Reproducibility and repeatability of the OcuSense TearLab™ osmometer.

Frank Eperjesi1, Maana Aujla, Hannah Bartlett.   

Abstract

BACKGROUND: Some studies report that increased tear osmolarity is a reliable indicator of dry eye syndrome (DES). The OcuSense TearLab™ osmometer requires less than a 100-nl sample of tears and provides an instant quantitative result. Our aim was to clinically evaluate this instrument in terms of its reproducibility and repeatability.
METHODS: Twenty-nine participants who ranged in age from 19 to 49 years (mean ± SD: 23.3 ± 5.5 years) were recruited. Osmolarity readings were collected by two operators, in two sessions separated by 1 or 2 weeks in order to assess test reproducibility and repeatability.
RESULTS: The coefficient of reproducibility was 39 mOsms/l; the coefficient of repeatability was 33 mOsms/l.
CONCLUSIONS: Our mean coefficient of variation over four readings for 29 subjects is 2.9%, which compares well with that reported by the manufacturer. Our results inform practitioners about the level of change over time that can be considered clinically relevant for healthy subjects. This value is 33mOsms/l; any change smaller than this could be attributed to measurement noise.

Entities:  

Mesh:

Year:  2012        PMID: 22350061     DOI: 10.1007/s00417-012-1961-4

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  16 in total

1.  Vapor pressure osmometry: minimum sample microvolumes.

Authors:  C D Pensyl; W J Benjamin
Journal:  Acta Ophthalmol Scand       Date:  1999-02

2.  Electrolytic conductivity, osmotic pressure, and hydrogen ion concentration of human lachrymal fluid.

Authors:  K PEDERSEN-BJERGAARD; B C SMIDT
Journal:  Acta Derm Venereol Suppl (Stockh)       Date:  1952

3.  Hyperosmolarity of the tear film in dry eye syndrome.

Authors:  Elisabeth M Messmer; Meral Bulgen; Anselm Kampik
Journal:  Dev Ophthalmol       Date:  2010-05-18

4.  The repeatability of best corrected acuity in normal and amblyopic children 4 to 12 years of age.

Authors:  Sean I Chen; Arvind Chandna; Anthony M Norcia; Mark Pettet; Deborah Stone
Journal:  Invest Ophthalmol Vis Sci       Date:  2006-02       Impact factor: 4.799

5.  Barriers to clinical uptake of tear osmolarity measurements.

Authors:  Santosh Khanal; Thomas J Millar
Journal:  Br J Ophthalmol       Date:  2011-05-23       Impact factor: 4.638

6.  Osmotic pressure of the tears during adaptation to contact lenses.

Authors:  N P Uniacke; R M Hill
Journal:  J Am Optom Assoc       Date:  1970-11

7.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

8.  Comparison of human tear film osmolarity measured by electrical impedance and freezing point depression techniques.

Authors:  Alan Tomlinson; Louise C McCann; Edward I Pearce
Journal:  Cornea       Date:  2010-09       Impact factor: 2.651

9.  Osmolarity of tear microvolumes in keratoconjunctivitis sicca.

Authors:  J P Gilbard; R L Farris; J Santamaria
Journal:  Arch Ophthalmol       Date:  1978-04

10.  Tear osmolarity measurement using the TearLab Osmolarity System in the assessment of dry eye treatment effectiveness.

Authors:  Umberto Benelli; Marco Nardi; Chiara Posarelli; Timothy G Albert
Journal:  Cont Lens Anterior Eye       Date:  2010-02-12       Impact factor: 3.077

View more
  6 in total

Review 1.  TFOS DEWS II Tear Film Report.

Authors:  Mark D P Willcox; Pablo Argüeso; Georgi A Georgiev; Juha M Holopainen; Gordon W Laurie; Tom J Millar; Eric B Papas; Jannick P Rolland; Tannin A Schmidt; Ulrike Stahl; Tatiana Suarez; Lakshman N Subbaraman; Omür Ö Uçakhan; Lyndon Jones
Journal:  Ocul Surf       Date:  2017-07-20       Impact factor: 5.033

2.  Variability of Tear Osmolarity in Patients With Dry Eye.

Authors:  Vatinee Y Bunya; Nicole M Fuerst; Maxwell Pistilli; Bridgette E McCabe; Rebecca Salvo; Ilaria Macchi; Gui-Shuang Ying; Mina Massaro-Giordano
Journal:  JAMA Ophthalmol       Date:  2015-06       Impact factor: 7.389

3.  What is the value of incorporating tear osmolarity measurement in assessing patient response to therapy in dry eye disease?

Authors:  Francisco Amparo; Yiping Jin; Pedram Hamrah; Debra A Schaumberg; Reza Dana
Journal:  Am J Ophthalmol       Date:  2013-09-21       Impact factor: 5.258

Review 4.  Diagnostic procedures and management of dry eye.

Authors:  Snježana Kaštelan; Martina Tomić; Jasminka Salopek-Rabatić; Branko Novak
Journal:  Biomed Res Int       Date:  2013-08-19       Impact factor: 3.411

5.  Estimating the Osmolarities of Tears During Evaporation Through the "Eyes" of the Corneal Nerves.

Authors:  Harumitsu Hirata; Kamila Mizerska; Valentina Dallacasagrande; Mark I Rosenblatt
Journal:  Invest Ophthalmol Vis Sci       Date:  2017-01-01       Impact factor: 4.799

6.  Response to "Variability of Tear Osmolarity Measurements With a Point-of-Care System in Healthy Subjects-Systematic Review".

Authors:  Benjamin David Sullivan
Journal:  Cornea       Date:  2019-06       Impact factor: 2.651

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.