Literature DB >> 22383019

High rate of ceramic sandwich liner fracture.

Ronny Lopes1, Jean M Philippeau, Norbert Passuti, François Gouin.   

Abstract

BACKGROUND: Ceramic bearing surfaces for THA were introduced to reduce the risk of wear. However, owing to liner fracture in some of the early series and presumption that the fractures were the result of the modulus mismatch of the implant and the bone, a ceramic sandwich liner with lower structural rigidity was introduced. Fractures of these devices also were reported subsequently, although the incidence is unclear and it is unknown whether there are any risk factors associated with the fractures. QUESTIONS/PURPOSES: We therefore determined the incidence of these fractures.
METHODS: We retrospectively reviewed 298 active patients in whom we implanted 353 ceramic-polyethylene sandwich liner acetabular components between November 1999 and February 2008. The mean age of the patients was 53.6 years (range, 17-84 years). The minimum followup was 6 months (mean, 41 months; range, 6-106 months). All patients were assessed clinically and radiographically.
RESULTS: Seven of the 353 (2%) ceramic sandwich liners fractured at a mean of 4.3 years (range, 1.3-7.6 years) after surgery without trauma. Neither patient-related factors nor radiographic position of the implants were risk factors for fracture.
CONCLUSIONS: Owing to the high rate of fractures of the sandwich ceramic polyethylene liners in our patients, we have discontinued use of this device. LEVEL OF EVIDENCE: Level IV, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22383019      PMCID: PMC3348322          DOI: 10.1007/s11999-012-2279-6

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  17 in total

1.  Five-year follow-up with a ceramic sandwich cup in total hip replacement.

Authors:  F Ravasi; V Sansone
Journal:  Arch Orthop Trauma Surg       Date:  2002-02-02       Impact factor: 3.067

2.  Ceramic acetabular liner fracture in total hip arthroplasty with a ceramic sandwich cup.

Authors:  Masahiro Hasegawa; Akihiro Sudo; Hitoshi Hirata; Atsumasa Uchida
Journal:  J Arthroplasty       Date:  2003-08       Impact factor: 4.757

3.  Functional results of hip arthroplasty with acrylic prosthesis.

Authors:  R M D'AUBIGNE; M POSTEL
Journal:  J Bone Joint Surg Am       Date:  1954-06       Impact factor: 5.284

Review 4.  Concerns about ceramics in THA.

Authors:  Robert L Barrack; Corey Burak; Harry B Skinner
Journal:  Clin Orthop Relat Res       Date:  2004-12       Impact factor: 4.176

5.  [Total arthroplasty of the hip by fritted aluminum prosthesis. Experimental study and 1st clinical applications].

Authors:  P Boutin
Journal:  Rev Chir Orthop Reparatrice Appar Mot       Date:  1972 Apr-May

6.  The long-term results of low-friction arthroplasty of the hip performed as a primary intervention.

Authors:  J Charnley
Journal:  J Bone Joint Surg Br       Date:  1972-02

7.  [Alumina and its use in surgery of the hip. (Experimental study)].

Authors:  P Boutin
Journal:  Presse Med       Date:  1971-03-20       Impact factor: 1.228

8.  Ceramic failure after total hip arthroplasty with an alumina-on-alumina bearing.

Authors:  Youn-Soo Park; Sung-Kwan Hwang; Won-Sik Choy; Yong-Sik Kim; Young-Wan Moon; Seung-Jae Lim
Journal:  J Bone Joint Surg Am       Date:  2006-04       Impact factor: 5.284

9.  Three-dimensional polyethylene wear of a press-fit titanium prosthesis. Factors influencing generation of polyethylene debris.

Authors:  P A Devane; J G Horne; K Martin; G Coldham; B Krause
Journal:  J Arthroplasty       Date:  1997-04       Impact factor: 4.757

10.  Cementless total hip arthroplasty with ceramic-on-ceramic bearing in patients younger than 45 years with femoral-head osteonecrosis.

Authors:  Young-Hoo Kim; Yoowang Choi; Jun-Shik Kim
Journal:  Int Orthop       Date:  2009-09-26       Impact factor: 3.075

View more
  8 in total

1.  Experience with sandwich liner and its high rate of failure.

Authors:  John Junghun Shin; Ji-Soo Shin; Jae-Hyun Kim; Sung Kwan Hwang
Journal:  Eur J Orthop Surg Traumatol       Date:  2014-07-16

2.  Letter to the editor: High rate of ceramic sandwich liner fracture.

Authors:  Yong-Chan Ha
Journal:  Clin Orthop Relat Res       Date:  2012-11       Impact factor: 4.176

Review 3.  Safe clinical use of carbon nanotubes as innovative biomaterials.

Authors:  Naoto Saito; Hisao Haniu; Yuki Usui; Kaoru Aoki; Kazuo Hara; Seiji Takanashi; Masayuki Shimizu; Nobuyo Narita; Masanori Okamoto; Shinsuke Kobayashi; Hiroki Nomura; Hiroyuki Kato; Naoyuki Nishimura; Seiichi Taruta; Morinobu Endo
Journal:  Chem Rev       Date:  2014-04-10       Impact factor: 60.622

4.  Navigated Acetabular Cup Fixation for Acetabular Deformity or Revision Total Hip Arthroplasty.

Authors:  Ho Hyun Yun; Jung-Ro Yoon; Jung Jin Yu; Hyo-Sung Seo
Journal:  Hip Pelvis       Date:  2014-09-29

5.  Nine year follow-up of a ceramic-on-ceramic bearing total hip arthroplasty utilizing a layered monoblock acetabular component.

Authors:  David Mayor; Savan Patel; Clayton Perry; Norman Walter; Stephen Burton; Theresa Atkinson
Journal:  Iowa Orthop J       Date:  2014

6.  Do the Reasons for Ceramic-on-ceramic Revisions Differ From Other Bearings in Total Hip Arthroplasty?

Authors:  Henri Migaud; Sophie Putman; Grégory Kern; Ronald Isida; Julien Girard; Nassima Ramdane; Christian P Delaunay; Moussa Hamadouche
Journal:  Clin Orthop Relat Res       Date:  2016-10       Impact factor: 4.176

7.  Comparison of the risk of revision in cementless total hip arthroplasty with ceramic-on-ceramic and metal-on-polyethylene bearings.

Authors:  Claus Varnum; Alma B Pedersen; Per Kjærsgaard-Andersen; Søren Overgaard
Journal:  Acta Orthop       Date:  2015-01-30       Impact factor: 3.717

8.  Total Hip Arthroplasty Using Hybrid Ceramic Bearing: A Minimum 10-Year Followup Study.

Authors:  Jun-Il Yoo; Yong-Chan Ha; Dong-Hyun Kim; Young-Kyun Lee; Kyung-Hoi Koo
Journal:  Indian J Orthop       Date:  2019 Sep-Oct       Impact factor: 1.251

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.