AIM: This study aimed to determine the effect of a cuff properly sized for mid-upper arm circumference on blood pressure measurement in obese surgical patients.Entities:
Keywords: Blood pressure measurement; cuff size; mid-upper arm circumference; obese patients; obesity
Year: 2020 PMID: 34263199 PMCID: PMC8152164 DOI: 10.5152/FNJN.2020.19119
Source DB: PubMed Journal: Florence Nightingale J Nurs ISSN: 2687-6442
Figure 1The sample and study protocol flow chart
BP: Blood pressure; BMI: Body mass index; MUAC: Mid-upper arm circumference
Descriptive characteristics of patients (n=100)
| Characteristics (n=100) | n | % | |
|---|---|---|---|
| Age (Mean±SD) (Min-Max) 52.82±13.95 (22–75 years) | |||
| Gender | Female | 81 | 81.0 |
| Male | 19 | 19.0 | |
| Marital status | Married | 81 | 81.0 |
| Single | 19 | 19.0 | |
| Diagnosis | Cancer | 32 | 32.0 |
| Cholelithiasis | 37 | 37.0 | |
| Multinodular goiter | 31 | 31.0 | |
| Classification of obesity | 30–34.99 kg/m2 (Class I) | 37 | 37.0 |
| 35–39.99 kg/m2 (Class III) | 24 | 24.0 | |
| >40 kg/m2 (Class III-Morbid obesity) | 39 | 39.0 | |
| Classification of MUACd | 27–34 cm | 33 | 33.0 |
| 35–44 cm | 67 | 67.0 | |
SD: Standard deviation; Min: Minimum; Max: Maximum; MUAC: Mid–upper arm circumference
Comparison of blood pressure readings according to different cuff sizes
| BP | BP differences (%) | Cuff size | Test | p | |||
|---|---|---|---|---|---|---|---|
|
| |||||||
| Small adult cuff | Proper cuff for MUAC | Difference | |||||
|
| |||||||
| >10 mmHg | >20 mmHg | Mean (SD) mmHg | Mean (SD) mmHg | Mean (SD) mmHg | |||
| Systolic | 93 | 53 | 134.03 (18.74) | 113.25 (16.09) | 20.78 (11.07) | 18.775 | <0.001 |
| Diastolic | 66 | 18 | 78.88 (11.93) | 68.74 (7.88) | 10.15 (10.26) | 9.888 | <0.001 |
BP: Blood pressure; SD: Standard deviation; MUAC: Mid–upper arm circumference.
Paired sample t test.
Figure 2Hypertension prevalence by different cuff sizes readings
* Pearson Chi-square test
MUAC: Mid-upper arm circumference