Quantitative comparison of contrast and imaging depth of ultrahigh-resolution optical coherence tomography images in 800-1700 nm wavelength region.

Shutaro Ishida1, Norihiko Nishizawa.   

Abstract

We investigated the wavelength dependence of imaging depth and clearness of structure in ultrahigh-resolution optical coherence tomography over a wide wavelength range. We quantitatively compared the optical properties of samples using supercontinuum sources at five wavelengths, 800 nm, 1060 nm, 1300 nm, 1550 nm, and 1700 nm, with the same system architecture. For samples of industrially used homogeneous materials with low water absorption, the attenuation coefficients of the samples were fitted using Rayleigh scattering theory. We confirmed that the systems with the longer-wavelength sources had lower scattering coefficients and less dependence on the sample materials. For a biomedical sample, we observed wavelength dependence of the attenuation coefficient, which can be explained by absorption by water and hemoglobin.

Entities:  

Keywords:  (110.4500) Optical coherence tomography; (170.3880) Medical and biological imaging

Year:  2012        PMID: 22312581      PMCID: PMC3269845          DOI: 10.1364/BOE.3.000282

Source DB:  PubMed          Journal:  Biomed Opt Express        ISSN: 2156-7085            Impact factor:   3.732


1. Introduction

Optical coherence tomography (OCT) is a non-invasive optical imaging technique for micrometer-scale cross-sectional imaging of biological tissue and materials [1-3]. Initially, it was a clinically useful diagnostic technique in ophthalmology [4,5]. Since then, OCT instrumentation has undergone substantial improvement and is now well-positioned for wide adoption in various other clinical and research applications [6]. For instance, this technology has been actively used in other clinical applications, such as cardiology [7-9], oncology [10,11], visualization of dental structure [12-14], and so on. Although OCT provides many advantages in medicine, the low penetration depth is a serious limitation for other applications [2]. Because the optical properties of tissue depend on the wavelength used, it is necessary to choose the proper wavelength to maximize the light penetration and enhance the image contrast at deeper depths. Around the near-infrared wavelength region, the first clinical applications of OCT for ophthalmology normally used 800 nm OCT systems. In other applications, such as observation of turbid biological tissue samples, 1300 nm OCT systems have been shown to have superior penetration depth compared with 800 nm OCT systems [12,15-18]. Recently, the long-wavelength window around 1700 nm has attracted attention for OCT imaging. Enhanced penetration depth at 1700 nm in skin, 10% intralipid solution, and rubber was confirmed by Sharma et al. [19]. On the other hand, the dependence of the scattering coefficient on intralipid concentration in water for 1300 nm and 1600 nm systems was examined by Kodach et al. [20]. The advantages of using this longer wavelength region are enhanced penetration depth due to lower scattering in tissue and enhanced imaging contrast at deeper penetration depths where multiple-scattered photons dominate over ballistic photons. However, there have been no studies comparing the performance of ultrahigh-resolution optical coherence tomography (UHR-OCT) for the same sample over a wide wavelength range. In this study, we constructed UHR-OCT systems using supercontinuum sources at five wavelengths in the range 800–1700 nm. We investigated the wavelength dependence of the images for several different samples and quantitatively compared their optical properties.

