cr 6.0 |
| 414 (2) | Prerenal azotemia | Serum Cr 1.8 |
| 918 (3) | Prerenal azotemia | Serum Cr 2.2 |
| 923 (4) | Class V lupus nephritis | Serum Cr 5.1 |
Urine components for individual patient urine samples.
| Urine 1 (90) | Urine 2 (414) | Urine 4 (923) | Urine 3 (918) | PBS [0.5%BSA] | |
|---|---|---|---|---|---|
| Ca (mg/dl) | 6.5 | < 2 | 2 | 2 | 0 |
| Ucr (mg/dl) | 15 | 112 | 210 | 90 | 0 |
| K (mM) | 7.7 | 44.2 | 30.3 | 24.8 | 2.7 |
| Na (mM) | 106 | 74 | 14 | < 10 | 138 |
| UUN (mg/dl) | < 50 | 975 | 348 | 1406 | 0 |
| Osm | 252 | 549 | 273 | 629 | 270 |
| pH | 7.8 | 6.7 | 4.6 | 5.3 | 7.4 |
Standard diluent, calculated values urines were measured by the central Lab at the Medical University Hospital.