Abstract

BACKGROUND: Recently, several cases of symptomatic and/or electrically detectable intracardiac inside-out abrasions in silicon-coated Riata® and Riata® ST leads have been described. However, the prevalence in asymptomatic patients with unremarkable implantable cardioverter defibrillator (ICD) interrogation is unknown. The aim of this study was to determine the prevalence of asymptomatic and electrically undetectable intracardiac inside-out abrasion in silicon-coated Riata® and Riata® ST leads.
METHODS: All 52 patients with an active silicone-coated Riata® and Riata® ST lead followed up in our outpatient clinic were scheduled for a premature ICD interrogation and a biplane chest radiograph. When an intracardiac inside-out abrasion was suspected, this finding was confirmed by fluoroscopy.
RESULTS: Mean time since implantation was 71 ± 18 months. An intracardiac inside-out abrasion was confirmed by fluoroscopy in 6 patients (11.5%). Mean time from lead implantation to detection of intracardiac inside-out abrasion was 79 ± 14 months. In all patients with an intracardiac inside-out abrasion, ICD interrogation showed normal and stable electrical parameters. Retrospectively, in 4 of these 6 patients, a coronary angiography performed 25 ± 18 months before diagnosis of intracardiac inside-out abrasion already showed the defect. Despite undetected intracardiac inside-out abrasion, 2 of these 4 patients experienced adequate antitachycardia pacing and ICD-shocks. ICD leads were replaced in all 6 patients.
CONCLUSIONS: The prevalence of asymptomatic intracardiac inside-out abrasion in silicon-coated Riata® and Riata® ST leads is higher than 10% when assessed by fluoroscopy, and most intracardiac inside-out abrasions are not detectable by ICD interrogation.
Copyright © 2011 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22236515     DOI: 10.1016/j.ijcard.2011.12.076

Source DB:  PubMed          Journal:  Int J Cardiol        ISSN: 0167-5273            Impact factor:   4.164


  7 in total

Review 1.  Considerations for cardiac device lead extraction.

Authors:  Oussama Wazni; Bruce L Wilkoff
Journal:  Nat Rev Cardiol       Date:  2016-01-29       Impact factor: 32.419

2.  Prospective evaluation of cinefluoroscopy and chest radiography for Riata lead defects: implications for future lead screening.

Authors:  Peem Lorvidhaya; Ivan Mendoza; Sharmila Sehli; Michael K Atalay; Michael H Kim
Journal:  J Interv Card Electrophysiol       Date:  2013-09-08       Impact factor: 1.900

3.  What did we learn from Riata™?

Authors:  Avi Fischer; C Jenney
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2015-06-09

4.  Increasing lead burden correlates with externalized cables during systematic fluoroscopic screening of Riata leads.

Authors:  Liza P Moorman; J Randall Moorman; John P DiMarco; Rohit Malhotra; Andrew Darby; Kenneth Bilchick; John D Ferguson; J Michael Mangrum; Sandeep Kamath; Pamela K Mason
Journal:  J Interv Card Electrophysiol       Date:  2012-12-20       Impact factor: 1.900

5.  Long-term performance of the St Jude Riata 1580-1582 ICD lead family.

Authors:  S D A Valk; D A M J Theuns; L Jordaens
Journal:  Neth Heart J       Date:  2013-03       Impact factor: 2.380

6.  First time description of early lead failure of the Linox Smart lead compared to other contemporary high-voltage leads.

Authors:  Vanessa Weberndörfer; Tobias Nyffenegger; Ian Russi; Miriam Brinkert; Benjamin Berte; Stefan Toggweiler; Richard Kobza
Journal:  J Interv Card Electrophysiol       Date:  2018-05-01       Impact factor: 1.900

7.  A case of riata® dual coil defibrillator lead failure in a patient with ventricular fibrillation.

Authors:  Kyu Choi; Jung Hee Kim; Hyo Jin Kim; Se-Ok Lee; Eun-Young Jang; June Soo Kim
Journal:  Korean Circ J       Date:  2013-05-31       Impact factor: 3.243

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.