2. Experimental setup

2.1. Supercontinuum sources: Design and output characteristics

We constructed supercontinuum (SC) sources at wavelengths 800 nm, 1060 nm, 1300 nm, 1550 nm, and 1700 nm. For the 800 nm source, we used a Kerr-lens mode-locked Ti:sapphire laser (Spectra Physics Mai-Tai HP). The chirping in the output pulses from the laser was compensated for using a prism pair, and the chirp-compensated pulses were coupled into polarization maintaining fiber (PMF) to generate the SC [21]. For the 1060 nm source, we used a commercially available, compact, diode-pumped soliton mode-locked femtosecond Nd:glass laser (High Q Laser Production). The 100 fs output pulses were coupled into high-numerical-aperture fiber to generate the SC [22]. For the 1300 nm and 1700 nm sources, we used the same laser system, consisting of a passively mode-locked, Er-doped fiber laser (IMRA femtolite B-5) that emitted 100 fs ultrashort pulses at a center wavelength of 1.56 μm as the seed source, an ultrashort-pulse fiber amplifier pumped by three high-power laser diodes at 1480 nm, and a large-mode-area photonic crystal fiber (LMA-PCF:LMA25, Thorlabs) to compensate for the chirping of the output high-power ultrashort pulses [23]. For the 1300 nm source, high-peak-power ultrashort pulses were coupled into normal-dispersion highly nonlinear fiber (ND-HNLF) to generate the SC. A bandpass filter was used at the output of the ND-HNLF to generate a Gaussian-like SC. For the 1700 nm source, high-peak-power ultrashort pulses were coupled into PMF to produce high-power, sech2-shaped soliton pulses with a center wavelength of 1700 nm. The output soliton pulses were then coupled into ND-HNLF to generate the SC [23]. For the 1550 nm source, we used a custom-made Er-doped high-power ultrashort-pulse mode-locked fiber laser, similar to that used in Ref. [24]. The chirping in the output pulses from this laser was compensated for using LMA-PCF (LMA35, Thorlabs), and the chirp-compensated pulses were coupled into ND-HNLF to produce the SC. Figure 1(a) shows the output spectra of the SC sources together with the spectrum for absorption by water. The SC spectra fit in the valleys or flat regions of the water absorption spectrum, except for the spectrum of the 1550 nm SC. The spectral bandwidths of the SC sources were 121 nm at 800 nm, 101 nm at 1060 nm, 223 nm at 1300 nm, 159 nm at 1550 nm, and 400 nm at 1700 nm. The average output powers were 30.0 mW, 31.0 mW, 12.7 mW, 19.4 mW, and 30.0 mW, respectively. Figure 1(b) shows the theoretical longitudinal resolution at each wavelength, estimated from 2ln 2/π (λc2/Δλ), where λc is the center wavelength, and Δλ is the bandwidth of the light source in the case of a Gaussian-shaped spectrum. The circles in Fig. 1(b) show the estimated longitudinal resolutions in air for the generated supercontinuum spectra.
Fig. 1

(a) Output spectra of supercontinuum sources together with water absorption spectrum and (b) corresponding theoretical longitudinal resolutions at all wavelengths as a function of bandwidth.

(a) Output spectra of supercontinuum sources together with water absorption spectrum and (b) corresponding theoretical longitudinal resolutions at all wavelengths as a function of bandwidth.

2.2. OCT setup and system characteristics

We constructed a time-domain (TD) OCT system for all wavelengths used in this study. The set-up is depicted schematically in Fig. 2 . The light from each SC source was coupled into a Michelson interferometer based on three broadband optical fiber couplers. The input beam was divided into signal and reference beams at these couplers. The signal beam was introduced into the sample arm containing a two-dimensional scanning system formed of a galvanometer mirror pair. To maintain the longitudinal resolution, the dispersion in the two arms of the interferometer was carefully balanced by using glass plates (BK7, fused silica, and so on) and an achromatic lens. The power of each SC source was 2.1–3.2 mW at the sample. Inline polarization controllers were used to adjust the polarization state of each beam. The reference arm was scanned by using a corner cube prism mounted on a galvanometer. The reference and signal beams were then combined with Coupler 1 in Fig. 2, and the interference signals were detected at a balanced detection system to cancel out the average noise.
Fig. 2

Experimental setup for time-domain optical coherence tomography.

Experimental setup for time-domain optical coherence tomography. Figure 3 shows the interference signals for all wavelengths. Since the spectral shape was nearly Gaussian, the magnitude of the sidelobes was very small on a linear scale. The observed longitudinal resolutions were 3.3–7.0 μm in air and 2.4–5.0 μm in tissue. The longitudinal resolutions in tissue at all wavelengths were under 5.0 μm, demonstrating that ultrahigh resolution was achieved.
Fig. 3

(a–e) Interference signals with mirror as sample at (a) 800 nm, (b) 1060 nm, (c) 1300 nm, (d) 1550 nm, and (e) 1700 nm UHR-OCT in air. (f) Observed longitudinal resolutions (circles) together with theoretical predictions (solid lines).

