Literature DB >> 22307271

Evaluation of stent visibility by flat panel detector CT in patients treated for intracranial aneurysms.

Frédéric Clarençon1, Michel Piotin, Silvia Pistocchi, Drazenko Babic, Raphaël Blanc.   

Abstract

INTRODUCTION: This study aimed to evaluate the visibility of stents using high-resolution computed tomography (CT) acquisitions acquired with flat panel detector (XperCT, Allura series, Philips Healthcare, The Netherlands) for endovascular treatment of intracranial aneurysms.
METHODS: On a 24-month period, 48 patients endovascularly treated by coiling and stenting (59 stents) for intracranial aneurysms were explored by flat panel detector CT technique. A sequence of 620 2D images was acquired over an angle of 240° using a 1,024 × 1,024 pixel matrix detector within a 48-cm field of view. The images were retrospectively analyzed independently by two neuroradiologists. Evaluation criteria were percentage of visualization of the stents and stent deployment (kinking or unsatisfactory deployment of the stent).
RESULTS: Evaluation of the stent was feasible for all the patients. Stent visibility by XperCT was overall estimated at 76% of the stent length. Difficulties to analyze the stents were related to coil artifacts but not to packing density or aneurysm location. Stent length visualization was higher when the acquisition was performed before additional coiling (P < 0.0001). Mild kinking/misdeployment was noticed in 22% of the cases.
CONCLUSION: XperCT technique provides multiplanar and 3D reconstructions that allows for a satisfying visualization of intracranial stents. This CT-like acquisition should be performed after the stent deployment and before coiling, in order to obtain better stent visualization.

Entities:  

Mesh:

Year:  2012        PMID: 22307271     DOI: 10.1007/s00234-011-1002-8

Source DB:  PubMed          Journal:  Neuroradiology        ISSN: 0028-3940            Impact factor:   2.804


  17 in total

1.  Angiographic CT in cerebrovascular stenting.

Authors:  Götz Benndorf; Charles M Strother; Benjamin Claus; Ramin Naeini; Hesham Morsi; Richard Klucznik; Michael E Mawad
Journal:  AJNR Am J Neuroradiol       Date:  2005-08       Impact factor: 3.825

2.  Stent conformity in curved vascular models with simulated aneurysm necks using flat-panel CT: an in vitro study.

Authors:  N Ebrahimi; B Claus; C-Y Lee; A Biondi; G Benndorf
Journal:  AJNR Am J Neuroradiol       Date:  2007-05       Impact factor: 3.825

3.  Endovascular reconstruction with the Neuroform stent as monotherapy for the treatment of uncoilable intradural pseudoaneurysms.

Authors:  David Fiorella; Felipe C Albuquerque; Vivek R Deshmukh; Henry H Woo; Peter A Rasmussen; Thomas J Masaryk; Cameron G McDougall
Journal:  Neurosurgery       Date:  2006-08       Impact factor: 4.654

4.  Dense packing of cerebral aneurysms: an in vitro study with detachable platinum coils.

Authors:  M Piotin; S Mandai; K J Murphy; K Sugiu; P Gailloud; J B Martin; D A Rüfenacht
Journal:  AJNR Am J Neuroradiol       Date:  2000-04       Impact factor: 3.825

5.  Brain imaging with a flat detector C-arm : Technique and clinical interest of XperCT.

Authors:  M Söderman; D Babic; S Holmin; T Andersson
Journal:  Neuroradiology       Date:  2008-06-17       Impact factor: 2.804

6.  Stent-assisted coiling of intracranial aneurysms: clinical and angiographic results in 216 consecutive aneurysms.

Authors:  Michel Piotin; Raphaël Blanc; Laurent Spelle; Charbel Mounayer; Rhelen Piantino; Paul J Schmidt; Jacques Moret
Journal:  Stroke       Date:  2009-12-03       Impact factor: 7.914

7.  Spontaneous delayed proximal migration of enterprise stent after staged treatment of wide-necked basilar aneurysm: technical case report.

Authors:  Sean D Lavine; Philip M Meyers; E Sander Connolly; Robert S Solomon
Journal:  Neurosurgery       Date:  2009-05       Impact factor: 4.654

Review 8.  Advances in interventional neuroradiology.

Authors:  Ajay K Wakhloo; Michael J Deleo; Martin M Brown
Journal:  Stroke       Date:  2009-05       Impact factor: 7.914

9.  Neuroform stent-assisted coil embolization of wide-neck intracranial aneurysms: strategies in stent deployment and midterm follow-up.

