Literature DB >> 9037647

Accuracy of infrared ear thermometry in adult patients.

K Stavem1, H Saxholm, N Smith-Erichsen.   

Abstract

OBJECTIVE: To assess (1) the agreement between infrared ear thermometry and core reference temperature (in the pulmonary artery). (2) the agreement between measurements in the right and left ears, and (3) the screening validity of infrared tympanic thermometry in detecting rectal fever.
DESIGN: Temperatures were measured in both ears with an infrared thermometer, in one group of patients by simultaneous measurements with thermistors inserted in the pulmonary artery, esophagus, and rectum, and in the other group with a rectal glass-mercury thermometer.
SETTING: An intensive care unit and a department of internal medicine in a secondary care hospital. PATIENTS AND PARTICIPANTS: Two samples: 16 adult patients admitted to the intensive care unit and 103 consecutive patients admitted to the department of medicine. MEASUREMENTS: The major outcome measures were (a) the agreement between infrared ear thermometry and thermistor pulmonary artery temperature and (b) the sensitivity and specificity for detecting fever, using rectal measurement as reference.
RESULTS: Both rectal and esophageal thermistor measurements showed better agreement with the pulmonary artery reference temperature than single ear tympanic thermometry. The sensitivity and specificity of ear thermometry for detecting fever (> or = 38.0 degrees C rectal reference) were 0.58 and 0.94, respectively. Double ear thermometry had a sensitivity of 0.61 and a specificity of 0.95, when using the mean value.
CONCLUSIONS: Both rectal and esophageal thermistor measurements showed better agreement with pulmonary artery temperature than single ear themometry. Using the mean of two ear measurements improves the agreement and screening validity for detecting fever by rectal temperature. If temperature measurements are critical, esophageal measurements achieve excellent agreement with pulmonary artery temperatures.

Entities:  

Mesh:

Year:  1997        PMID: 9037647     DOI: 10.1007/s001340050297

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  13 in total

1.  Effect of pre-warmed intravenous fluids on perioperative hypothermia and shivering after ambulatory surgery under monitored anesthesia care.

Authors:  Gahyun Kim; Myung Hee Kim; Sangmin M Lee; Soo Joo Choi; Young Hee Shin; Hee Joon Jeong
Journal:  J Anesth       Date:  2014-04-01       Impact factor: 2.078

Review 2.  Critical care in the emergency department: monitoring the critically ill patient.

Authors:  F J Andrews; J P Nolan
Journal:  Emerg Med J       Date:  2006-07       Impact factor: 2.740

3.  Comparison of temporal artery, rectal and esophageal core temperatures in children: Results of a pilot study.

Authors:  Fahad Al-Mukhaizeem; Upton Allen; Luba Komar; Basem Naser; Larry Roy; Derek Stephens; Stanley Read; Christina Kim; Suzanne Schuh
Journal:  Paediatr Child Health       Date:  2004-09       Impact factor: 2.253

4.  [Accidental hypothermia in the household environment. Importance of preclinical temperature measurement].

Authors:  S Russo; A Timmermann; O Radke; T Kerren; A Bräuer
Journal:  Anaesthesist       Date:  2005-12       Impact factor: 1.041

5.  Temperature measurement in intensive care patients: comparison of urinary bladder, oesophageal, rectal, axillary, and inguinal methods versus pulmonary artery core method.

Authors:  J-Y Lefrant; L Muller; J Emmanuel de La Coussaye; M Benbabaali; C Lebris; N Zeitoun; C Mari; G Saïssi; J Ripart; J-J Eledjam
Journal:  Intensive Care Med       Date:  2003-02-08       Impact factor: 17.440

6.  Oral and tympanic membrane temperatures are inaccurate to identify Fever in emergency department adults.

Authors:  Barbara J Barnett; Stacy Nunberg; Julia Tai; Martin L Lesser; Vladmir Fridman; Patricia Nichols; Richard Powell; Robert Silverman
Journal:  West J Emerg Med       Date:  2011-11

7.  Evaluation of performance and uncertainty of infrared tympanic thermometers.

Authors:  Wenbin Chung; Chiachung Chen
Journal:  Sensors (Basel)       Date:  2010-03-31       Impact factor: 3.576

8.  Effect of pre-warming on perioperative hypothermia during holmium laser enucleation of the prostate under spinal anesthesia: a prospective randomized controlled trial.

Authors:  Joo-Hyun Jun; Mi Hwa Chung; Eun Mi Kim; In-Jung Jun; Jung Hwa Kim; Joon-Sang Hyeon; Mi Hyeon Lee; Hye Sun Lee; Eun Mi Choi
Journal:  BMC Anesthesiol       Date:  2018-12-22       Impact factor: 2.217

9.  Accuracy and precision of four common peripheral temperature measurement methods in intensive care patients.

Authors:  Simin Asadian; Alireza Khatony; Gholamreza Moradi; Alireza Abdi; Mansour Rezaei
Journal:  Med Devices (Auckl)       Date:  2016-09-01

10.  Intra-rater reliability when using a tympanic thermometer under different self-measurement conditions.

Authors:  Won-Gyu Yoo
Journal:  J Phys Ther Sci       Date:  2016-07-29
View more