(a–e) Interference signals with mirror as sample at (a) 800 nm, (b) 1060 nm, (c) 1300 nm, (d) 1550 nm, and (e) 1700 nm UHR-OCT in air. (f) Observed longitudinal resolutions (circles) together with theoretical predictions (solid lines). We used a suitable balanced detector for each wavelength: a Si PIN photodiode-based balanced detector (PDB150A, Thorlabs) for 800 nm, an InGaAs PIN photodiode-based balanced detector (PDB150C, Thorlabs) for 1060 nm, 1300 nm, and 1550 nm, and an extended InGaAs-based balanced detector (PDA10D, Thorlabs) for 1700 nm. The detected interference signal was bandpass filtered, logarithmically demodulated, low-pass filtered, and acquired by a computer. We confirmed higher than 95 dB sensitivity at all wavelengths. In Figs. 3(c, e), the interference signals have the sidelobes caused by the non-ideal spectral shape of SC sources shown in Fig. 1(a). In the same way, 1500 nm UHR-OCT system has sidelobes in the interference signal at longer distance. They cause the small artifact in OCT images especially at the sample surface.

2.3. Phantom materials

We imaged several phantom and biological samples at all wavelengths. The tested samples were a plastic cover of a semiconductor memory card, made of polycarbonate (PC), an eraser, made of polyvinyl chloride (PVC), a magnet, made of chlorinated polyethylene (CPE) and anisotropic ferrite, a human tooth, and a pig trachea specimen. These samples were chosen to evaluate the dependence of absorption by water in typical industrially used materials and biological samples examined with UHR-OCT. For direct comparison, the OCT systems we constructed had almost the same architecture for all wavelengths. To reduce unwanted errors in estimating the attenuation coefficients, we adjusted the lateral resolution to 20.0 μm for all wavelengths. The focal point was set at the top of the sample. Figure 4 shows the spatial beam profiles at the sample surface for all wavelengths. The lateral resolutions at the sample surface were 17.6 μm at 800 nm, 19.6 μm at 1060 nm, 19.4 μm at 1300 nm, 20.6 μm at 1550 nm, and 21.4 μm at 1700 nm.
Fig. 4

Observed spatial beam profiles at the top of surface of the sample at all wavelengths.

Observed spatial beam profiles at the top of surface of the sample at all wavelengths.

2.4. Numerical analysis and measurement setting

We used the single-backscattering model to determine the total attenuation coefficient μt, which is the sum of the absorption and the scattering coefficient [25]. The principal assumption is that the direction of a wave with power P incident upon the sample is reversed only once by a single backscattering event. While the wave travels in the sample, attenuation occurs along the path of the wave before and after it is backscattered. From this model, the backscattered power returning from depth z can be expressed as where K is a constant that depends on the source coherence function, A(z) is the beam divergence function, and μ is the backscattering coefficient. Because the magnitude of the OCT signal is proportional to the amplitude of the field returning from the sample, the OCT signal on a dB scale is given by From these equations, we used the maximum likelihood estimation to extract the total attenuation coefficient by finding the condition that minimizes the quantity χ2 given by [26] We repeatedly measured each sample, and extracted the attenuation coefficient from the average of A-line scans to achieve a high degree of precision. In addition, the OCT signals between the surface and several tens of micrometers below the surface were eliminated to avoid the artifact caused by non-ideal spectral shape of light source. In the same way, the OCT signals between the tail or the changing point of slope and several tens of micrometers above there were eliminated.

3. Results

3.1. Plastic cover

Figure 5(a) shows a photograph of a memory card with a plastic cover. From the OCT images at 800 nm in Fig. 5(b) and 1700 nm in Fig. 5(c), we confirmed a deeper penetration depth in the longer wavelength region. Figures 5(d–h) show the results of averaged A-line scans from OCT signals measured at 800 nm, 1060 nm, 1300 nm, 1550 nm, and 1700 nm for quantitative comparison. The total attenuation coefficients were 4.8 mm−1 at 800 nm, 4.6 mm−1 at 1060 nm, 4.8 mm−1 at 1300 nm, 2.6 mm−1 at 1550 nm, and 2.8 mm−1 at 1700 nm. From these results, we confirmed lower attenuation coefficients in longer wavelength regions, demonstrating the usefulness of our system for measuring inside plastic samples. The OCT signal had a peak intensity of almost 20 dB at the surface of the sample, and then the reflected power decreased monotonically in all wavelength regions. The longer the center wavelength was, the lower the slope of the OCT signal intensity versus depth was.
Fig. 5

(a) Photograph of semiconductor memory card with plastic cover. (b, c) OCT images obtained at (b) 0.8 μm and (c) 1.7 μm. (d–h) Depth profiles averaged over 250 A-line scans and more than 50 iterative measurements at (d) 800 nm, (e) 1060 nm, (f) 1300 nm, (g) 1550 nm, and (h) 1700 nm. The slopes shown by the red lines were used to determine the total attenuation coefficients.

