Literature DB >> 22231359

Developing an algorithm for pulse oximetry derived respiratory rate (RR(oxi)): a healthy volunteer study.

Paul S Addison1, James N Watson, Michael L Mestek, Roger S Mecca.   

Abstract

OBJECTIVE: The presence of respiratory information within the pulse oximeter signal (PPG) is a well-documented phenomenon. However, extracting this information for the purpose of continuously monitoring respiratory rate requires: (1) the recognition of the multi-faceted manifestations of respiratory modulation components within the PPG and the complex interactions among them; (2) the implementation of appropriate advanced signal processing techniques to take full advantage of this information; and (3) the post-processing infrastructure to deliver a clinically useful reported respiratory rate to the end user. A holistic algorithmic approach to the problem is therefore required. We have developed the RR(OXI) algorithm based on this principle and its performance on healthy subject trial data is described herein.
METHODS: Finger PPGs were collected from a cohort of 139 healthy adult volunteers monitored during free breathing over an 8-min period. These were subsequently processed using a novel in-house algorithm based on continuous wavelet transform technology within an infrastructure incorporating weighted averaging and logical decision making processes. The computed oximeter respiratory rates (RR(oxi)) were then compared to an end-tidal CO2 reference rate RR(ETCO2).
RESULTS: RR(ETCO2) ranged from a lowest recorded value of 2.97 breaths per min (br/min) to a highest value of 28.02 br/min. The mean rate was 14.49 br/min with standard deviation of 4.36 br/min. Excellent agreement was found between RR(oxi) and RR(ETCO2), with a mean difference of -0.23 br/min and standard deviation of 1.14 br/min. The two measures are tightly spread around the line of agreement with a strong correlation observable between them (R2 = 0.93).
CONCLUSIONS: These data indicate that RR(oxi) represents a viable technology for the measurement of respiratory rate of healthy individuals.

Entities:  

Mesh:

Year:  2012        PMID: 22231359      PMCID: PMC3268017          DOI: 10.1007/s10877-011-9332-y

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  20 in total

Review 1.  Forehead oximetry in critically ill patients: the case for a new monitoring site.

Authors:  Richard D Branson; Paul D Mannheimer
Journal:  Respir Care Clin N Am       Date:  2004-09

2.  Respiratory rate extraction via an autoregressive model using the optimal parameter search criterion.

Authors:  J Lee; K H Chon
Journal:  Ann Biomed Eng       Date:  2010-05-25       Impact factor: 3.934

3.  An algorithm for the detection of individual breaths from the pulse oximeter waveform.

Authors:  Paul Leonard; Neil R Grubb; Paul S Addison; David Clifton; James N Watson
Journal:  J Clin Monit Comput       Date:  2004-12       Impact factor: 2.502

4.  Extracting breathing rate information from a wearable reflectance pulse oximeter sensor.

Authors:  W S Johnston; Y Mendelson
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2004

5.  Estimation of respiratory rate from photoplethysmogram data using time-frequency spectral estimation.

Authors:  Ki H Chon; Shishir Dash; Kihwan Ju
Journal:  IEEE Trans Biomed Eng       Date:  2009-04-14       Impact factor: 4.538

6.  Monitoring of respiratory and heart rates using a fibre-optic sensor.

Authors:  L G Lindberg; H Ugnell; P A Oberg
Journal:  Med Biol Eng Comput       Date:  1992-09       Impact factor: 2.602

7.  An automated algorithm for determining respiratory rate by photoplethysmogram in children.

Authors:  Paul A Leonard; David Clifton; Paul S Addison; James N Watson; Tom Beattie
Journal:  Acta Paediatr       Date:  2006-09       Impact factor: 2.299

8.  Monitoring of heart and respiratory rates in newborn infants using a new photoplethysmographic technique.

Authors:  A Johansson; P A Oberg; G Sedin
Journal:  J Clin Monit Comput       Date:  1999-12       Impact factor: 2.502

9.  Extracting respiratory data from pulse oximeter plethysmogram traces in newborn infants.

Authors:  D Wertheim; C Olden; E Savage; P Seddon
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2008-11-17       Impact factor: 5.747

10.  Measurement of respiratory rate from the photoplethysmogram in chest clinic patients.

Authors:  David Clifton; J Graham Douglas; Paul S Addison; James N Watson
Journal:  J Clin Monit Comput       Date:  2006-11-25       Impact factor: 1.977

View more
  22 in total

Review 1.  Journal of Clinical Monitoring and Computing 2015 end of year summary: respiration.

Authors:  D S Karbing; S E Rees; M B Jaffe
Journal:  J Clin Monit Comput       Date:  2015-12-30       Impact factor: 2.502

2.  Exploration of key stakeholders' preferences for pre-hospital physiologic monitoring by emergency rescue services.

Authors:  Alasdair J Mort; Gordon F Rushworth
Journal:  J Clin Monit Comput       Date:  2013-05-26       Impact factor: 2.502

3.  Acoustic sensor versus electrocardiographically derived respiratory rate in unstable trauma patients.

Authors:  Shiming Yang; Ashley Menne; Peter Hu; Lynn Stansbury; Cheng Gao; Nicolas Dorsey; William Chiu; Stacy Shackelford; Colin Mackenzie
Journal:  J Clin Monit Comput       Date:  2016-06-07       Impact factor: 2.502

4.  Robust respiration detection from remote photoplethysmography.

Authors:  Mark van Gastel; Sander Stuijk; Gerard de Haan
Journal:  Biomed Opt Express       Date:  2016-11-03       Impact factor: 3.732

5.  Comparison of Nellcor™ PM1000N and Masimo Radical-7® for detecting apnea in volunteers.

Authors:  Chanatthee Kitsiripant; Tomoko Fukada; Hiroko Iwakiri; Yuri Tsuchiya; Makoto Ozaki; Minoru Nomura
Journal:  J Anesth       Date:  2017-07-20       Impact factor: 2.078

6.  Non-invasive continuous respiratory monitoring using temperature-based sensors.

Authors:  Daniel E Hurtado; Angel Abusleme; Javier A P Chávez
Journal:  J Clin Monit Comput       Date:  2019-06-03       Impact factor: 2.502

7.  Identification of apnea during respiratory monitoring using support vector machine classifier: a pilot study.

Authors:  Paruthi Pradhapan; Muthukaruppan Swaminathan; Hari Krishna Salila Vijayalal Mohan; N Sriraam
Journal:  J Clin Monit Comput       Date:  2012-11-21       Impact factor: 2.502

8.  Toward a Robust Estimation of Respiratory Rate From Pulse Oximeters.

Authors:  Marco A F Pimentel; Alistair E W Johnson; Peter H Charlton; Drew Birrenkott; Peter J Watkinson; Lionel Tarassenko; David A Clifton
Journal:  IEEE Trans Biomed Eng       Date:  2016-11-18       Impact factor: 4.538

9.  Increasing signal processing sophistication in the calculation of the respiratory modulation of the photoplethysmogram (DPOP).

Authors:  Paul S Addison; Rui Wang; Alberto A Uribe; Sergio D Bergese
Journal:  J Clin Monit Comput       Date:  2014-09-11       Impact factor: 2.502

10.  Calculation of the Respiratory Modulation of the Photoplethysmogram (DPOP) Incorporating a Correction for Low Perfusion.

Authors:  Paul S Addison; Rui Wang; Scott J McGonigle; Alberto A Uribe; Sergio D Bergese
Journal:  Anesthesiol Res Pract       Date:  2014-08-07
View more