Evaluation of the impact of raw materials on the fatigue and mechanical properties of ProFile Vortex rotary instruments.

Yong Gao1, James L Gutmann, Kevin Wilkinson, Randall Maxwell, Dan Ammon.   

Abstract

INTRODUCTION: In this in vitro study, raw materials (including stainless steel, conventional superelastic nickel-titanium [NiTi], M-Wire NiTi, and Vortex Blue NiTi) were used to create ProFile Vortex designed 25/.06 instruments and subject these instruments to testing for fatigue resistance, torsional properties, flexibility, and Vickers microhardness. The comparative results in this study will enable the clinician's understanding of the performance of these materials for better choices in application during endodontic procedures.
METHODS: Cyclic fatigue testing was performed by rotating files in an artificially constructed stainless steel canal with a 5-mm radius and a 90° angle of curvature at 500 rpm. Torsional properties and flexibility in bending were assessed according to specification ISO 3630-1. Vickers microhardness was measured on the cross section of instruments with 300-g load and 15-second dwell time.
RESULTS: There were significant differences in the average fatigue life and flexibility for instruments made of different materials (P < .05). Vortex Blue was ranked first in both fatigue and flexibility, followed by M-Wire, superelastic wire, and stainless steel. For torsional strength and microhardness, stainless steel and M-Wire were ranked first and second, respectively. There was no statistically significant difference between superelastic wire and Vortex Blue. Vortex Blue showed the greatest distortion angle at break, whereas the other 3 materials showed comparable degree of rotation in the torque test.
CONCLUSIONS: Under the limitations of this study, NiTi shape memory alloy appeared to be a superior material option compared with stainless steel for its use in the application of endodontic rotary instruments. Vortex Blue and M-Wire offered functional advantages over conventional superelastic NiTi. Vortex Blue showed improved fatigue resistance and flexibility compared with ProFile Vortex M-Wire. Copyright Â
© 2012 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22341083     DOI: 10.1016/j.joen.2011.11.004

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  32 in total

1.  Performance of NiTi endodontic instrument under different temperatures.

Authors:  Ahmed Jamleh; Yoshio Yahata; Arata Ebihara; Amre R Atmeh; Turki Bakhsh; Hideaki Suda
Journal:  Odontology       Date:  2015-08-22       Impact factor: 2.634

2.  Shaping ability of reciprocating single-file systems in severely curved canals: WaveOne and Reciproc versus WaveOne Gold and Reciproc blue.

Authors:  Sebastian Bürklein; Stefanie Flüch; Edgar Schäfer
Journal:  Odontology       Date:  2018-05-18       Impact factor: 2.634

3.  Both operator and heat treatment determine the centring ability of Reciproc® files in vitro.

Authors:  N Hofmann
Journal:  Clin Oral Investig       Date:  2018-05-12       Impact factor: 3.573

4.  Canal transportation and centering ratio after preparation in severely curved canals: analysis by micro-computed tomography and double-digital radiography.

Authors:  Ane Poly; Fouad AlMalki; Fernando Marques; Bekir Karabucak
Journal:  Clin Oral Investig       Date:  2019-03-05       Impact factor: 3.573

5.  Cyclic fatigue resistance of XP-endo Shaper compared with different nickel-titanium alloy instruments.

Authors:  Amr Elnaghy; Shaymaa Elsaka
Journal:  Clin Oral Investig       Date:  2017-10-13       Impact factor: 3.573

6.  Cyclic fatigue and torsional strength of three different thermally treated reciprocating nickel-titanium instruments.

Authors:  Murilo Priori Alcalde; Marco Antonio Hungaro Duarte; Clovis Monteiro Bramante; Bruno Carvalho de Vasconselos; Mario Tanomaru-Filho; Juliane Maria Guerreiro-Tanomaru; Jader Camilo Pinto; Marcus Vinicius Reis Só; Rodrigo Ricci Vivan
Journal:  Clin Oral Investig       Date:  2017-12-09       Impact factor: 3.573

7.  Screw-in forces during instrumentation by various file systems.

Authors:  Jung-Hong Ha; Sang Won Kwak; Sung-Kyo Kim; Hyeon-Cheol Kim
Journal:  Restor Dent Endod       Date:  2016-11-08

8.  Cyclic fatigue resistance of R-Pilot, WaveOne Gold Glider, and ProGlider glide path instruments.

Authors:  Cangül Keskin; Uğur İnan; Murat Demiral; Ali Keleş
Journal:  Clin Oral Investig       Date:  2018-02-17       Impact factor: 3.573

9.  Root canal shaping of curved canals by Reciproc Blue system and Pro Taper Gold: A micro-computed tomographic study.

Authors:  Rogério-Vieira Silva; Murilo-Priori Alcalde; Martinho-Campolina-Ribeiro Horta; Clarissa-Teles Rodrigues; Frank-Ferreira Silveira; Marco-Antônio Duarte; Eduardo Nunes
Journal:  J Clin Exp Dent       Date:  2021-02-01

10.  An in vitro cyclic fatigue resistance comparison of conventional and new generation nickel-titanium rotary files.

Authors:  Celia Ruiz-Sánchez; Vicente Faus-Matoses; Teresa Alegre-Domingo; Ignacio Faus-Matoses; Vicente-José Faus-Llácer
Journal:  J Clin Exp Dent       Date:  2018-08-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.