(a) Photograph of semiconductor memory card with plastic cover. (b, c) OCT images obtained at (b) 0.8 μm and (c) 1.7 μm. (d–h) Depth profiles averaged over 250 A-line scans and more than 50 iterative measurements at (d) 800 nm, (e) 1060 nm, (f) 1300 nm, (g) 1550 nm, and (h) 1700 nm. The slopes shown by the red lines were used to determine the total attenuation coefficients.

3.2. Eraser

Figure 6(a) show a photograph of an eraser. Figures 6(b–f) show the results of averaged A-line scans from OCT signals measured at 800 nm, 1060 nm, 1300 nm, 1550 nm, and 1700 nm for quantitative comparison. The total attenuation coefficients were 10.1 mm−1 at 800 nm, 8.0 mm−1 at 1060 nm, 6.6 mm−1 at 1300 nm, 6.7 mm−1 at 1550 nm, and 6.3 mm−1 at 1700 nm. For 1550 nm, we estimated the attenuation coefficient from the slope obtained by disregarding artifacts due to the properties of the light source, as indicate by the dashed line in Fig. 6(e). From these results, similar to the results in Section 3.1, we confirmed lower attenuation coefficient in longer wavelength regions, demonstrating the usefulness of the longer wavelength systems for performing measurement through rubber type materials.
Fig. 6

(a) Photograph of eraser. (b–f) Depth profiles averaged over 250 A-lines and more than 50 iterative measurements at (b) 800 nm, (c) 1060 nm, (d) 1300 nm, (e) 1550 nm, and (f) 1700 nm. The slopes shown by the red lines were used to determine the total attenuation coefficients.

(a) Photograph of eraser. (b–f) Depth profiles averaged over 250 A-lines and more than 50 iterative measurements at (b) 800 nm, (c) 1060 nm, (d) 1300 nm, (e) 1550 nm, and (f) 1700 nm. The slopes shown by the red lines were used to determine the total attenuation coefficients.

3.3. Magnet

Figure 7(a) shows a photograph of the magnet. Figures 7 (b–f) show the results of averaged A-line scans from OCT signals measured at 800 nm, 1060 nm, 1300 nm, 1550 nm, and 1700 nm for quantitative comparison. The total attenuation coefficients were 24.0 mm−1 at 800 nm, 22.3 mm−1 at 1060 nm, 13.9 mm−1 at 1300 nm, 9.6 mm−1 at 1550 nm, and 6.0 mm−1 at 1700 nm. From these results, we confirmed the strong wavelength dependence of the attenuation coefficient, which is different from the other materials in Figs. 5 and 6. For 1550 nm, similar to the results in Section 3.2, we extracted the slope of the OCT signal by disregarding artifacts due to the properties of the light source, as indicated by the dashed line in Fig. 7(e).
Fig. 7

(a) Photograph of magnet. (b–f) Depth profiles averaged over 250 A-lines and more than 50 iterative measurements at (b) 800 nm, (c) 1060 nm, (d) 1300 nm, (e) 1550 nm, and (f) 1700 nm. The slopes shown by the red lines were used to determine the total attenuation coefficients.

(a) Photograph of magnet. (b–f) Depth profiles averaged over 250 A-lines and more than 50 iterative measurements at (b) 800 nm, (c) 1060 nm, (d) 1300 nm, (e) 1550 nm, and (f) 1700 nm. The slopes shown by the red lines were used to determine the total attenuation coefficients.