Authors:  Alessandra Biondi; Vallabh Janardhan; Jeffrey M Katz; Kimberly Salvaggio; Howard A Riina; Y Pierre Gobin
Journal:  Neurosurgery       Date:  2007-09       Impact factor: 4.654

10.  Angiographic CT with intravenous administration of contrast medium is a noninvasive option for follow-up after intracranial stenting.

Authors:  Jan-Hendrik Buhk; Paul Lingor; Michael Knauth
Journal:  Neuroradiology       Date:  2008-02-02       Impact factor: 2.804

View more
  12 in total

1.  Wall Apposition Is a Key Factor for Aneurysm Occlusion after Flow Diversion: A Histologic Evaluation in 41 Rabbits.

Authors:  A Rouchaud; C Ramana; W Brinjikji; Y-H Ding; D Dai; T Gunderson; J Cebral; D F Kallmes; R Kadirvel
Journal:  AJNR Am J Neuroradiol       Date:  2016-07-07       Impact factor: 3.825

2.  Artifact reduction of different metallic implants in flat detector C-arm CT.

Authors:  S-C Hung; C-C Wu; C-J Lin; W-Y Guo; C-B Luo; F-C Chang; C-Y Chang
Journal:  AJNR Am J Neuroradiol       Date:  2014-01-23       Impact factor: 3.825

3.  CT reconstruction and MRI fusion of 3D rotational angiography in the evaluation of pediatric cerebrovascular lesions.

Authors:  Prakash Muthusami; Nicholas Shkumat; Vanessa Rea; Albert H Chiu; Manohar Shroff
Journal:  Neuroradiology       Date:  2017-03-27       Impact factor: 2.804

4.  Impact of a new metal artefact reduction algorithm in the noninvasive follow-up of intracranial clips, coils, and stents with flat-panel angiographic CTA: initial results.

Authors:  Marios-Nikos Psychogios; Bernhard Scholz; Christopher Rohkohl; Yiannis Kyriakou; Alexander Mohr; Peter Schramm; Dorothee Wachter; Katrin Wasser; Michael Knauth
Journal:  Neuroradiology       Date:  2013-04-16       Impact factor: 2.804

5.  Contrast-enhanced angiographic cone-beam computed tomography without pre-diluted contrast medium.

Authors:  K I Jo; S R Kim; J H Choi; K H Kim; P Jeon
Journal:  Neuroradiology       Date:  2015-08-21       Impact factor: 2.804

6.  Metal artifact reduction algorithm for image quality improvement of cone-beam CT images of medium or large cerebral aneurysms treated with stent-assisted coil embolization.

Authors:  Satoshi Murai; Masafumi Hiramatsu; Yuji Takasugi; Yu Takahashi; Naoya Kidani; Shingo Nishihiro; Yukei Shinji; Jun Haruma; Tomohito Hishikawa; Kenji Sugiu; Isao Date
Journal:  Neuroradiology       Date:  2019-11-07       Impact factor: 2.804

7.  Visual outcomes with flow-diverter stents covering the ophthalmic artery for treatment of internal carotid artery aneurysms.

Authors:  A Rouchaud; O Leclerc; Y Benayoun; S Saleme; Y Camilleri; F D'Argento; M-P Boncoeur; P-Y Robert; C Mounayer
Journal:  AJNR Am J Neuroradiol       Date:  2014-10-22       Impact factor: 3.825

8.  Fatal arterial rupture during angioplasty of a flow diverter in a recurrent, previously Y-stented giant MCA bifurcation aneurysm.

Authors:  Jean-Christophe Gentric; Robert Fahed; Tim E Darsaut; Igor Salazkin; Daniel Roy; Jean Raymond
Journal:  Interv Neuroradiol       Date:  2016-02-17       Impact factor: 1.610

9.  Effect of structural remodeling (retraction and recoil) of the pipeline embolization device on aneurysm occlusion rate.

Authors:  L-D Jou; B D Mitchell; H M Shaltoni; M E Mawad
Journal:  AJNR Am J Neuroradiol       Date:  2014-04-10       Impact factor: 3.825

Review 10.  Toward the era of a one-stop imaging service using an angiography suite for neurovascular disorders.

Authors:  Sheng-Che Hung; Chung-Jung Lin; Wan-Yuo Guo; Feng-Chi Chang; Chao-Bao Luo; Michael Mu-Huo Teng; Cheng-Yen Chang
Journal:  Biomed Res Int       Date:  2013-05-14       Impact factor: 3.411

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.