| Literature DB >> 22231209 |
Martin Szegedi1, Vikren Sarkar, Prema Rassiah-Szegedi, Brian Wang, Y Jessica Huang, Hui Zhao, Bill Salter.
Abstract
In the AAPM Report 80, the imaging modality of 4D CT and respiration-correlated CT was declared a "promising solution for obtaining high-quality CT data in the presence of respiratory motion". To gather anatomically correct data over time, the existence of correlation between the internal organ movement and an external surrogate has to be assumed. For the in-house evaluation of such correlation, we retrospectively analyzed 21 four-dimensional computer tomography (4D CT) scans of five patients, out of which the artifacts experienced in three patients are shown here. To provide context and a baseline for the analysis of patient motion, a real-tissue liver phantom was used with a solid water block and liver tissue. The superior-inferior motion of fiducials in phantom and patients was correlated to the recorded anterior-posterior motion of an external surrogate marker on the chest. The use of a solid water block yielded a measurable correlation coefficient of 0.98 or better using a sinusoidal animation pattern. With sinusoidally-animated liver tissue, the minimum correlation observed was 0.96. Comparing this to retrospective patient data, we found three cases of a change in the correlation coefficient, or simply a low correlation. The source of this low correlation was investigated by careful examination of the breathing traces and the CT-phase assignments used to reconstruct the datasets. Consequences of nonregular breathing are elaborated on. We demonstrate the impact of wrong phase assignments and missing image information in the 4D CT phase sampling processes. We also show how daily patient-based correlation analysis can indicate changes in breathing traces, which can be significant enough to decrease, or completely eliminate, previously existing correlation.Entities:
Mesh:
Year: 2012 PMID: 22231209 PMCID: PMC5716128 DOI: 10.1120/jacmp.v13i1.3564
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Liver phantom containment. The arrows indicate piston motion direction.
Maximum motion envelope for the marker centroids in all three directions and its summary 3D vector length. The maximum phase error as given by the GE Advantage Windows software is listed in the last column.
|
|
|
|
|
| |
|---|---|---|---|---|---|
| Pat. 1 | 11.9 | 4.0 | 2.5 | 12.8 | 23 |
| Pat. 2 | 4.6 | 2.9 | 1.3 | 5.6 | 13 |
| Pat. 3 | 10.3 | 4.4 | 0.8 | 11.2 | 11 |
| Pat. 4 | 12.5 | 6.9 | 4.2 | 14.9 | 35 |
| Pat. 5 | 11.6 | 7.5 | 1.3 | 13.9 | 26 |
Maximum motion envelope of the porcine liver for the fiducial centroid in all three directions and its summary 3D vector length. The maximum phase error as given by the GE Advantage Windows software is listed in the last column.
|
|
|
|
|
| |
|---|---|---|---|---|---|
| Porcine liver sinusoidal trace | 17.2 | 4.0 | 1.8 | 17.7 | 2 |
| Porcine live patient trace | 17.2 | 4.0 | 1.8 | 17.7 | 4 |
| Rigid object sinuisoidal trace | 20 |
|
| 20 | 2 |
| Rigid object patient trace | 20 |
|
| 20 | 2 |
Correlation values of the 4D CT internal fiducial in SI direction to external surrogate. Values in bold & italics indicate correlation values of less than or equal to 0.5. The last two columns show correlation data measured with the phantom for a porcine liver and a rigid object.
|
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|
| Fraction 1 |
| 0.77 | 0.92 | 0.79 | 0.77 | 0.99 sin | 1.00 sin |
| Fraction 2 | 0.97 | 0.90 | 0.86 |
| 0.91 | 0.71 pat | 0.82 pat |
| Fraction 3 | 0.87 | 0.85 | 0.94 |
| 0.87 | 0.96 sin | 0.98 sin |
| Fraction 4 | 0.87 | 0.80 | 0.91 | 0.70 | 0.69 | 0.83 pat | 0.90 pat |
| Fraction 5 |
|
Due to lack of fiducials in one CT phase, this correlation has been calculated with two averaged positions in this CT phase.
Figure 2Correlation of the centroid of fiducials in porcine liver to an external marker for SIN‐ and PAT‐animation.
Figure 3Typical inhale pattern of patient 2 as captured from the RPM device from the 4D CT acquired for fraction 2. The trace shows the total time of image generation.
Figure 4Typical inhale pattern of patient 4 as captured from the RPM device from the 4D CT acquired for the last fraction. The trace shows the total time of image generation.