3.4 .Total attenuation coefficients of low-water-absorption industrially used materials

Table 1 shows the total attenuation coefficients of the three samples of industrially used materials (plastic cover, eraser, and magnet), which were obtained from the UHR-OCT depth profiles. Figure 8 shows the wavelength dependence of the total attenuation coefficients of these samples. The ranges of variation of the attenuation coefficients over all wavelengths were 2.0 dB/mm for the plastic cover, 3.8 dB/mm for the eraser, and 18.0 dB/mm for the magnet. The magnet showed much larger wavelength dependence.
Table 1

Comparison of total attenuation coefficients of industrially used materials

WavelengthPlastic cover
Eraser
Magnet
Range
μt (mm−1)μt (mm−1)μt (mm−1)Max – Min (mm−1)
800 nm4.810.124.019.2
1060 nm4.68.022.317.7
1300 nm4.76.613.99.2
1550 nm2.66.79.67.0
1700 nm2.86.36.03.2
Max – Min2.03.818.0 
Fig. 8

Wavelength dependence of total attenuation coefficient for industrially used materials

Wavelength dependence of total attenuation coefficient for industrially used materials The longer the wavelength was, the smaller the range of variation of the total attenuation coefficient was. From Table 1 and Fig. 8, we confirmed the effectiveness of longer-wavelength UHR-OCT for samples showing low absorption by water. Moreover, there was less dependence of the total attenuation coefficient on the sample material at longer wavelengths.

3.5. Human baby tooth

Next, we examined some biological samples. Figure 9(a) shows an image of a human baby tooth. We measured the OCT signal at the red line in Fig. 9(a). Figures 9(b–f) show UHR-OCT images and corresponding A-line scans of the OCT signals at the red dashed line in each OCT image. The OCT signals in Figs. 9(b–f) have more than two slopes. The red line shows the slope at the enamel layer, and the blue line shows the slope at the dentine layer. Figure 9(g) shows the wavelength dependence of the total attenuation coefficients of the enamel and dentine layers. From 800 nm to 1550 nm, the total attenuation coefficient of the enamel layer decreased monotonically. The variation of the total attenuation coefficient of the dentine layer was 0.5 dB/mm, which is much less than that of the enamel layer, namely, 4.6 dB/mm. Thus, the enamel layer had stronger wavelength dependence than the dentine layer. Figures 9(c, d, f) reveal the existence of signals under the dentine layer. The corresponding total attenuation coefficient shown by en2 in Fig. 9(f) was 3.2 dB/mm, which is larger than the total attenuation coefficient of the dentine layer. It is considered that this signal originated from the enamel layer at the bottom side of the thin human baby tooth.
Fig. 9

(a) Photograph of human tooth sample. (b–f) OCT images at red line in (a) and depth profiles obtained at red dashed line in each of the OCT images at (b) 800 nm, (c) 1060 nm, (d) 1300 nm, (e) 1550 nm, and (f) 1700 nm. The slopes shown by the red and blue solid lines were used to determine the total attenuation coefficients. (g) Wavelength dependence of total attenuation coefficients of enamel and dentine layers in the sample. Important features inside the sample can be distinguished, such as the enamel layer (en) and the dentin layer (d).

(a) Photograph of human tooth sample. (b–f) OCT images at red line in (a) and depth profiles obtained at red dashed line in each of the OCT images at (b) 800 nm, (c) 1060 nm, (d) 1300 nm, (e) 1550 nm, and (f) 1700 nm. The slopes shown by the red and blue solid lines were used to determine the total attenuation coefficients. (g) Wavelength dependence of total attenuation coefficients of enamel and dentine layers in the sample. Important features inside the sample can be distinguished, such as the enamel layer (en) and the dentin layer (d). From these results and from comparison with the results of the industrially used materials described above, we confirmed the lower absorption by water in the human tooth sample and greater imaging depth with high contrast at a wavelength of 1700 nm.

3.6. Pig trachea

Figure 10(a) shows a photograph of the pig trachea sample. Figures 10(b–f) show the UHR-OCT images and A-line scans of the OCT signals at the red dashed line in each OCT image used for quantitative comparison. In Figs. 10(b–f), we can distinguish the epithelium, mucosa, and cartilage in the sample. The solid red line shows the slope at the mucosa layer. Figure 10(g) shows the wavelength dependence of the attenuation coefficient of the mucosa in the pig trachea sample. In Fig. 10(g), the total attenuation coefficient is highest at 800 nm, due to absorption by hemoglobin in the mucosa layer. At longer wavelengths, 1060 nm and above, the total attenuation coefficient increased. This was due to the influence of absorption by water around 1400 nm and above 1800 nm.
Fig. 10

(a) Photograph of pig trachea sample. (b–f) OCT images and depth profiles at the red dashed line in each OCT image at (b) 800 nm, (c) 1060 nm, (d) 1300 nm, (e) 1550 nm and (f) 1700 nm . The slopes shown by the solid red lines were used to determine the total attenuation coefficients. (g) Wavelength dependence of total attenuation coefficients of mucosa in trachea. (h) Three-dimensional (3D) image at 1700 nm (Media 1). A 3D image at 800 nm is also shown as a movie in Media 2. Important features inside the sample can be distinguished, such as the epithelium layer (ep), the mucosa layer (m), and cartilage (ca).

(a) Photograph of pig trachea sample. (b–f) OCT images and depth profiles at the red dashed line in each OCT image at (b) 800 nm, (c) 1060 nm, (d) 1300 nm, (e) 1550 nm and (f) 1700 nm . The slopes shown by the solid red lines were used to determine the total attenuation coefficients. (g) Wavelength dependence of total attenuation coefficients of mucosa in trachea. (h) Three-dimensional (3D) image at 1700 nm (Media 1). A 3D image at 800 nm is also shown as a movie in Media 2. Important features inside the sample can be distinguished, such as the epithelium layer (ep), the mucosa layer (m), and cartilage (ca). From these results, we confirmed the influence of absorption by water and hemoglobin on the total attenuation coefficient in the mucosa of the pig trachea sample. The total attenuation coefficient of the mucosa was smallest at 1060 nm. On the other hand, the image contrast between the epithelium, mucosa, and cartilage was much clearer at 1700 nm than at the other wavelengths. Figure 10(h) shows a reconstructed 3D image of the pig trachea at 1700 nm. Reconstructed 3D images at 800 nm and 1700 nm are also shown as movies in Media 1 and Media 2. We can see the difference in image contrast and penetration depth in the two movies.

4. Discussion

We observed wavelength dependence of the total attenuation coefficients of various samples in UHR-OCT. Figure 11(a) shows the total attenuation coefficients of the industrially used homogeneous materials. The wavelength dependence of the magnet was much stronger than those of the other samples. The dashed line in Fig. 11(a) shows the fitted curve obtained from Rayleigh scattering theory. The total attenuation coefficients were fitted to this curve by the least-squares method. They were proportional to λ−4, where λ is the center wavelength of the light source. The wavelength dependence of the total attenuation coefficients of the plastic cover and eraser were well-fitted to the λ−4 dependence. It is considered that the slight different behavior of attenuation coefficient of these two materials is caused by the material properties such as hardness, molecular weight, and density. However, the fitness for the magnet was modest. The variation of the total attenuation coefficients at longer wavelengths for the industrially used homogeneous samples was much smaller than the others, in terms of the sample dependence shown in Fig. 11(a). Therefore, it is considered that a longer-wavelength system is useful for examining many kinds of industrially used materials with only a single apparatus. For more precise analysis of optical properties in the materials, it is necessary to compensate the chromatic dispersion inside the material, especially in the shorter wavelength region. In addition, it is expected that the phantoms made of polystyrene sphere in several liquid conditions are also useful to decouple the contribution of the absorption and scattering coefficients. In the human tooth sample, the enamel and dentine layers showed almost the same characteristics as the industrial samples. This is considered to be because of low absorption by water in human teeth. The total attenuation coefficient of the mucosa layer in the pig trachea sample, depicted by the circles in Fig. 11(b), was dependent on the sum of absorption and scattering. The total attenuation coefficient at 800 nm was affected by absorption by hemoglobin, which is described by the red line in Fig. 11(b). On the other hand, the total attenuation coefficient at wavelengths longer than 1060 nm was affected by absorption by water, which resulted in a higher total attenuation coefficient as the wavelength increased. From the imaging results of the pig trachea in Fig. 10(b–f), the OCT images at 800 nm and 1700 nm had clear boundaries between the epithelium, mucosa, and cartilage. On the other hand, the imaging contrast at 1060 nm, 1300 nm, and 1550 nm did not show any dependence on wavelength. From these results, in biomedical measurement, it is considered that there is a trade-off between imaging contrast and total attenuation coefficient, which leads to the greatest possible penetration depth. In addition, it is considered that the evaluation of the wavelength dependence of OCT imaging of skin and blood is effective to investigate the contribution of water and hemoglobin absorptions.
Fig. 11

Total attenuation coefficients of (a) industrially used materials and (b) biomedical samples. The absorption coefficients of water [27] and hemoglobin [28] are shown in blue and red lines with arbitrary unit.

Total attenuation coefficients of (a) industrially used materials and (b) biomedical samples. The absorption coefficients of water [27] and hemoglobin [28] are shown in blue and red lines with arbitrary unit.

5. Conclusion

We investigated the wavelength dependence of the optical properties of samples with ultrahigh resolution optical coherence tomography (UHR-OCT) at wavelengths in the range 0.8–1.7 μm. We constructed supercontinuum sources in five wavelength regions and corresponding optical coherence tomography systems designed to match the broadband characteristics of each supercontinuum source. With these systems, we compared the total attenuation coefficients, which are mainly the sum of the scattering coefficient and absorption of the samples. In the samples of industrially used homogeneous materials showing low absorption by water, the total attenuation coefficients of the samples were fitted by Rayleigh scattering theory. The longer-wavelength systems had low scattering coefficient, leading to high-penetration-depth imaging and less dependence of the total attenuation coefficient on the sample material. One of the biological samples, the human tooth, which had low absorption by water, showed nearly the same total attenuation coefficient and penetration depth behavior. On the other hand, in the pig trachea sample, which showed absorption by water and hemoglobin, the total attenuation coefficient depended on the wavelength, due to absorption and scattering in the mucosa layer, and the imaging contrast at 800 nm and 1700 nm was much higher than at the other wavelengths.
  22 in total

1.  Dental optical coherence tomography: a comparison of two in vitro systems.

Authors:  L L Otis; B W Colston; M J Everett; H Nathel
Journal:  Dentomaxillofac Radiol       Date:  2000-03       Impact factor: 2.419

2.  Real-time optical coherence tomography of the anterior segment at 1310 nm.

Authors:  S Radhakrishnan; A M Rollins; J E Roth; S Yazdanfar; V Westphal; D S Bardenstein; J A Izatt
Journal:  Arch Ophthalmol       Date:  2001-08

Review 3.  Clinical imaging with optical coherence tomography.

Authors:  Brett E Bouma; G J Tearney
Journal:  Acad Radiol       Date:  2002-08       Impact factor: 3.173

Review 4.  Optical coherence tomography: feasibility for basic research and image-guided surgery of breast cancer.

Authors:  Stephen A Boppart; Wei Luo; Daniel L Marks; Keith W Singletary
Journal:  Breast Cancer Res Treat       Date:  2004-03       Impact factor: 4.872

5.  Real-time, ultrahigh-resolution, optical coherence tomography with an all-fiber, femtosecond fiber laser continuum at 1.5 microm.

Authors:  N Nishizawa; Y Chen; P Hsiung; E P Ippen; J G Fujimoto
Journal:  Opt Lett       Date:  2004-12-15       Impact factor: 3.776

6.  Retinal nerve fiber layer thickness map determined from optical coherence tomography images.

Authors:  Mircea Mujat; Raymond Chan; Barry Cense; B Park; Chulmin Joo; Taner Akkin; Teresa Chen; Johannes de Boer
Journal:  Opt Express       Date:  2005-11-14       Impact factor: 3.894

7.  Quantitative measurement of attenuation coefficients of weakly scattering media using optical coherence tomography.

Authors:  Dirk Faber; Freek van der Meer; Maurice Aalders; Ton van Leeuwen
Journal:  Opt Express       Date:  2004-09-20       Impact factor: 3.894

8.  Noninvasive imaging of living human skin with dual-wavelength optical coherence tomography in two and three dimensions.

Authors:  Y Pan; D L Farkas
Journal:  J Biomed Opt       Date:  1998-10       Impact factor: 3.170

9.  Optical coherence tomography.

Authors:  D Huang; E A Swanson; C P Lin; J S Schuman; W G Stinson; W Chang; M R Hee; T Flotte; K Gregory; C A Puliafito
Journal:  Science       Date:  1991-11-22       Impact factor: 47.728

10.  Quantification of macrophage content in atherosclerotic plaques by optical coherence tomography.

Authors:  Guillermo J Tearney; Hiroshi Yabushita; Stuart L Houser; H Thomas Aretz; Ik-Kyung Jang; Kelly H Schlendorf; Christopher R Kauffman; Milen Shishkov; Elkan F Halpern; Brett E Bouma
Journal:  Circulation       Date:  2003-01-07       Impact factor: 29.690

View more
  18 in total

1.  High-resolution three-dimensional in vivo imaging of mouse oviduct using optical coherence tomography.

Authors:  Jason C Burton; Shang Wang; C Allison Stewart; Richard R Behringer; Irina V Larina
Journal:  Biomed Opt Express       Date:  2015-06-30       Impact factor: 3.732

2.  Multimodal evaluation of ultra-short laser pulses treatment for skin burn injuries.

Authors:  Moises Oliveira Dos Santos; Anne Latrive; Pedro Arthur Augusto De Castro; Wagner De Rossi; Telma Maria Tenorio Zorn; Ricardo Elgul Samad; Anderson Zanardi Freitas; Carlos Lenz Cesar; Nilson Dias Vieira Junior; Denise Maria Zezell
Journal:  Biomed Opt Express       Date:  2017-02-16       Impact factor: 3.732

3.  Noninvasive, in vivo imaging of subcortical mouse brain regions with 1.7  μm optical coherence tomography.

Authors:  Shau Poh Chong; Conrad W Merkle; Dylan F Cooke; Tingwei Zhang; Harsha Radhakrishnan; Leah Krubitzer; Vivek J Srinivasan
Journal:  Opt Lett       Date:  2015-11-01       Impact factor: 3.776

4.  Tuning of successively scanned two monolithic Vernier-tuned lasers and selective data sampling in optical comb swept source optical coherence tomography.

Authors:  Dong-Hak Choi; Reiko Yoshimura; Kohji Ohbayashi
Journal:  Biomed Opt Express       Date:  2013-11-22       Impact factor: 3.732

5.  Development of a high power supercontinuum source in the 1.7 μm wavelength region for highly penetrative ultrahigh-resolution optical coherence tomography.

Authors:  H Kawagoe; S Ishida; M Aramaki; Y Sakakibara; E Omoda; H Kataura; N Nishizawa
Journal:  Biomed Opt Express       Date:  2014-02-26       Impact factor: 3.732

6.  Rapid chemically selective 3D imaging in the mid-infrared.

Authors:  Eric O Potma; David Knez; Yong Chen; Yulia Davydova; Amanda Durkin; Alexander Fast; Mihaela Balu; Brenna Norton-Baker; Rachel W Martin; Tommaso Baldacchini; Dmitry A Fishman
Journal:  Optica       Date:  2021-07-07       Impact factor: 11.104

7.  Four-dimensional live imaging of hemodynamics in mammalian embryonic heart with Doppler optical coherence tomography.

Authors:  Shang Wang; David S Lakomy; Monica D Garcia; Andrew L Lopez; Kirill V Larin; Irina V Larina
Journal:  J Biophotonics       Date:  2016-03-21       Impact factor: 3.207

8.  Computational analysis of six optical coherence tomography systems for vocal fold imaging: A comparison study.

Authors:  Tiffany T Pham; Lily Chen; Andrew E Heidari; Jason J Chen; Alisa Zhukhovitskaya; Yan Li; Urja Patel; Zhongping Chen; Brian J F Wong
Journal:  Lasers Surg Med       Date:  2019-01-25       Impact factor: 4.025

9.  Three-dimensional, non-invasive, cross-sectional imaging of protein crystals using ultrahigh resolution optical coherence tomography.

Authors:  Norihiko Nishizawa; Shutaro Ishida; Mika Hirose; Shigeru Sugiyama; Tsuyoshi Inoue; Yusuke Mori; Kazuyoshi Itoh; Hiroyoshi Matsumura
Journal:  Biomed Opt Express       Date:  2012-03-15       Impact factor: 3.732

10.  Enhanced antitumor activity of the photosensitizer meso-Tetra(N-methyl-4-pyridyl) porphine tetra tosylate through encapsulation in antibody-targeted chitosan/alginate nanoparticles.

Authors:  Sharif M Abdelghany; Daniela Schmid; Jill Deacon; Jakub Jaworski; Francois Fay; Kirsty M McLaughlin; Julie A Gormley; James F Burrows; Daniel B Longley; Ryan F Donnelly; Christopher J Scott
Journal:  Biomacromolecules       Date:  2013-01-31       Impact factor: 6.988